Turbidity monitor and controller BAMOPHOX 436 E & M (Technical information and Manual for LOGGER /RS422 version are on a specific document) Table of contents 1. TECHNICAL SPECIFICATIONS Page 3 2. CAUTION 3. DIMENSIONS 4. WIRING 5. FRONT PANEL SCROLLING MENU...
1. TECHNICAL FEATURES Displayed parameters: Measurement values of Turbidity - Configuration Menu - Temperature value Display: Back lighted - 2 lines of 16 alphanumerical characters ; 9,2 mm high Indication: LED alarms status Configuration: 8 push buttons keyboard on front face - Keyword protected Units: NTU, FAU, mg/L, g/L, FNU Scale...
Page 5
Screwing terminals + mA ouput / measurement - mA ouput / measurement - mA ouput / temperature + mA ouput / temperature + 24 V To switch the regulation on stand-by - 24 V - 24 V Power supply = 20 mA + 24 V Pt 100 Ω...
Page 6
Screwing terminals + mA ouput / measurement - mA ouput / measurement - mA ouput / temperature + mA ouput / temperature + 24 V To switch the regulation on stand-by - 24 V - 24 V Power supply = 20 mA + 24 V Pt 100 Ω...
5. FRONT PANEL S1, S2, S3, and S4 2 lines /16 alphanumeric characters indicate relays status: 9.22 mm high - Back lighted LED lighting = contact ON LED OFF = contact OFF LED flashing = Timer in use Key “A” To display the parameters of upper line.
SCROLLING MENU Display the turbidity value MENU ABOUT BAMOPHOX To consult serial number, instrument type MENU CONSULTATION / MODIFICATION MENU SCALE PARAMETERS MENU ADUST ALARM 1 MENU ADUST ALARM 2 MENU ADUST ALARM 3 MENU OUTPUT mA TURB MENU OUTPUT mA TEMP MENU FORCED DISPLAY MENU...
MENU SCALE PARAMETERS ADJUST ALARM 1 page 10 ENTER UNIT Choose between NTU FAU, mg/l, g/l, FNU ENTER x-x,xxx NTU Decimal point selection ENTER SCALE ENTER MINI : 00,00 NTU Minimal range value ENTER MAXI : 00,00 NTU Maximal range value ENTER TRANSMITTER ENTER...
Page 10
ADJUST ALARM 1 MENU ADJUST ALARM 2 ENTER ALARM 1 ON/OFF ENTER ALARM 1 MEASURE/TEMP MEASURE= Threshold against pH/mV measured value TEMP.= Threshold against temperature measured value ENTER ALARM 1 LOW/HIGH HIGH= Contact closes when value goes over the limit LOW= Contact closes when the value goes under the limit ENTER ON 0000 NTU / °C...
ADJUST ALARM 3 MENU OUTPUT mA TURB ENTER ALARM 3 ON/OFF ENTER ALARM 3 MEASURE/TEMP MEASURE= Threshold against pH/mV measured value TEMP.= Threshold against temperature measured value ENTER ALARM 3 LOW/HIGH HIGH= Contact closes when value goes over the limit LOW= Contact closes when the value goes under the limit ENTER ON 0000 NTU / °C...
OUTPUT mA TEMP. MENU MARCHE FORCEE Caution: If PID regulation is active, this step menu would not appears ENTER HIGHER 0000 NTU Value corresponding to 20,00 mA ENTER LOWER 0000 NTU Value corresponding to 00,00 mA or 04,00 mA ENTER OUTPUT 4/20 mA / 0/20 mA Output type ENTER...