Telestream shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Telestream service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
Page 3
DEFINITIONS "Telestream" means Telestream, LLC, a Delaware limited liability company, and its Affiliates , or the local Telestream legal entity that is supplying the equipment. "Program" means the Telestream software product (executable program and/or data) enclosed with this Agreement or included within the equipment with which this Agreement is packed.
Page 4
Title to the Program and all copies thereof, but not the media on which the Program or copies may reside, shall be and remain with Telestream or others for whom Telestream has obtained a respective licensing right. Customer shall pay when due all property...
Page 5
Telestream. This Agreement and the license granted herein shall be governed by the laws of the state of California. nAll questions regarding this Agreement or the license granted herein should be directed to the nearest Telestream Sales Office. PRISM MPI2-10-MPX2-10 User Manual...
Page 7
Configure to Decode Timecode Configure Reference Settings Set and Recall Instrument Presets Manage Trace and Graticule Intensity Set Time and Date Upgrade Instrument Firmware Upgrade Software License Troubleshooting Functions Headphone/Speaker Volume, Balance, Device, and Source Adjustment PRISM MPI2-10-MPX2-10 User Manual...
Page 8
CIE Application Dolby Status Application PTP Operational Overview PTP Introduction Five Basic PTP Timing Messages Profiles and Domains One-step and Two-step Operation Multicast, Unicast, and Mixed Communication Modes BMCA Operation Compensating for Causes of Asymmetric Delay PRISM MPI2-10-MPX2-10 User Manual...
Page 9
Measurement, Control, and Laboratory Use - Part 1: General Requirements. Environmental Perchlorate Materials: this product contains one or more type CR lithium batteries. According to the state of California, CR lithium batteries are classified as perchlorate materials and require special handling. www.dtsc.ca.gov/hazardouswaste/perchlorat for additional information. PRISM MPI2-10-MPX2-10 User Manual...
Page 10
SLOVENIA, SPAIN, SWEDEN, UNITED KINGDOM, !CELANO, LICHTENSTEIN, NORWAY, SWITZERLAND Declaration of Conformity Marking by the “CE” symbol indicates compliance with the Essential Requirements of the EMC Directive of the European Union 2014/30/EU This equipment meets the following conformance standards: PRISM MPI2-10-MPX2-10 User Manual...
Page 12
RoHS 3 Directives 2011 /65/ EU and EU 2015/863 until July 22, 2021. This product does, however, comply with the RoHS 2 Directive 2011/65/EU. Korea Compliance Statement PRISM MPI2-10-MPX2-10 User Manual...
Only qualified personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment. Before use, always check the product with a known source to be sure it is operating correctly. PRISM MPI2-10-MPX2-10 User Manual...
Page 14
Use only specified replacement parts. Do not operate in wet/damp conditions: Be aware that condensation may occur if a unit is moved from a cold to a warm environment. Do not operate in an explosive atmosphere. PRISM MPI2-10-MPX2-10 User Manual...
Page 15
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries. Use only the Telestream rackmount hardware specified for this product. Service Safety Summary The Service safety summary section contains additional information required to safely perform service on the product.
Page 16
WARNING: To prevent injury or death, power off the instrument and disconnect it from line voltage before cleaning. WARNING: To reduce the risk of fire and shock, ensure that the mains supply voltage fluctuations do not exceed 10% of the operating voltage range. PRISM MPI2-10-MPX2-10 User Manual...
Page 17
Caution: After the instrument has started the upgrade process, DO NOT remove power from the instrument. If you do so, the instrument flash will be corrupted. The instrument will have to be sent to a Telestream factory service center to have the system firmware restored.
"How to use instrument menus. Where to find more information The full User Manual, Release Notes, and other information about your product are available for download at www.telestream.net/video/resources.htm. Table 1 lists all the documentation for the PRISM SDI / IP Waveform Monitor. Product documentation...
Web Site: http://www.telestream.net/telestream-support Support Email: videosupport@telestream.net US and Canada Toll-free: 1-844-219-5329 Outside of US and Canada: 1-503-967-9833 Telestream Web Site: www.telestream.net Sales and Marketing Email: info@telestream.net Telestream 848 Gold Flat Road, Suite 1 Nevada City, CA. USA 95959 PRISM MPI2-10-MPX2-10 User Manual...
Page 20
Preface Resource Contact Information International Web Site: www.telestream.net Distributor Support See the Telestream Web site for your regional authorized Telestream distributor. Telestream Technical Email: techwriter@telestream.net Writers If you have comments or suggestions about improving this document, or other Telestream documents—or if you've discovered an error or omission, please email us.
Getting Started This section helps you set up and begin to use the PRISM SDI / IP Waveform Monitor. It is divided into subsections: Product description describes your instrument and provides a list of key features. Controls and connectors shows you how to verify that you have received all of the parts of your instrument and provides a list of standard and optional accessories and a list of options that are available for your instrument.
The PRISM SDI / IP Waveform Monitor is available in two form factors: MPI2-10 The PRISM MPI2-10 is 3RU half-rack width with integrated 9-inch HD display and touch panel. Option PHY is available to add physical layer measurement sets. MPX2-10 The PRISM MPX2-10 is 1RU full-rack width with optional integrated speakers.
Page 23
Monitor PTP trend graphs to ensure proper IP system setup for robust sync system • Telestream patented Timing display, the SDI versus Analog reference and ST2022-6 / ST2110-20 / SDI / Analog reference versus PTP reference makes facility timing easy •...
50/59.94/60p Base instrument 3G-SDI Level B 1920x1080 4:2:2 YCbCr 50/59.94/60p Base instrument Quad Link HD-SDI 3840x2160 4:2:2 YCbCr 23.98/24/25/ MP2-FMT-4K Square Division 29.97/30p, and Quad Link 3G-SDI 3840x2160 4:2:2 YCbCr 50/59.94/60p MP2-FMT-4K Level A, Square Division PRISM MPI2-10-MPX2-10 User Manual...
Page 25
YCbCr 59.94i Base instrument 625i 4:2:2 YCbCr Base instrument 1920x1080 4:2:2 YCbCr 50/59.94/60i Base instrument 1280x720 4:2:2 YCbCr 50/59.94/60p Base instrument 1920x1080 4:2:2 YCbCr 50/59.94/60p Base instrument 3840x2160 4:2:2 YCbCr MP2-FMT-4K 3840x2160 4:2:2 YCbCr 59.94/60p MP2-FMT-4K PRISM MPI2-10-MPX2-10 User Manual...
MP2-QC adds a node-locked license for quality control feature sets, this includes Datalist, CC decode, and ANC Session. • MP2-PROD adds a node-locked license for production feature sets, this includes Stop, False Color, Light meter, HDR/WCG Conversion, CIE, and HDR Measurements. PRISM MPI2-10-MPX2-10 User Manual...
Page 27
Getting Started Product description • MP2-GEN adds a node-locked license for SDI/IP signal generator, this includes IP/ SDI Generator application. • MP2-EXTNDSP adds a node-locked license for enabling extended desktop. PRISM MPI2-10-MPX2-10 User Manual...
MPI-RACK-MM cabinet. MPX2-10 Monitor For the PRISM MPX2-10 monitor, there is an optional 1RU full-rack kit that can be used for either portable or rackmount applications. PRISM MPI2-10-MPX2-10 User Manual...
Page 29
MPX2. A software license is required to enable the extended display feature. For information on how to install and use this extended display, see the PRISM MPX2-DUALDSP Installation and Safety Instructions. PRISM MPI2-10-MPX2-10 User Manual...
Page 30
Caution: To minimize the risk of damage to the instrument, we strongly recommend the power cord be connected to the instrument before the power cord is connected to the AC power source. PRISM MPI2-10-MPX2-10 User Manual...
Page 31
Note: If the PRISM monitor was previously powered off by a power interruption or by removing the power cord from the rear of the instrument, the instrument powers on when power is reapplied. 2. Press the Power/Standby button on the instrument front panel to turn the instrument on. PRISM MPI2-10-MPX2-10 User Manual...
Page 32
2. To completely remove power from the instrument, disconnect the power cord from the instrument. The power cord has a locking mechanism to keep it attached to the instrument. Push the button on the cord housing to release the locking MPI2-10 Power Cord Removal.) mechanism. (See PRISM MPI2-10-MPX2-10 User Manual...
Page 33
Caution: To minimize the risk of damage to the instrument, it is strongly recommended that the power cord be connected to the instrument before the power cord is connected to the AC power source. PRISM MPI2-10-MPX2-10 User Manual...
Page 34
2. To completely remove power from the instrument, disconnect the power cord from the instrument. The power cord has a locking mechanism to keep it attached to the instrument. Push the button on the cord housing to release the locking mechanism. PRISM MPI2-10-MPX2-10 User Manual...
SFP module in an anti-static bag or container. Caution: To prevent possible damage to the PRISM monitor and SFP modules, remove the SFP modules before transporting the PRISM monitor. PRISM MPI2-10-MPX2-10 User Manual...
DHCP, you must manually enter the address for the instrument. To get a fixed address, talk to your LAN administrator. The default IP address of the PRISM monitor is 192.168.1.2. To connect your instrument to a network and access it with a remote PC: PRISM MPI2-10-MPX2-10 User Manual...
