4.0 Operation
• Use the main switch to start the unit.
• Readiness for operation after 1 second.
• Completely press the hose tip onto the handle.
If the hose tip is not completely attached onto the handle,
!
damage could be caused to the patient's eardrum.
4.1 Adjusting temperatures
• Two variable temperature levels (20 °C ‒ 47 °C) (48 °C +
49 °C only for testing purposes).
) The lowest achievable stimulation temperature is approx.
2 °C over the ambient temperature.
• Temperature setting by coding switches (
- Left switch: for adjusting the "ten" partition
- Right switch: for adjusting the "one" partition
ª Lower keys (+): temperature increase
ª Upper keys (-): temperature decrease
• Standard settings:
- Level for cold stimulation: 30 °C
- Level for warm stimulation: 44 °C
Avoid bending the tip of the hose within the ear / auditory
!
canal. Otherwise, the error "F1" or "F7" may occur.
4.2 Selecting temperature levels
• For selecting the desired temperature level, use the
respective key (, , Fig. 1).
ª Display of the active level via LEDs.
ª Display of the air temperature (current value) in °C.
• For switching off the heating system, press the respective
key (, , Fig. 1) of the active temperature level.
ª LED of the temperature level goes out.
ª Display of the air temperature (current value) in °C.
4.3 Adjusting stimulation time
• By means of coding switch (, Fig. 1).
4.4 Description of operating modes
4.4.1 Preparation mode
Purpose:
The temperature set by the user is adjustable.
Properties:
• Temperature: corresponds to the preselected cold or
• Air flow: 5.0 l/min.
Activation:
• By operating one of the key switches for selecting a
,
, Fig. 1).
• If the active temperature push button is pressed repeat-
Deactivation:
• Operating the key switch heating/air flow ON/OFF (,
• Automatic switch to standby mode when the unit is not
warm stimulation level.
temperature level (, , Fig. 1) or by operating the key
switch for heating/air flow ON/OFF (, Fig. 1) in the
standby mode.
edly, the heating is switched off.
ª Air with a temperature that almost corresponds to the
ambient air temperature is available.
Fig. 1) switches the unit to the standby mode.
used within a 3-minute period.
9