Dissolved Oxygen monitor
and controller
INSTRUCTION MANUAL
INSTRUCTION MANUAL
MES
Dissolved oxygen monitoring
MESURES
BAMOPHOX 450
22, Rue de la Voie des Bans - Z.I. de la Gare - 95100 ARGENTEUIL
450-02
Tél : (+33) 01 30 25 83 20 - Web : www.bamo.fr
/1
Fax : (+33) 01 34 10 16 05 - E-mail : info@bamo.fr
11-10-2007
450 M1 02 B

Advertisement

Table of Contents
loading

Summary of Contents for Bamo BAMOPHOX 450 E

  • Page 1: Adjust Threshold

    22, Rue de la Voie des Bans - Z.I. de la Gare - 95100 ARGENTEUIL 450-02 Tél : (+33) 01 30 25 83 20 - Web : www.bamo.fr Fax : (+33) 01 34 10 16 05 - E-mail : info@bamo.fr...
  • Page 2: Table Of Contents

    Dissolved Oxygen monitor and controller BAMOPHOX 450 E & M (Technical information and Manual for LOGGER /RS422 version are on a specific document) Content 1. TECHNICAL FEATURES Page 3 2. DIMENSIONS 3. WIRING 4. FRONT PANEL PRESENTATION AND SCROLLING MENU...
  • Page 3: Technical Features

    1. TECHNICAL FEATURES Displayed parameters: Measurement values - Configuration Menu - Temperature value Display: Back lighted - 2 lines of 16 alphanumerical characters; 9.2 mm high Indication: LED alarms status Configuration: 8 push buttons keyboard on the front - Keyword protected Scales: 0 to 100% –...
  • Page 4: Wiring

    3. WIRING WALL MOUNTING MODEL OPTION : LOGGER & RS 422 PANEL MOUNTING MODEL OPTION : LOGGER (access by the upper cover) & RS 422 Page 450 M1 02 B 11-10-2007...
  • Page 5 Screwing terminals + mA ouput / measurement - mA ouput / measurement - mA ouput / temperature + mA ouput / temperature + 24 V To switch the regulation on stand-by - 24 V - 24 V Sensor power supply = 20 mA + 24 V Pt 100 Ω...
  • Page 6: Front Panel

    4. FRONT PANEL S1, S2, S3, and S4 2 lines /16 alphanumeric characters indicate relays status: 9.22 mm high - Back lighted LED lighting = contact ON Key “A” LED OFF = contact OFF To display the parameters of upper line. LED flashing = Timer in use (main BAMOPHOX) Key “B”...
  • Page 7: Presentation And Scrolling Menu

    SCROLLING MENU 00.00 m/L (Display during measurement) MENU ABOUT BAMOPHOX MENU CONSULTATION / MODIFICATION MENU MEASUREMENT PARAMETERS MENU ATMOSPHERIC PRESSURE MENU SALINITY MENU ADJUST ALARM 1 MENU ADJUST ALARM 2 MENU ADJUST ALARM 3 MENU RELAY REGULATION MENU REGUL. PID MENU OUTPUT mA Oxygen MENU...
  • Page 8: About Bamophox

    ABOUT Bamophox ABOUT BAMOPHOX ENTER VERSION 2.04 ENTER SERIAL N° ENTER 20451 05 ENTER CONSULTATION / MODIFICATION CONSULTATION MODIFICATION ENTER CODE ? 0000 ENTER CODE ? 5105 Last 4 digits (of serial number) are the key code to access the MODIFICATION menu. ENTER When wrong code is entered, a message "ERROR"...
  • Page 9: Measurement Parameters

    MEASUREMENT PARAMETERS ENTER DISPLAY mg /I Choose the right scale, confirm with "Enter" then save. ENTER (OFFSET TEMPERATURE) Enter the difference between real value and reference temperature OT +00,00 °C Example: Real value 22 °C Reference 20 °C Entry +2,00 °C ENTER SAVING ? VALID...
  • Page 10 ADJUST ALARM 1 MENU ADJUST ALARM 2 ENTER ALARM 1 ON/OFF ENTER ALARM 1 MEASURE/TEMP MEASURE= Threshold against pH/mV measured value TEMP.= Threshold against temperature measured value ENTER ALARM 1 LOW/HIGH HIGH= Contact closes when value goes over the limit LOW= Contact closes when the value goes under the limit ENTER ON 00,00 mg/l / °C...
  • Page 11 ADJUST ALARM 3 MENU RELAY REGULATION please, see page 12 ENTER ALARM 3 ON/OFF ENTER ALARM 3 MEASURE/TEMP MEASURE= Threshold against pH/mV measured value TEMP.= Threshold against temperature measured value ENTER ALARM 3 LOW/HIGH HIGH= Contact closes when value goes over the limit LOW= Contact closes when the value goes under the limit ENTER ON 00,00 mg/l / °C...
  • Page 12: On/Off Regulation

    RELAY REGULATION MENU REGUL. PID Please see page 13 ENTER ENTER ON/OFF MODE REGUL. PID REGUL ON/OFF ENTER ENTER ENTER SET VAL. 00,00 mg/l Set point ENTER T. CYCLE 0000 SEC Cycle time ENTER Highest limit for HIGH PB 00,00 mg/l proportional bandwith ENTER Lowest value...
  • Page 13: Pid Regulation

    REGUL PID MENU Output mA Oxygen Please see page 14 This operating mode allows a PID regulation with an analogic output 0/20 or 4/20 mA REGUL ON/OFF To switch ON or OFF the regulation mode ENTER REGUL AUTO/MANU MANU=MANUAL to be able to check the relays output ENTER SET VAL 00,00 mg/l Set point value...
  • Page 14: Output Ma Dissolved Oxygen

    Output mA Oxygen Output mA TEMP. MENU Measurement signal copy on the analog output ENTER HIGHER 00,00 mg/l Value corresponding to 20,00 mA ENTER LOWER 00,00 mg/l Value corresponding to 00,00 or 04,00 mA ENTER OUTPUT 4/20 mA / 0/20 mA Output type ENTER SAVING ?
  • Page 15: Sensor Calibration

    CALIBRATION MENU FORCED RELAYS Please see page 16 ENTER SOL ZERO Immerse the sensor in a solution of 17 g of sodium sulphite anhydrous (Na2SO3) in distilled water. See also the probe manual. ENTER PROBE IN AIR Let the probe in the air, moving it to have a better exchange. See also the probe manual ENTER COEF 0000...
  • Page 16: Set Up Of Alarms

    FORCED RELAY ADJUST ALARM MENU ENTER ALARM 1 ON/OFF ENTER ALARM 2 ON/OFF ENTER Diagnostic mode to test the threshold configurations ALARM 3 ON/OFF ENTER ALARM 4 ON/OFF VALID ADJUST ALARM MENU LANGUAGE When in use the S4 contact is active. This mode allows to detect a malfunction on ENTER and S2 contacts ;...