Page 37
This may cause a non-deterministic operation, such as IGMP leaves failing or other network connectivity issues. 7. Select the Settings icon ( ) or select the Close icon ( ) to close the Settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 38
MPI2-10 Rear Panel Connectors MPI2-10 Rear Panel Connectors The illustration shows the external connections on the rear panel of the MPI2–10 instrument. A description of each connector is provided. Descriptions of numbered PRISM MPI2-10 rear-panel connectors Item Connector label, number...
Page 39
Getting Started MPI2-10 Rear Panel Connectors Descriptions of numbered PRISM MPI2-10 rear-panel connectors Item Connector label, number symbol, or end Description SDI SFP+ Two optional SFP outputs are for SDI signals (HD-BNC transmitter SFP modules shown). SDI SFP outputs 3 and 4 supports HD/3G signals.
Page 40
Getting Started MPI2-10 Rear Panel Connectors Descriptions of numbered PRISM MPI2-10 rear-panel connectors Item Connector label, number symbol, or end Description Ethernet port is a standard RJ-45 connector for 10/100/1000Base- T Ethernet cable. Audio input port is for future use.
Page 41
Quadlink inputs (4k/8k) the Aux out is link 1 ST2110-20/30 inputs are supported. 10GbE Two optional SFP ports for 10 GbE Ethernet applications. SFP+ 2-1 Two optional SFP outputs for SDI signals. SDI SFP inputs 3 and 4 support HD/3G signals. SFP+ 4-3 PRISM MPI2-10-MPX2-10 User Manual...
Page 42
REF OUT is a pass through of the REF IN. Audio input port is for future use. Audio output (1/8-inch—3.5mm—line out) port is for headphones to listen to the selected audio channel pair. Microphone input port is for future use. PRISM MPI2-10-MPX2-10 User Manual...
Page 43
2. 2 Audio output is an 1/8-inch (3.5mm) diameter port for a headset. It has connections for L and R audio channels. Headset plugs that are an 1/8-inch and have an additional contact for a microphone may not work because the headphone ports are not set for the spacing requirements of a third connection on the plug. PRISM MPI2-10-MPX2-10 User Manual...
Display Elements The illustration shows the key elements of the MPI2-10 internal monitor. Descriptions of the elements are provided in the Descriptions of PRISM Display Elements table. Descriptions of numbered PRISM MPI2-10 Display Elements Item number Description Application tiles 1-4...
Page 45
Getting Started Display Elements Descriptions of numbered PRISM MPI2-10 Display Elements Power / Standby button: Press the button to turn the instrument on or off. To completely remove power from the instrument, remove the power cord. Caution: To prevent data loss, we strongly recommend you first shut down the instrument before disconnecting the power cord.
Page 46
2. Select the Eject icon ( ) to unmount the USB drive from the instrument. Note: If there is no USB drive in the instrument, the Eject USB Drive text and eject icon do not appear. 3. Select the Activity icon ( ) again to close the window. PRISM MPI2-10-MPX2-10 User Manual...
Use Activity DASHBOARD To open the Activity DASHBOARD, select the Activity icon ( To close the Activity DASHBOARD, select the Activity icon ( The alarm group buttons and the alarm titles are color-coded. PRISM MPI2-10-MPX2-10 User Manual...
Page 48
Activity icon. No color change or alarm occurs. The alarms are still logged in the Event Log application and sent to the Syslog server, if enabled. 1. In the status bar, select the Settings icon and then select Alarms. PRISM MPI2-10-MPX2-10 User Manual...
Page 49
– To turn on any specific alarm or alarms, select the title of a group of alarms to see the individual alarms and select the checkbox of any specific alarm. To turn an alarm off, clear the checkbox. PRISM MPI2-10-MPX2-10 User Manual...
Page 50
DASHBOARD, select Acknowledge Group. All the alarms in the selected group and the group button are set to green, unless an alarm is still asserted, in that case the alarm, and the related group button, remains red. PRISM MPI2-10-MPX2-10 User Manual...
Page 51
The Event Log application displays in tile 1. • The errors are filtered by the group currently selected Dashboard Alarm Group. To filter Event Log alarms by group Select an alarm group and select View in tile 1. PRISM MPI2-10-MPX2-10 User Manual...
Page 52
Getting Started Activity Center PRISM MPI2-10-MPX2-10 User Manual...
Methods of Operation This section describes the primary methods of operating the instrument: • MPI2-10 front panel touchscreen • MPX2-10 front panel controls • External touchscreen display • Keyboard and mouse • Remote control using VNC • Remote control using API commands...
Page 54
If you are using a mouse and keyboard you must “click”, “scroll” and use the scroll wheel on the mouse. Select and Control Application to Display Open a different application in a tile: 1. In the Status bar, select the Tiles icon ( ) to open the Application toolbar. PRISM MPI2-10-MPX2-10 User Manual...
Page 55
– Select Clear All to remove all the applications from the Application toolbar. – Select an application that is not marked with a number to add it to an empty space in the application toolbar. PRISM MPI2-10-MPX2-10 User Manual...
Page 56
When a tile is in vertical extended mode, select the quarter- screen icon ( ) in the menu bar to return to quarter-screen mode. • If available, select the Reset icon ( ) to clear or reset the selected display. PRISM MPI2-10-MPX2-10 User Manual...
Position. For instructions on how to use each on-screen tool, see these application sections: • Waveform Application On-screen Tools • Vector Application On-screen Tools • Stop Display Application On-screen Tools • Lightning Application On-screen Tools PRISM MPI2-10-MPX2-10 User Manual...
Page 58
USB ports: There are two USB 3.0 ports for connecting a mouse and keyboard, importing or exporting instrument presets, upgrading the instrument firmware, or saving screen and stream captures. PRISM MPI2-10-MPX2-10 User Manual...
Page 59
To select an input to monitor using the front panel: 1. Press the INPUT button to prepare the instrument for an input selection. 2. Press the SELECTIONS numbered button corresponding to the input you want to monitor. PRISM MPI2-10-MPX2-10 User Manual...
Selects Input 1 or selects Preset 1 from the selected preset group. Select 2 Selects Input 2 or selects Preset 2 from the selected preset group. Select 3 Selects Input 3 or selects Preset 3 from the selected preset group. PRISM MPI2-10-MPX2-10 User Manual...
Page 61
ALT and at the same time and press p three times to make a selection from Preset Group C. 2. Press a number from 1 to 6 to select the input or to select a preset from the selected Preset Group (A-F). PRISM MPI2-10-MPX2-10 User Manual...
Page 62
Connect the output of the external device to one of the USB ports on the PRISM monitor. Note: A monitor with a 1920×1080 capable display works best with the instrument. Note: We recommend you reboot the instrument after connecting an external touchscreen. PRISM MPI2-10-MPX2-10 User Manual...
Page 63
Half-Screen Tiles There are several half screen configurations. Select the Tiles icon ( ) and then select the expand icon ( ) to enlarge the tile vertically. PRISM MPI2-10-MPX2-10 User Manual...
Page 64
Double click or double tap an application to change between quarter-screen tile and full-screen mode. • Select the Tiles icon ( ) to open the application toolbar and select the quarter- screen icon ( ) or the full-screen icon ( ) to change between modes. PRISM MPI2-10-MPX2-10 User Manual...
Page 65
To extend a tile vertically (half-screen mode): • In the application toolbar, select the expand icon ( ) to expand the tile vertically; • Select the compress icon ( ) to collapse the tile to a quarter-screen. PRISM MPI2-10-MPX2-10 User Manual...
Page 66
1. On the PRISM monitor to connect to, select the Settings icon ( ), then select Network, and find the Control IP Port IP address of the instrument. 2. Enter the hostname or IP address in a web browser. The PRISM HOME page displays. PRISM MPI2-10-MPX2-10 User Manual...
Page 67
5. In the Settings menu, use the Scaling Mode drop-down list to select Local Scaling, and then select Apply. 6. In the Control bar, select the Connect icon ( ) to open the Connect menu. 7. In the Settings menu, enter the password for the PRISM, and then select Connect. PRISM MPI2-10-MPX2-10 User Manual...
Page 68
Note: If the correct Scaling Mode to Local Scaling is not selected, you must disconnect using the computer icon at the top of the viewer to gain access the Settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 69
Access the API Documentation 1. Find the PRISM hostname: on the Status bar, under the Telestream logo, next to the Activity icon. Or find the IP address: select Settings, then Network, and under CONTROL IP PORT find the IP Address.
Page 70
Remote Control through API Commands Syslog PRISM supports Syslog protocol (RFC 5424) for alarms and events. Setup Syslog 1. Select the Settings icon ( ) to open the Settings menu. 2. select Network, and select REMOTE ACCESS. PRISM MPI2-10-MPX2-10 User Manual...
• GET and POST are allowed commands. • In these examples, replace "xxx.xxx.xxx.xxx" with the IP address of the PRISM the sending the traps and "yyy.yyy.yyy.yyy" with the IP address of the computer receiving the traps. PRISM MPI2-10-MPX2-10 User Manual...
Page 72
Check the Traps 1. Download the prism_mon.mib from the PRISM web page. a. Open the PRISM instrument home page. b. Enter the PRISM IP address in a browser. c. Select SNMP in the PRISM menu. d. Select Download. PRISM MPI2-10-MPX2-10 User Manual...
Page 73
From the SNMP manager, configure the SNMP agent (PRISM) IP address. b. Load the prism.mib file. c. Open the trap receiver window. d. Confirm the traps are received on the trap receiver window of the MIB browser application on the destination machine (on port 162). PRISM MPI2-10-MPX2-10 User Manual...
