11.0 Notes on EMC

• Medical electrical equipment is subject to special precautions with regard to EMC and must be installed according to the follow-
ing EMC notes.
• Portable and mobile RF communication facilities can influence medical electrical equipment.
• The use of other accessories, converters, and cables than stated may lead to an increased emission or a reduced interference
immunity of the equipment or system.
11.1 Guidelines and Manufacturer's Declaration ‒ Emissions
The Varioair 3 is designed for operation in the environment specified below. The customer or user of the Varioair 3 should ensure
that it is used in such an environment.
Emissions Test
RF Emissions acc.to CISPR 11
RF Emissions acc. to CISPR 11
Harmonic emissions according to
IEC 61000-3-2
Voltage fluctuations/flicker according
to IEC 61000-3-3
The device may not be used directly next to other devices or piled up with other devices. If operation next to or piled with other
devices is necessary, please watch the device to check its intended operation in this arrangement.
11.2 Guidelines and Manufacturer's Declaration ‒ Immunity
The Varioair 3 is designed for operation in the electromagnetic environment specified below. The customer or user of the Varioair 3
should ensure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic dis-
charge (ESD) accord-
ing to IEC 61000-4-2
EFT IEC 61000-4-4
Surges IEC 61000-
4-5
Magnetic field at pow-
er frequency 50/60 Hz
acc. to IEC 61000-4-8
Compliance
Electromagnetic Environment
Group 1
The Varioair 3 uses HF energy for its internal functions. There-
fore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
Class A
The Varioair 3 is suitable for use in all establishments, including
domestic and those connected directly to a public power supply
network that supplies buildings used for residential purposes.
Corresponds
Compliance Level
± 6 kV Contact
± 8 kV Air
± 2 kV Mains
± 1 kV I/Os
1 kV
Common
2 kV
Differential
3 A/m
Guidance
Electromagnetic Environment
Floors should be made of wood or concrete or laid
with ceramic tiles. If floors are synthetic, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Power frequency magnetic fields should be that of a
typical commercial or hospital environment.
Guidance
17