J.P. SELECTA FRIGITERM TFT-10 Instruction Manual page 15

Recirculation baths
Table of Contents

Advertisement

Available languages

Available languages

MANUAL DE INSTRUCCIONES
SEL change to yellow
colour when programm is
selected
PROGRAM 04
TEMP: 50.0ºC
TIME: 00:00
UNDERTEMP: 0ºC
OVERTEMP: 0ºC
SETUP
Fig 7.4. Selecting a program
28.9
ºC
50.0
SET
Fig 7.5. Operation screen
Fig 7.6.1 INTERNAL circulation.
Fig 7.6.2 EXTERNAL circulation.
J
.P. SELECTA s.a.u. - Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
CODIGO 80324 REV 0
7.4 Selecting a program for operation
To select a stored program press over SEL
(SEL indication changes its color to yellow).
SEL
Once selected, this program in placed on the WORKING screen.
SAVE
HH:MM:SS DD/MM/YY
7.5 Operation screen
On the operation screen «TASK», the program selected on the programming
screen is shown (See 7.4) and it starts when pressing over
This screen shows:
PROG 04
SET -- h -- m
T
00 h 00m
Actual temperature
Temperature set
FLOW
Working time set (SET)
80%
Elapsed time
HH:MM:SS DD/MM/YY
7.6 Circulation pump
The liquid circulation pump could operate in two confi guration modes:
- INTERNAL circulation:
Water (or other) will be pumped from horizontal pump output.
(See fi g 7.6.1).
Use this confi guration to made water circulation into the bath vessel in order
to homogenize the temperature from sample load.
- EXTERNAL circulation:
Water (or other) will be pumped from vertical pump output.
(See fi g 7.6.2).
Use this confi guration to made water circulation out of the bath vessel in order
to temperate an external equipment.
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Nov/2015
(Sujetas a modifi caciones sin previo aviso)
Pag.: 15
.