Page 74
Methods of Operation Remote Control through API Commands PRISM MPI2-10-MPX2-10 User Manual...
Verify Firmware Upgrade Installed software and hardware versions: Note: Some of the configuration settings require you to enter values. A USB keyboard can be used to enter these values instead of using the touchscreen keypad on the instrument. PRISM MPI2-10-MPX2-10 User Manual...
Page 76
8K is present. If this option is not present, only a single-link input can be configured. Single Link Input Configuration 1. Select the Settings icon ( ) to open the Settings menu. 2. Select Inputs to open the inputs submenu. 3. Select an input and then select SDI. PRISM MPI2-10-MPX2-10 User Manual...
Page 77
Color Gamut settings define the characteristics of the video signal. Note: The preferred setting for all video options is Auto. At this setting the PRISM determines the correct settings from the signal or any metadata (if present). PRISM MPI2-10-MPX2-10 User Manual...
Page 78
3. If a 3-gigabit transport is being used, select the In 3G-SDI drop-down menu and choose: – Auto: The system selects the correct setting based on the information in the incoming signal. – Level A – Level B PRISM MPI2-10-MPX2-10 User Manual...
Page 79
– S-Log3 (Live HDR): The reference OOTF (Optical to Optical Transfer Function) is defined as Sony S-Log3. – Log C: The reference OETF is defined as ARRI Log C. The EOTF is the inverse of OETF. Note: Structure is set to Auto. PRISM MPI2-10-MPX2-10 User Manual...
Page 80
The Audio settings define how the audio is embedded in the SDI input signal. These settings are the same for Single and Quad Link input configurations, except the Quad Link, which is intended for 8K, can have 32 channels, instead of 16, in Program Configuration. 1. Select AUDIO. PRISM MPI2-10-MPX2-10 User Manual...
Page 81
(use of Dolby requires the MP2-DLBY license). Select PCM a. Select Program Configuration. b. Move the Program Configuration slider to On. c. Select Edit. The Edit button disappears and the Default, Clear, and Undo buttons replace it. PRISM MPI2-10-MPX2-10 User Manual...
Page 82
Dolby stream are automatically set up based on the Dolby metadata. Quad Link Input Configuration Note: All four SDI inputs must be connected to a cable on the back of the PRISM to use Quad Link Inputs. PRISM MPI2-10-MPX2-10 User Manual...
Page 83
9. Select Save and close the menu. Note: The 8K format trace applications—Waveform, Stop, Vector, Diamond, and Lightning Displays, enabled by MP2-FMT-8K—are rendered based on the resized 8K signal and are suitable for operational applications such as setting camera exposure PRISM MPI2-10-MPX2-10 User Manual...
Page 84
4. When you are done editing the input, select Save to save your changes. Configure IP Stream Inputs Configure and Monitor ST2022-6 1. Select the Settings icon ( ) to open the Settings menu. 2. Select Inputs to open the Inputs submenu. 3. Select the input from the list. PRISM MPI2-10-MPX2-10 User Manual...
Page 85
Note: Path 2 is only available if ST2022-7 Seamless Switching is enabled. You may have to scroll down to see Path 2. 8. To have the instrument ignore a parameter, select the Unmasked control next to the parameter to change the state to Masked. PRISM MPI2-10-MPX2-10 User Manual...
Page 86
ST2110 menu has a different layout. The settings define the characteristics of the video signal. 1. Choose the Gamma Curve setting from the drop-down menu. The selection characterizes the video signal on each virtual input. PRISM MPI2-10-MPX2-10 User Manual...
Page 87
– Log C: The reference OETF is defined as ARRI Log C. The EOTF is the inverse of OETF. 2. Select Color Gamut settings in the drop-down menu. The selection is the color space of the video signal. a. Rec. 709: Standard for HD. b. Rec. 2020: Standard for 4K. PRISM MPI2-10-MPX2-10 User Manual...
Page 88
Note: Rec. 709 gamut is automatically selected when the SD format is detected or selected in the video signal. Select Audio Settings 1. Select AUDIO. Select PCM Select DOLBY 2. Choose the type of audio. (use of Dolby requires the MP2-DLBY license). PRISM MPI2-10-MPX2-10 User Manual...
Page 89
7.1 assignment, the entire assignment is removed and is not displayed on the Audio application. Default resets the channels to 8 stereo pairs. Clear removes all assignments, including the default. e. Select Save and close the menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 90
If the streams are not identical, error messages are shown in the Status Bar. There is a difference between “port” and “path”. The port is the physical SFP port that is used to input the 10GE signal. Path 1 and Path 2 are the signal paths to be used for seamless switching. PRISM MPI2-10-MPX2-10 User Manual...
Page 91
Select Auto to provide an offset based on the current video standard. g. Select the Gamma Curve and Color Gamut settings for the video stream. See Select Video Settings for information on available Gamma and Gamut selections. 9. Configure the Audio (2110-30) stream tab: PRISM MPI2-10-MPX2-10 User Manual...
Page 92
Set the number of channels in the audio stream between 2 and 16. The number of channels must be an even number (2, 4, 6, etc.). A maximum of 16 audio channels, across all audio streams, are supported. PRISM MPI2-10-MPX2-10 User Manual...
Page 93
Configure and Select Signal Inputs Note: The Bit Depth is already set to 24. g. Select Save. Select Audio Settings 1. Choose the type of audio. Select PCM Select DOLBY (use of Dolby requires the MP2-DLBY license). PRISM MPI2-10-MPX2-10 User Manual...
Page 94
7.1 assignment, the entire assignment is removed and is not displayed on the Audio application. Default resets the channels to 8 stereo pairs. Clear removes all assignments, including the default. e. Select Save and close the menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 95
A guideline for the number of audio channels (assuming 24bit 48khz audio) is approximately 1 to 1.5 Mb/s per channel. See the IP Status Bitrate column. Note: To receive only audio, disable the video stream. PRISM MPI2-10-MPX2-10 User Manual...
Page 96
3. To begin monitoring this input, select the configured input. (See Select Input.) Configure RTP Monitoring general purpose RTP streams allows you to see physical and IP layer measurements, but none of the video, audio, and data payload is decoded and displayed. PRISM MPI2-10-MPX2-10 User Manual...
) from the Status bar at the bottom of the PRISM monitor. 2. Select a configured input from the list in the bottom bar. 3. Select the Home icon ( ) to close the Input selection controls. PRISM MPI2-10-MPX2-10 User Manual...
In the Stop Display Settings menu (select and hold in the application tile and select the Settings icon ( ), in Reference select Display Light or Scene Light depending on your Stop Display application. The trace represents light level in the vertical axis. (See Application.) PRISM MPI2-10-MPX2-10 User Manual...
Page 99
BT. 709 Gamma / Gamut displays. The conversion between input video and linear light is processed in Scene light. Note: Convert to Rec. 709 mode is not supported for SD signals. PRISM MPI2-10-MPX2-10 User Manual...
Page 100
Control selection is for the RJ-45 10/100/1000Mbps Ethernet port. Note: When changes are made to the control port IP address or the label of the instrument and NMOS is being used for configuration, we recommend rebooting the instrument to update the NMOS registration server. PRISM MPI2-10-MPX2-10 User Manual...
Page 101
Control selection is for the RJ-45 10/100/1000Mbps Ethernet port. Note: When changes are made to the control port IP address or the label of the instrument and NMOS is being used for configuration, we recommend rebooting the instrument to update the NMOS registration server. PRISM MPI2-10-MPX2-10 User Manual...
Page 102
3. Set Loudness Display to On. The Loudness Display appears in the right half of the tile. Note: The view varies based on the visibility selection of other auxiliary audio displays including Audio Session or Lissajous Display. PRISM MPI2-10-MPX2-10 User Manual...
Page 103
Loudness display with appropriate color code) LOUDNESS Loud is 0 to -30 and Sets the limits to trigger in-bar status for THRESHOLDS quiet and loud level Quiet 0 to -60 Full Scale Loudness PRISM MPI2-10-MPX2-10 User Manual...
Page 104
Loudness meter measures the levels based on that audio program. Note: Changing the selected Audio program that the Loudness meter measures changes the headphones or speaker pair, and changing the headphones or speaker PRISM MPI2-10-MPX2-10 User Manual...
Page 105
The green letters in the status bar (marked as 3) mark the channels the Loudness display is measuring and where they are in the channel series. Meter Controls • To start and stop the loudness meter: select the play ( ) or stop ( ) button. PRISM MPI2-10-MPX2-10 User Manual...
Page 106
• Mono: Reduces the stream to one channel. The center channel is reduced 3dB, the surround channels are each reduced 3dB, and the LFE is muted. All 6 channels are combined. PRISM MPI2-10-MPX2-10 User Manual...
Page 107
Line and RF are modes of Dynamic Range Control (DRC)—compression—factors that are applied when decoding Dolby content for monitoring or output. Select Downmix Dynamic Rng. Select Line or RF. These DRC compression factors are applied when downmixing to the various Dolby D listening modes. PRISM MPI2-10-MPX2-10 User Manual...
Page 108
SDI OUT loop-through of IP inputs is available for SDI, ST2022-6, and ST2110-20/30/31 Configure the Audio (2110-30) stream tab: (see for audio configuration details) including embedded audio streams. See the SDI Generator applications for instructions on how to configure the SDI Generator signal. PRISM MPI2-10-MPX2-10 User Manual...
Page 109
2. Select Anc Data to open the Anc Data submenu. 3. In Timecode Format select LTC or VITC as needed and close the menu. The selected timecode appears in the bottom border of the Status frame and all alarms and events. PRISM MPI2-10-MPX2-10 User Manual...
Page 110
Note: When a reference selection is made the reference changes in the status bar and in the Timing application. Configure the Analog Reference Settings 1. Select the Settings icon ( ) to open the Settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 111
3. To enable or disable PTP logging to the Event Log, select the control box to change the setting to the state. 4. Select the profile to configure: ST2059, AES67, or General. The default communication mode for each profile is Multicast. PRISM MPI2-10-MPX2-10 User Manual...
Page 112
) to remove the existing domain number. – Select the backspace key to delete characters by backspacing over them. 8. When you are done editing the domain number, select the Enter key. 9. Repeat steps 4 to 6 for each PTP profile. PRISM MPI2-10-MPX2-10 User Manual...
Page 113
(Rename and Save as) you can also recall the existing settings (Recall). When edit mode is off, selecting a preset button only recalls the preset settings assigned to that button, with no confirmation. PRISM MPI2-10-MPX2-10 User Manual...
Page 114
There are 6 presets in each of 6 groups. Use the arrow buttons ( ) or swipe left or right to move to the preset group (A–F). Note: When a preset has no content, <empty> is displayed on the preset button. PRISM MPI2-10-MPX2-10 User Manual...
Page 115
If Rename does not appear, Edit mode is not on. 4. Select Save. Rename Group of Presets 1. Select the Settings icon ( ) to open the Settings menu. 2. Select Presets to open the Presets submenu. PRISM MPI2-10-MPX2-10 User Manual...
Page 116
(this includes naming an <empty> preset). 5. Select the name box of the preset you want to name or rename. An on-screen keyboard displays. Note: You can also rename a preset group by selecting a group name box. PRISM MPI2-10-MPX2-10 User Manual...
Page 117
) and then select the Eject icon ( ) to unmount the USB drive from the instrument. 7. Remove the USB drive with the presets from the PRISM. 8. To upload the presets to another PRISM, go to Upload the Presets from a USB Memory Drive. PRISM MPI2-10-MPX2-10 User Manual...
Page 118
9. Repeat steps 1 to 9 for any other PRISMs to copy the presets. Download a File of Presets Locally Caution: This overwrites all existing presets in the PRISM. 1. Go to the PRISM HOME page for the PRISM with the presets. 2. Select Preset Download and Upload. PRISM MPI2-10-MPX2-10 User Manual...
Page 119
1. With the presets file downloaded locally, go to the PRISM HOME page for the PRISM to upload the presets. 2. Select Preset Download and Upload. 3. Select Upload to PRISM and then select Browse. A file explorer window displays. PRISM MPI2-10-MPX2-10 User Manual...
Page 120
Note: The preset still has all of the input settings, they are just not applied when the Recalled Saved Inputs setting is off. 1. Select the Settings icon ( ) to open the Settings menu. 2. Select Presets to open the Presets submenu. PRISM MPI2-10-MPX2-10 User Manual...
Page 121
4. Select OK to confirm the selection and to reset the instrument settings to the factory defaults. The next table lists some of the settings that are reset. Note: Custom presets are still available after recalling the factory preset. PRISM MPI2-10-MPX2-10 User Manual...
Page 122
SDI-In 4 Input type SDI 4 Input connector SDI-SFP 1 input configuration Input name SDI-SFP 1 Input type Input connector None selected SDI-SFP 2 input configuration Input name SDI-SFP 2 Input type Input connector None selected PRISM MPI2-10-MPX2-10 User Manual...
Many alarms also have user-definable thresholds to set specific limits for the alarm. Configure Alarms Open the Alarms Select Settings, and then select Alarms. • To turn on the group of alarms, select the checkbox under Dashboard for the group. PRISM MPI2-10-MPX2-10 User Manual...
Page 124
2. Select the UI checkbox of the alarm to monitor. Note: If the alarm shows information in the Threshold column (not N/A), you can change the setting. Change an Alarm Threshold 1. Select the alarm title. The Thresholds option displays. PRISM MPI2-10-MPX2-10 User Manual...
Page 125
S334-2 CEA 608 Closed Captions Lost: No valid S334-2 CEA 608 closed caption packets were found in the ancillary data space. • SMPTE 2031 Teletext Lost: No valid SMPTE 2031 ANC data packets were found in the ancillary data space. PRISM MPI2-10-MPX2-10 User Manual...
Page 126
RTP Timestamp Error: RTP timestamp is corrupt or is not updating at the correct frequency for the specified media type. • S2022/7 Path Missing: One of the datapaths in a S2022.7 seamless switching configuration is missing. PRISM MPI2-10-MPX2-10 User Manual...
Page 127
Using a known test signal, the checksum status can be used to verify integrity through a system. An error means that the Y Channel embedded checksum value does not match the calculated checksum value, which indicates that a transmission error has occurred. PRISM MPI2-10-MPX2-10 User Manual...
Page 128
Format setting (which is set to a specific line rate, rather than being set to Auto) in the SDI Input submenu of the Configure menu or the setting specified in the SMPTE352 payload (if present). • Video Signal Lost: Video input signal has not been detected at the selected input. PRISM MPI2-10-MPX2-10 User Manual...
Page 129
) to decrease intensity of the trace or graticules. Note: The intensity can also be adjusted by selecting anywhere in the slider or selecting and dragging the slider circle left or right. 4. Close the Display settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 130
5. Select the Time Zone drop-down list to select the correct time zone offset for your location. The selection names include the time offset from UTC and major cities in those time zones. 6. Select TIME AND DATE to open the submenu. PRISM MPI2-10-MPX2-10 User Manual...
Page 131
9. The message box displays and asks whether you want to reboot the instrument now or later to implement the time and date changes. Select one of the options: – Select OK to reboot the instrument immediately and implement the changes. – Select Cancel to end the operation. PRISM MPI2-10-MPX2-10 User Manual...
Upgrade Instrument Firmware Before You Begin Telestream releases updates to product firmware to add new features or to fix reported problems. Check the Telestream website regularly for new firmware releases. You do not need to perform a firmware upgrade if your instrument has the latest version of firmware already installed.
Caution: Removing the USB device or powering off the instrument before the upgrade is complete can cause upgrade failure. To prevent upgrade failure, wait until the Installation Complete message box is displayed before performing these actions. PRISM MPI2-10-MPX2-10 User Manual...
Page 134
) to open the Settings menu. 2. Select Utilities to open the Utilities submenu. 3. Select Version to open the Version display. 4. Verify the displayed firmware version number matches the version of the firmware upgrade package you installed. PRISM MPI2-10-MPX2-10 User Manual...
Page 135
1. Select the Settings icon ( ) to open the Settings menu. 2. Select Utilities to open the Utilities submenu. 3. Select Options to open the Options display. 4. Verify that the displayed option(s) match the option(s) you installed. PRISM MPI2-10-MPX2-10 User Manual...
You can download a zip file containing a service report of instrument diagnostics using the Settings menu. If your instrument needs service, you may be asked to provide the diagnostics file to Telestream to aid in troubleshooting problems with the instrument. 1. Select the Settings icon ( ) to open the Settings menu.
Page 137
Functions Functions This section describes additional functions available on the PRISM monitor. • Headphone and speaker volume, balance, and source adjustment • Stream Capture PRISM MPI2-10-MPX2-10 User Manual...
Page 138
If the input is SDI, select AUDIO. b. If the input is 2022, select AUDIO. 4. Check the Program Configuration is on and configured for the input. 5. Select Save. 6. Select the Volume icon ( ). The downmix button is visible. PRISM MPI2-10-MPX2-10 User Manual...
Page 139
(visible when only Program Configuration is on and set) Audio source: Audio channels to monitor WARNING: To avoid damaging your hearing, always turn the volume down to the minimum before you put on headphones and then turn the volume up slowly. PRISM MPI2-10-MPX2-10 User Manual...
1. From the Settings menu, select Capture. Note: If the Capture configuration menu does not appear under the Settings menu, option MP2-IP-MEAS is not installed, but screen captures are still functional. 2. Select Screenshot under SELECT CAPTURE TYPE. PRISM MPI2-10-MPX2-10 User Manual...
Page 141
3. On the application bar select the Capture icon ( A screenshot is saved to the device. Note: The screenshot API is for Telestream internal use only. If you need to remotely save a screenshot to a flash drive, use VNC.
Page 142
9. A screenshot and pcap file are saved to the attached USB device. The name of the file and the path to the file on the USB device are displayed after the save is complete. 10. Select OK to close the menu. PRISM MPI2-10-MPX2-10 User Manual...
To see the All Application menu, select the Tiles icon ( ) and then select the All Application menu icon ( Note: Some displays require a specific option to be installed. For example, the IP Graph display is not accessible unless you have Option MP2-IP-MEAS. PRISM MPI2-10-MPX2-10 User Manual...
Page 144
Event Log Application Use this to view a log of detected errors. IP Status Application Use this to see an overview of the monitored IP stream and see the status of each program in the stream. PRISM MPI2-10-MPX2-10 User Manual...
Page 145
2022-6 and 2110-20 video signals and PTP signals. Stream Timing Application Use this application to view the timing of the ST2110 video, audio, and data as it was received relative to the embedded RTP time stamps. PRISM MPI2-10-MPX2-10 User Manual...
Page 146
Application: Use this to see the digital data contained in SDI frames to perform in-depth monitoring and analysis of incoming SDI data ANC (Ancillary Data) Session Application: Use this to monitor all ancillary data present in a signal. PRISM MPI2-10-MPX2-10 User Manual...
Page 147
Application: Use this to reliably detect invalid colors. Lightning Application: Use this to show luma and chroma amplitudes and verify component timing using a color-bar signal. Application: Use this to check the chromaticity of the video against various color gamut limits. PRISM MPI2-10-MPX2-10 User Manual...
Indicates the signal parameter was in an error condition but the error has now cleared. Indicates the signal parameter is currently in an error condition. Indicates the signal parameter is not being monitored for an error condition. PRISM MPI2-10-MPX2-10 User Manual...
Page 149
) is in the settings menu header. Select this icon to move the settings left of right. Open the Waveform Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Waveform application. PRISM MPI2-10-MPX2-10 User Manual...
RGB: Displays the input as Red (R), Green (G), and Blue (B) components. • YPbPr: Displays the input as Luma (Y) and color difference (Pb, Pr) components. • YRGB: Displays the input as Luma (Y), Red (R), Green (G), and Blue (B) components. PRISM MPI2-10-MPX2-10 User Manual...
Turn this on to remove the trace of vertical / horizontal blanking data. Gain Select from the options to set to x1, x2, x5, or x10. Magnification Select from the options to set to the magnification to Best View, x1, x10, x20, x25, or x50. PRISM MPI2-10-MPX2-10 User Manual...
Page 152
Gain Increase or decrease the magnification of the trace display. 1. Select the Gain button to expand the available magnification options drop down list. 2. Select one of the preset magnifications for fixed gain. PRISM MPI2-10-MPX2-10 User Manual...
For fine increments, select the plus or minus buttons to adjust the gain. • The scroll wheel on a mouse adjusts the variable gain. • Use the Reset icon ( ) to remove variable gain adjustments. Horizontal and Vertical Position Adjust the position of the trace display. PRISM MPI2-10-MPX2-10 User Manual...
Page 154
– Use a mouse scroll wheel to move the trace. – For incremental adjustments, use the plus or minus buttons to adjust the horizontal position. – Select the Reset icon ( ) to reset the horizontal position adjustment to default. PRISM MPI2-10-MPX2-10 User Manual...
Page 155
– Use a mouse scroll wheel to move the trace. – For incremental adjustments, use the plus or minus button to adjust the – vertical position. – Use the Reset icon ( ) to reset the vertical position adjustment to default. PRISM MPI2-10-MPX2-10 User Manual...
Page 156
Note: The time and voltage cursor positions are retained if the trace is moved horizontally or vertically, or if the gain is adjusted. PRISM MPI2-10-MPX2-10 User Manual...
Page 157
– Select YPbPr to display any combination of the Y, Pb, and Pr waveforms. – Select RGB to display any combination of the R, G, and B waveforms. – Select YRGB to display any combination of the Y, R, G, and B waveforms. PRISM MPI2-10-MPX2-10 User Manual...
– When the Parade display style is active you can select from 1 line or 1 field. – When Overlay display style is active you can select from 1 line, 2 line, 1 field, or 2 field. PRISM MPI2-10-MPX2-10 User Manual...
Page 159
• the Line number is the selected line, including vertical blanking, • if the signal is interlaced format, the Fields readout is below the Line number, • the Pic Line number is next to Line; the Pic Line is the active line number, or numbers, selected. PRISM MPI2-10-MPX2-10 User Manual...
Page 160
Note: In link, the letters (A, B, C, or D) refers to the physical links. The numbers (1, 2, 3, or 4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
2. Select the tile with the Waveform application. 3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Page 162
You can change the gain, horizontal position, vertical position, and bar targets without opening the Vector settings menu through on-screen tools. Select the tool/button for the setting you want to adjust. All the available tools are located on the top and bottom PRISM MPI2-10-MPX2-10 User Manual...
Page 163
Gain Increase or decrease the magnification of the trace display. 1. Select Gain to expand the available magnification options drop down list. 2. Select one of the preset magnifications for fixed gain. PRISM MPI2-10-MPX2-10 User Manual...
Page 164
• For fine increments, select the plus or minus button to adjust gain. The scroll wheel on a mouse adjusts the variable gain. • Select the Reset icon ( ) to remove variable gain adjustments. PRISM MPI2-10-MPX2-10 User Manual...
Page 165
– Use a mouse scroll wheel to move the trace. – For incremental adjustments, use the plus or minus button to adjust the horizontal position. – Select the Reset icon ( ) to reset the horizontal position adjustment to default. PRISM MPI2-10-MPX2-10 User Manual...
Page 166
Note: The same line is selected in any trace display if line select is enabled. To turn on Line Select (and choose the line to monitor): 1. Select the Lines button at the bottom of the tile. PRISM MPI2-10-MPX2-10 User Manual...
Page 167
• the Line number is the selected line, including vertical blanking, • if the signal is interlaced format, the Fields readout is below the Line number, • the Pic Line number is next to Line; the Pic Line is the active line number, or numbers, selected. PRISM MPI2-10-MPX2-10 User Manual...
Page 168
Note: In link, the letters (A-D) refers to the physical links. The numbers (1-4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
) is in the settings menu header. Select this icon to move the settings left of right. Open the Waveform Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Waveform application. 3. Select the Tile icon ( PRISM MPI2-10-MPX2-10 User Manual...
The Lissajous display is a visual of a user-specified pairing of channel inputs with a channel correlation meter. To turn on the Lissajous display, in the Audio Settings, in Aux Display, select Lissajous. The Lissajous display appears to the right of the Audio bars. PRISM MPI2-10-MPX2-10 User Manual...
The meter bars are displayed below the Test level in a green color. • Between the Test level and the Peak Program level, the meter bars are displayed in yellow. • Above the Peak Program level, the meter bars are displayed in red. PRISM MPI2-10-MPX2-10 User Manual...
Page 172
MUTE: The number of consecutive all-zero samples equals or exceeds the # Samples for Mute setting. PRISM MPI2-10-MPX2-10 User Manual...
2. Set the surround Dominance Indicator to ON or OFF as needed. 3. Set the Loudness Filter to the type preferred. A-weighting biases the response toward that of human hearing. 4. Use the level bar display to monitor level control. PRISM MPI2-10-MPX2-10 User Manual...
Page 174
Surround Sound display, indicates the location of potential phantom sound sources formed by adjacent channels. The white tic marks on these moving bar indicators show the phantom source locations. The bar length indicates the correlation between the adjacent channels. A short-to-medium length green bar PRISM MPI2-10-MPX2-10 User Manual...
Uncorrelated signals (example 1): This shows the same level in the L, C, R, Ls, and Rs channels Sine wave test tone (example 2): This has the same level in the L,C, R, Ls, and Rs channels. All signals are in phase, creating phantom sources between the adjacent channels. PRISM MPI2-10-MPX2-10 User Manual...
Page 176
It is a part of the Audio App in PRISM and it is configured in the main Settings menu. The Audio Loudness meter allows you to view audio loudness values associated with audio loudness measurements. This instrument maintains a running audio loudness session. PRISM MPI2-10-MPX2-10 User Manual...
Page 177
To turn on the audio loudness values, in the Audio Settings, in Aux Display, set Loudness Meter to On. The Loudness Meter appears to the right of the Audio bars. Audio Session It consists of several performance parameters for an overview of the audio input signal. PRISM MPI2-10-MPX2-10 User Manual...
Page 178
It allows you to select any one channel of a downmixed 5.1 or 7.1 stream and listen to it alone. This helps to isolate and identify an unexpected noise in a stream. To use Downmix Solo: 1. Set up a 5.1 or 7.1 stream for downmix. PRISM MPI2-10-MPX2-10 User Manual...
Page 179
6. Select a channel to isolate in the Audio application: – To change the solo channel, select another channel in the downmixed stream. – To return to the entire downmixed stream (without turning Downmix Solo off ), select the highlighted channel again. PRISM MPI2-10-MPX2-10 User Manual...
) is in the settings menu header. Select this icon to move the settings left of right. Open the Picture Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Waveform application. PRISM MPI2-10-MPX2-10 User Manual...
Set them as required. The source ID is limited to a maximum 15 ASCII characters. To enable the Source ID Overlay, open the Picture Setting menu and then set the Source ID Overlay to On. Select and drag the overlay to the correct location on the application. PRISM MPI2-10-MPX2-10 User Manual...
Page 182
% Area: A fixed false-color palette is applied to regions of the picture: Brightest, Area, and Darkest. • Luminance: Create a user defined false color palette based on luminance values. Luminance False Color.) (See Closed Caption/Subtitles Tools for closed captions/subtitles decode and display. PRISM MPI2-10-MPX2-10 User Manual...
Page 183
Note: Each Picture tile can select a closed caption format and a service. Picture Measurements An overlay containing a selectable set of picture measurements can be shown on the picture application, including HDR measurements. PRISM MPI2-10-MPX2-10 User Manual...
Page 184
See the Measurement Threshold menu for setting these thresholds. These measurements are available. Settable thresholds are indicated by x: • Total percent HDR area the Nits threshold for this measurement can be set. • X percent area, in nits PRISM MPI2-10-MPX2-10 User Manual...
Page 185
Lightmeter cursor in the Picture application. Proper operation requires selecting the appropriate Gamma Curve in the input settings. (See Choose the Gamma Curve option from the drop-down menu. The selection characterizes the video signal on each virtual input.) PRISM MPI2-10-MPX2-10 User Manual...
Page 186
3. Select the number of needed Lightmeter cursors (up to 5). Note: The Lightmeters all use the same method for measurements. The color of the Lightmeter cursor is just for identification. The Lightmeters are visible in the Picture application. PRISM MPI2-10-MPX2-10 User Manual...
Page 187
Default: By default, the false colors are defined according to the BT.2408 specifications if a display-based gamma curve is set for the input, or a typical camera specification if a camera-based gamma curve is set for the input. If changes are made to the false color PRISM MPI2-10-MPX2-10 User Manual...
Page 188
Enter to close. The false colors in the Picture and the Meter Overlay are updated as adjustments are made. If the range for one color overlaps the range for another color, the color that is higher in the table overlaps the colors lower in the table. PRISM MPI2-10-MPX2-10 User Manual...
Page 189
HH:MM:SS. DD is the number of days, HH is the number of hours, MM is the number of minutes, and SS is the number of seconds. • Reset icon: Select the Reset icon ( ) to restart the monitoring session. PRISM MPI2-10-MPX2-10 User Manual...
Page 190
For information about the errors in this section, view the help file in your instrument: When the Video Session is active in a tile, select the HELP button. • Status: Shows the status of the associated error as either OK, Invalid, Missing, or Error. PRISM MPI2-10-MPX2-10 User Manual...
Page 191
Use the BIT LEVEL tab display to view if there are any stuck bits in the input signal. Elements of the BIT LEVEL Tab Display • Y Stuck Bits: Displays which of the luminance video bits are stuck. If the readout is " ---", then none of the bits are stuck. PRISM MPI2-10-MPX2-10 User Manual...
Page 192
C Anc Checksum Error: Displays the status of the C channel ancillary embedded checksum value that was calculated for the video signal. Using a known test signal, the checksum status can be used to verify integrity through a system. An error PRISM MPI2-10-MPX2-10 User Manual...
Page 193
Err Secs: The number of seconds, since the last reset, that contained at least one error. • % Err Fields: Shows a calculated number showing the percentage of all fields since the last reset that contained at least one error. PRISM MPI2-10-MPX2-10 User Manual...
Open the Picture Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Picture application. 3. Select the Tile icon ( Measurement Mode Use the menu to change the measurement mode between options. PRISM MPI2-10-MPX2-10 User Manual...
Page 195
– Horizontal Offset: The timing difference between the reference and input signal, expressed as the line time which remains from the total delay after the vertical offset has been removed. This has a range of ± 1/2 a line for the input format. PRISM MPI2-10-MPX2-10 User Manual...
Page 196
If a properly timed SDI signal is applied to a gateway, then the timing measurement on the resulting IP flow displays the combined latency in the gateway and the network. Each ST2110 stream also carries time stamps. The Stream Timing application allows for the analysis of these time stamps. PRISM MPI2-10-MPX2-10 User Manual...
Page 197
Log is tracking new events. The newest events appear at the bottom of the Event Log. Scroll in the Event Log list or select-and-hold the slider and move to a previous event. Tracking new events is automatically paused. PRISM MPI2-10-MPX2-10 User Manual...
Page 198
) is in the settings menu header. Select this icon to move the settings left of right. Open the Event Log Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Event Log application. PRISM MPI2-10-MPX2-10 User Manual...
Page 199
To open the group filter, select the Group heading. Note: If there is any other filter applied (Event, Date, Time—marked by a full filter icon ) next to a heading), it limits what any other filter shows. For example, if no events PRISM MPI2-10-MPX2-10 User Manual...
Page 200
To open the event filter, select the Event heading. Note: If there is any other filter applied (Group, Date, Time—marked by a filled filter icon-- --next to a tile), it limits what any other filter shows. For example, if no groups PRISM MPI2-10-MPX2-10 User Manual...
Page 201
– If any group is not selected (a filter is applied), the filter icon is full. • To select or unselect any single checkbox, select anywhere in the line of the title. • To scroll through the list of events, select the arrows at the top and bottom of the menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 202
– Select in the Date From field and either enter a date (yyyy-mm-dd) with a keyboard, or select the date using the on-screen Adjust Date tool. – Repeat selecting a date in the Date To field, if needed. PRISM MPI2-10-MPX2-10 User Manual...
Page 203
– Select in the Time From field and either enter a time (hh-mm-ss) with a keyboard—or select the time using the on-screen Adjust Time tool—and select Enter. – Repeat selecting a time in the Time To field, if needed. Note: Other filters (duration, date) are not required. PRISM MPI2-10-MPX2-10 User Manual...
Page 204
– If no date, duration, or time filter is entered (no filter), the filter icon is open. – If a date, duration, or time filter is entered (a filter is applied), the filter icon is filled. PRISM MPI2-10-MPX2-10 User Manual...
Page 205
1. Select the Tiles icon ( 2. Select the tile with the IP Status application. 3. Select the Tile icon ( In the IP Status settings menu select Both to monitor both Port 1 and Port 2. PRISM MPI2-10-MPX2-10 User Manual...
Page 206
HH is the number of hours, MM is the number of minutes, and SS is the number of seconds. • Clear icon: Select the Clear icon ( ) to reset the status session. After a reset, items appear in the list in the order they are detected. PRISM MPI2-10-MPX2-10 User Manual...
Page 207
• SEQ ERR: Shows the number of sequencing errors in the monitored stream. • RTP CLK: Shows the RTP clock of the monitored stream. • RTP MARK: Shows the RTP marker frequency of the monitored stream. PRISM MPI2-10-MPX2-10 User Manual...
Page 208
Path 1 and Path 2 to use a single Port. Elements of the IP Session Header The elements at the top of the IP Session application are shared between several tabs: PRISM MPI2-10-MPX2-10 User Manual...
Page 209
) to stop the session. When the session is stopped (no error data collection and no display updates), select the Play icon ( ) to restart the session. LAYER 1/2 Tab Display Select the LAYER 1/2 tab to view the status of these signal parameters. PRISM MPI2-10-MPX2-10 User Manual...
Page 210
ERR SECS: An errored second is an interval of a second during which any error has occurred, regardless of whether that error was a single bit error, or a complete loss of communication for that entire second. PRISM MPI2-10-MPX2-10 User Manual...
Page 211
– Destination Addr: Lists the destination addresses for the L3 IP layer. • L4 UDP: – Source Port: Lists the source ports for the L4 UDP layer. – Destination Port: Lists the destination ports for the L4 UDP layer. PRISM MPI2-10-MPX2-10 User Manual...
Page 212
Note: For 2022-6 signals, the status indicator turns red if the Time Stamp field does not change at the rate of 27 MHz on every packet. For 2110 signals, the status indicator PRISM MPI2-10-MPX2-10 User Manual...
Page 213
Reserved (n): When n is a number other than 0 or 1, indicates that a reserved value is being used. Frame Count: The Video Frame counter changes in value for the next RTP sequence and rolls over after 255. PRISM MPI2-10-MPX2-10 User Manual...
Page 214
2.970 GHz signal (0x00), Unknown/Unspecified frame rate 2.970/1.001 GHz signal (0x01), Unknown/Unspecified frame rate 1.485 GHz signal (0x02), Unknown/Unspecified frame rate 1.485/1.001 GHz signal (0x03), Unknown/ Unspecified frame rate 0.270 GHz signal (0x04), Reserved (0x04-0x0f ), 60 (0x10), PRISM MPI2-10-MPX2-10 User Manual...
Page 215
Select the AUDIO tab to view the status of the audio in the stream. Set the Audio Stream 1. Select the Tile icon ( ) in the application toolbar to open the IP Session Settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 216
The value is set to 0 for all other Media Datagrams and should be 0 for most packets. – Payload type: Identifies the type of RTP payload. – Sequence number: RTP sequence counter that changes by one for each RTP Media Datagram that is sent. PRISM MPI2-10-MPX2-10 User Manual...
Page 217
– Destination Addr: Lists the destination addresses for the L3 IP layer. • L4 UDP – Source Port: Lists the source ports for the L4 UDP layer. – Destination Port: Lists the destination ports for the L4 UDP layer. PRISM MPI2-10-MPX2-10 User Manual...
Page 218
90 kHz on every packet. – SSRC: Identifies the Synchronization Source ID and is set in compliance with RFC3550. This identifier is chosen randomly and should be unique so no two SSRC have the same value with an RTP session. PRISM MPI2-10-MPX2-10 User Manual...
Page 219
Ethernet connection. • Profile: Shows the profile being used: General, AES67, or ST2059. This is the user- selected profile set on the Settings > Reference configuration menu page and is not detected on the input. PRISM MPI2-10-MPX2-10 User Manual...
Page 220
SMPTE mixed mode without negotiation. In later firmware releases, other communication modes will be supported. For some profiles, such as some telecom profiles and the SMPTE ST2059 profile, a mixture of Multicast and Unicast are allowed. On the SMPTE profile unique mixed PRISM MPI2-10-MPX2-10 User Manual...
Page 221
Note: For SMPTE mixed mode, PRISM sets the delay Request rate equal to the Sync message rate received from the master. – Delay Response: Displays the rate of the Delay Response messages received from the master. Each Delay Request should have a matching Delay Response, so PRISM MPI2-10-MPX2-10 User Manual...
Page 222
Registration Server List: Select the Registration Server List tab to view the Host Name, IP Address, and Priority of the servers registered to this instrument using NMOS. Note: Select Return to go back to the main IP Session NMOS menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 223
Last SDP File Received: Select the Last SDP File Received tab view video, audio, and data SDP files. These files contain information used to configure input signals from the registration servers. Note: Select Return to go back to the main IP Session NMOS menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 224
3. Select the Tile icon ( Trend Interval Select the drop-down menu to change the interval to 1 minute, 5 minute, 10 minute, 30 minute, 1 hour, 3 hour, 6 hour, 12 hour, or 1 day. PRISM MPI2-10-MPX2-10 User Manual...
Page 225
Stream type: Select this button to change to Video, Audio, or Data stream IP graphs. • Path: Select the Path button to change to Path 1, Path 2, or both Path 1 and 2. • Res: Lists the resolution of the graph display at the bottom right of the graph. PRISM MPI2-10-MPX2-10 User Manual...
Page 226
Free Min: Shows the minimum available bandwidth that is not in use. • Free Max: Shows the maximum available bandwidth that is not in use. Session Bit Rate Graph The Session Bit Rate graph shows the data rate of the currently selected input stream. PRISM MPI2-10-MPX2-10 User Manual...
Page 227
The number of reconstructed path errors also applies to situations when ST2022-7 is turned off. Packets could pass uncorrected if they are dropped or if reordering results in an input buffer overflow. PRISM MPI2-10-MPX2-10 User Manual...
Page 228
Path 1 minus Path 2. A positive value means that Path 2 is lagging Path 1. Note: The measurement range of Path1-Path2 differential measurement is 1 second. Note: The Path1-Path2 differential measurement is not supported for ST2110-30. PRISM MPI2-10-MPX2-10 User Manual...
Page 229
The VRX Buffer section provides a trend graph for this type of modeling. Both trend graphs help engineers properly setup the packet delivery timing in the RTP packet sender. PRISM MPI2-10-MPX2-10 User Manual...
Page 230
3. Select the Tile icon ( Y-Axis Scaling: Select Linear or Log for the y-axis scaling. Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Page 231
Std Dev: Shows the standard deviation of the packet interval time. Use the pinch and pinch-out gestures on the graph to zoom in and out. After using the zoom-in gesture, the text "Zoomed" is displayed on the x axis of the graph. PRISM MPI2-10-MPX2-10 User Manual...
Page 232
Note: When using a keyboard and mouse, click and drag to zoom in. Select the Reset button ( ) to set the graph back to autoscale mode. Note: The Reset button is only displayed after zooming in. PRISM MPI2-10-MPX2-10 User Manual...
Page 233
Elements of the PTP Graph The PTP graphs are useful in observing the effects of traffic on the PTP messages and the operation of transparent and boundary clocks. Non-PTP aware switches may have PRISM MPI2-10-MPX2-10 User Manual...
Page 234
As the PTP is locking, the slave clock adjusts, so the trace may have large swings or jumps. As the system locks, the master-to-slave and slave-to-master delays PRISM MPI2-10-MPX2-10 User Manual...
Page 235
The PTP phase lag is the phase error of the phase-locked loop, which controls the slave clock. The phase lag is a reasonable estimate of how well the PTP slave is locked to the master; however, the phase lag does not indicate if there is any asymmetry in the network. PRISM MPI2-10-MPX2-10 User Manual...
Page 236
3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. Trend Interval: Select the drop-down menu to choose the interval. PRISM MPI2-10-MPX2-10 User Manual...
Page 237
Path 2. Stream type tabs: Select these tabs to switch between the stream offset graphs: VIDEO AUDIO Note: The number in the AUDIO tab is the selected stream. DATA Res: Lists the resolution of the graph display. PRISM MPI2-10-MPX2-10 User Manual...
Page 238
PTP according to the time stamps. This is a time-based graph of the Timing Display application. Vid-RTP Offset The graph shows timing of the video stream as it was received relative to the embedded RTP time stamps. PRISM MPI2-10-MPX2-10 User Manual...
Page 239
Note: The audio measurements are logged to a single graph. If you change the audio stream being monitored, we recommend resetting the graph to only see the measurements relative to the current stream. To change the monitored stream, open the Tile Settings menu. PRISM MPI2-10-MPX2-10 User Manual...
Page 240
Data Offset Tab Data-Vid Offset The graph shows the relative delay between data and video streams. This value indicates the amount of delay that must be applied to realign the signals according to the time stamps. PRISM MPI2-10-MPX2-10 User Manual...
Page 241
Application Information Stream Timing Application Data-RTP Offset The graph shows timing of the data stream as it was received relative to the embedded RTP time stamps. PRISM MPI2-10-MPX2-10 User Manual...
Page 242
2. Select the tile with the IP Generator application. 3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Page 243
– 720p 60: available for format HD 720 only – 1080i 50: available for format HD 1080 – 1080p 50: available for format 3G LevelA – 1080i 59.94: available for format HD 1080 – 1080p 59.94: available for format 3G LevelA PRISM MPI2-10-MPX2-10 User Manual...
Page 244
VIDEO PORT SFP+ 1 and SFP+ 2 Video port SFP 1 and SFP 2 can be Enabled or Disabled. When enabled, each port is configured using the Destination Address and Destination Port. PRISM MPI2-10-MPX2-10 User Manual...
Page 245
Sample Depth (all Channels): This setting lists the sample depth of all channels. The next table shows the audio test signal values in each channel when both Frequency and Amplitude, in the settings menu, are set to Multiple. PRISM MPI2-10-MPX2-10 User Manual...
Page 246
Note: Each generated IP stream must have a unique RTP Payload Type number. AUDIO PORT SFP+ 1 and SFP+ 2 Audio port SFP 1 and SFP 2 can be Enabled or Disabled. When enabled, each port is configured using the Destination Address and Destination Port. PRISM MPI2-10-MPX2-10 User Manual...
Page 247
Elements of the IP Generator Display Three display tabs provide information regarding the generated IP streams. Select one of the tabs to view the associated display. TRANSPORT Tab Display Select TRANSPORT to view the status of the generated IP streams. PRISM MPI2-10-MPX2-10 User Manual...
Page 248
Application Information IP Generator Application VIDEO Tab Display Select VIDEO to view the settings of the generated video stream. AUDIO Tab Display Select AUDIO to view the settings of the generated audio stream. PRISM MPI2-10-MPX2-10 User Manual...
Page 249
Note: The SDI Generator application requires Option MP2-GEN. SDI Generator Settings Menu Open the SDI Generator Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the SDI Generator application. 3. Select the Tile icon ( PRISM MPI2-10-MPX2-10 User Manual...
Page 250
720p 59.94: available for format HD 720 only • 720p 60: available for format HD 720 only • 1080i 50: available for format HD 1080 only • 1080p 50: available for format 3G LevelA or 3G LevelB only PRISM MPI2-10-MPX2-10 User Manual...
Page 251
Select to enable or disable horizontal motion of the color bars. Elements of the SDI Generator Display Select the VIDEO tab to view the settings of the generated SDI stream. Note: SDI audio and data generation is not yet available. PRISM MPI2-10-MPX2-10 User Manual...
Run/Pause button in the application banner. While the application is running or paused, you can scroll through the SDI data using a touchscreen or mouse. Scroll vertically by touch-swipe or use the mouse scroll wheel. PRISM MPI2-10-MPX2-10 User Manual...
Page 253
SAV button and then select the fast forward button. Run/Pause control: Select this button to run and pause the SDI frame updates. Pausing allows you to examine a single SDI frame in more detail. PRISM MPI2-10-MPX2-10 User Manual...
Page 254
SDID: Secondary data identifier of the packet; permissible values range from 0 through 0xFF (255) inclusive. LINK: Virtual link or stream the ANC data is present on. LOCATION(S): Transport line number the ANC data is present on. OCCURRENCE: Shows the last time a packet occurred. PRISM MPI2-10-MPX2-10 User Manual...
Page 255
When the pause button ( ) is visible, it means data collecting-monitoring is running. Multiple Operation codes and the titles Op code Op title 0x0101 splice_request_data() 0x0102 splice_null_request_data() 0x0104 time_signal_request_data() 0x0109 insert_DTMF_descriptor_request_data() 0x010A insert_avail_descriptor_request_data() 0x010B insert_segmentation_descriptor_request_data() 0x010C proprietary_command_request_data() PRISM MPI2-10-MPX2-10 User Manual...
Page 256
The number in the left column of the SCTE104 page is the number of operations received. Select the Plus icon ( ) or Minus icon ( ) to scroll through the list of up to the last 10 received messages. PRISM MPI2-10-MPX2-10 User Manual...
Page 257
2. Select the tile with the Jitter Display application. 3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Use the drop-down menu to select 1 Line, 2 Line, 1 Field, or 2 Field. Jitter Meter Select this to hide or show the jitter meter at the top of the display. Note: Eye measurements are displayed when the Jitter Display application is in full screen mode. PRISM MPI2-10-MPX2-10 User Manual...
Page 259
2. Select the tile with the Eye Display application. 3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Page 260
Use the drop-down menu to select 3 Eye, 20 Eye, 1 Field, or 2 Field. Jitter Meter Use the button to hide or show the jitter meter above the eye pattern. Note: Eye measurements are displayed when the Eye Display application is in full screen mode. PRISM MPI2-10-MPX2-10 User Manual...
Page 261
Configure and Select Signal Inputs section. Stop Display Settings Menu Open the Stop Display Settings menu: 1. Select the Tiles icon ( 2. Select the tile with the Stop Display application. 3. Select the Tile icon ( PRISM MPI2-10-MPX2-10 User Manual...
Page 262
Y: Displays the input as Luma (Y) components. • RGB: Displays the input as Red (R), Green (G), and Blue (B) components. • YRGB: Displays the input as Luma (Y), Red (R), Green (G), and Blue (B) components. PRISM MPI2-10-MPX2-10 User Manual...
Page 263
All the available tools are located on the top and bottom of the display and are highlighted for a few seconds when the application is first opened or if you select anywhere in the middle of the application tile. PRISM MPI2-10-MPX2-10 User Manual...
Page 264
2. Select one of the preset fixed-gain values from the drop-down list. Choose from 1.0x, 2.0x, 5.0x, or 10.0x. 3. Select the Reset icon ( ) to reset the gain to 1.0x. Horizontal and Vertical Position Adjust the position of the trace display. PRISM MPI2-10-MPX2-10 User Manual...
Page 265
– Use a mouse scroll wheel to move the trace. – For incremental adjustments, select the plus or minus button to adjust the vertical position. – Select the Reset icon ( ) to reset the vertical position adjustment to default. PRISM MPI2-10-MPX2-10 User Manual...
Page 266
1. Select the button in the bottom right corner of the application to expand the available display format options. 2. Select the Mag option to adjust the magnification to Best View, x1, x10, x20, x25, or x50. PRISM MPI2-10-MPX2-10 User Manual...
Page 267
– To scroll through lines, select the + and – buttons. – To enter a specific line to monitor: • select Line Sel. and enter the line on the keypad, • in the Picture application, use the line-select cursor (for active lines). PRISM MPI2-10-MPX2-10 User Manual...
Page 268
Note: In link, the letters (A-D) refers to the physical links. The numbers (1-4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
Page 269
Diamond: The normal Diamond display shows Gamut violations of the SDI input if translated to RGB color space. • Split Diamond: The split Diamond display offsets the two halves of the Diamond to better see negative RGB Gamut errors. PRISM MPI2-10-MPX2-10 User Manual...
Errors in green amplitude affect both diamonds equally, while blue amplitude errors affect only the top diamond and red errors affect only the bottom diamond. PRISM MPI2-10-MPX2-10 User Manual...
Page 271
Monochrome signals appear as vertical lines. Nonlinear component processing, such as from a gamma corrector that alters white balance, can cause deviations along the vertical axis. PRISM MPI2-10-MPX2-10 User Manual...
Page 272
• the Line number is the selected line, including vertical blanking, • if the signal is interlaced format, the Fields readout is below the Line number, • the Pic Line number is next to Line; the Pic Line is the active line number, or numbers, selected. PRISM MPI2-10-MPX2-10 User Manual...
Page 273
Note: In link, the letters (A-D) refers to the physical links. The numbers (1-4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
Page 274
2. Select the tile with the Waveform application. 3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. PRISM MPI2-10-MPX2-10 User Manual...
Page 275
If 1 Line is selected, the display only shows results for the selected line in the picture. Line Select Function. Or the line The line can be selected using the on-screen tools; see can be selected directly on the Picture. PRISM MPI2-10-MPX2-10 User Manual...
Page 276
1. Select the button to in the top left corner to expand the available options drop down list. 2. Select one of the preset magnifications for fixed horizontal and vertical gain using the H Gain and V Gain options. PRISM MPI2-10-MPX2-10 User Manual...
Page 277
– 5.0x has a range of 1.25x to 10.00x gain. – 10.0x has a range of 2.50x to 20.00x gain To adjust the gain: • With a touchscreen, use a pinch gesture to change the scale of the trace for large incremental changes. PRISM MPI2-10-MPX2-10 User Manual...
Page 278
– Use a mouse scroll wheel to move the trace. – For incremental adjustments, use the plus or minus button to adjust the horizontal position. – Select the Reset icon ( ) to reset the horizontal position adjustment to default. PRISM MPI2-10-MPX2-10 User Manual...
Page 279
) to reset the vertical position adjustment to default. Bar Targets Change the scaling of the Lightning Display. 1. Select Bars in the bottom right corner of the application to expand the available options. 2. Select 75% or 100% scaling. PRISM MPI2-10-MPX2-10 User Manual...
Page 280
• the Line number is the selected line, including vertical blanking, • if the signal is interlaced format, the Fields readout is below the Line number, • the Pic Line number is next to Line; the Pic Line is the active line number, or numbers, selected. PRISM MPI2-10-MPX2-10 User Manual...
Page 281
Note: In link, the letters (A-D) refers to the physical links. The numbers (1-4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
3. Select the Tile icon ( Note: The Move icon ( ) is in the settings menu header. Select this icon to move the settings left of right. Color Space Select either CIE 1931 or CIE 1976 diagrams. PRISM MPI2-10-MPX2-10 User Manual...
Page 283
Note: The same line is selected in any trace display if line select is enabled. To turn on Line Select (and choose the line to monitor): 1. Select the Lines button at the bottom of the tile. PRISM MPI2-10-MPX2-10 User Manual...
Page 284
• the Line number is the selected line, including vertical blanking, • if the signal is interlaced format, the Fields readout is below the Line number, • the Pic Line number is next to Line; the Pic Line is the active line number, or numbers, selected. PRISM MPI2-10-MPX2-10 User Manual...
Page 285
Note: In link, the letters (A-D) refers to the physical links. The numbers (1-4) refers to the virtual links. For example, A3/62 indicates it is physical link 1 and virtual link 3 (in a 2-sample interleave, 12G transport). PRISM MPI2-10-MPX2-10 User Manual...
Page 286
If your instrument has Dolby monitoring capabilities, it can decode and monitor audio signals that are based on Dolby digital surround sound formats. These formats are Dolby D/D Plus (AC-3) compression (designed for distribution) and Dolby E compression (designed for production). The next sample shows Dolby E data. PRISM MPI2-10-MPX2-10 User Manual...
Page 287
Dolby E audio, or VANC metadata if the Dolby E is externally decoded. • Auto: Selects the Dolby E audio by default as the metadata source, if it is available. If Dolby E audio is not available, it uses VANC audio. PRISM MPI2-10-MPX2-10 User Manual...
Page 288
The audio channels making up an audio program are assigned to consecutively numbered or consecutively arranged audio channel pairs. The VANC packets carrying the audio metadata for a specific audio program are identified by setting the SDID PRISM MPI2-10-MPX2-10 User Manual...
Page 289
(Surround) channels. If the LFE Channel is present, an L is appended at the end of the channel mode. Program Configuration Determines how the audio channels are grouped within a Dolby E bitstream. Up to eight channels can be grouped together in individual programs, where each program contains its own metadata. PRISM MPI2-10-MPX2-10 User Manual...
Rate, the rate value is displayed in red. Dolby E Frame Loc Displays the location (line number and sample number) of the SMPTE337 preamble. Dialogue Level Represents the long-term, A-weighted average level of dialog within a presentation, Leq(A). PRISM MPI2-10-MPX2-10 User Manual...
Page 291
PTP Operational Overview PTP Operational Overview This section provides an overview of PTP network operation. Not all of the functionality described is currently available in the PRISM monitor. PRISM MPI2-10-MPX2-10 User Manual...
The BMCA is still used to select the active master and the network still takes the place of the ECO in the legacy Black-burst/Tri- level network. The PRISM monitor is a slave-only device. PRISM MPI2-10-MPX2-10 User Manual...
Page 293
After the slave receives the delay response, it has the second pair of time stamps needed to calculate the second delay measurement of the slave to master delay. PRISM MPI2-10-MPX2-10 User Manual...
PTP Operational Overview Five Basic PTP Timing Messages Message Timing The illustration shows the relationships between the five PTP timing messages. PRISM MPI2-10-MPX2-10 User Manual...
Page 295
PTP on different profiles. For example, domain 0 might be an AES67 profile, domain 1 could be a master using the AVB (802.1AS) profile, and domain 127 could be a master on the SMPTE ST2059 profile. PRISM MPI2-10-MPX2-10 User Manual...
Page 296
One type of device which often requires two-step operation is a Transparent Clock, since it has to calculate the residence time of a message and insert that time into the message or the follow-up message. PRISM MPI2-10-MPX2-10 User Manual...
Page 297
AMTs of all the slaves. In Unicast PTP systems, the slave must have the address of the master in its acceptable master table because the slave initiates the communication with the master. The slave PRISM MPI2-10-MPX2-10 User Manual...
Page 298
To use this combination with an SPG8000A master clock generator, set the master and slave to any of the modes: Multicast, Mixed, or Mixed without negotiation. The masters support all of the slave modes simultaneously. PRISM MPI2-10-MPX2-10 User Manual...
If slaves on a given network have a different list of masters in their AMT, then they may choose a different master. Mixed Mode In a SMPTE ST2059 mixed environment, the announce message is Multicast. Therefore, the BMCA can follow the Multicast conventions. PRISM MPI2-10-MPX2-10 User Manual...
Page 300
To reduce this effect, a user should avoid large loading on the PTP network, use high-quality switches that avoid this delay variance, or use PTP Transparent mode switches which insert residence time information to allow the slave to correct for this delay. PRISM MPI2-10-MPX2-10 User Manual...
Page 301
The PTP system is incapable of removing this effect, so the best way to prevent this is to use high-quality cable with a minimum variation in the physical propagation delay. PRISM MPI2-10-MPX2-10 User Manual...
Page 302
PTP Operational Overview Compensating for Causes of Asymmetric Delay PRISM MPI2-10-MPX2-10 User Manual...