Amperometric signal monitor BAMOPHOX 194 (Technical information and Manual for LOGGER /RS422 version are on a specific document) Table of contents Page 1. TECHNICAL SPECIFICATIONS 2. DIMENSIONS 3. WIRING 4. FRONT PANEL SCROLLING MENU ABOUT BAMOPHOX CONSULTING / MODIFYING TESTING RELAY OUTPUTS...
1. TECHNICAL FEATURES Displayed parameters: Measured values - Configuration Menu - Temperature Display: Back lighted - 2 lines of 16 alphanumerical characters ; 9 mm high Indication: LED alarms status Configuration: 8 push buttons keyboard on front board - Keyword protected Sensor input: For amperometric sensors 4-20 mA (2 wire technique) or 0/-2 V (4 wire technique) Scales:...
3. WIRINg PANEL MOUNTINg INSTRUMENT OPTION: LOGGER & RS 422 Signal input WALL MOUNTINg INSTRUMENT OPTION: LOGGER (access by upper cover) & RS 422 Signal input Page 4 194 M1 01 F 12-13-2013...
Page 5
Screwing terminals + mA ouput / measurement - mA ouput / measurement - mA ouput / temperature or Regulation PID + mA ouput / temperature or Regulation PID + 24 V - 24 V Flow controller connection - 24 V + 24 V Pt 100 Ω...
Page 6
TERMINAL FOR SENSOR CONNECTION TERMINAL FOR SENSOR CONNECTION Load resistor supplied with the BAMOPHOX 194 4-20 mA sensor 0 to -2000 mV sensor Alim = 24 V Shunt Brown = + Blue = – Black = output – Flow rate sensor (194 832)
4. FRONT PANEL S1, S2, S3, and S4 2 lines /16 alphanumeric characters indicate relays status: 9.22 mm high - Back lighted LED lighting = contact ON LED OFF = contact OFF LED flashing = Timer in use Key “A” To display the parameters of upper line.
PRESENTATION ET DEFILEMENT DU MENU Press ENTER to change the display ENTER MESU. 00,00 ppm +20,0°C from MEASUREMENT or TEMPERATURE MENU ABOUT BAMOPHOX MENU CONSULTATION / MODIFICATION MENU SENSOR CONFIGURATION MENU ADJUST ALARM 1 MENU ADJUST ALARM 2 MENU ADJUST ALARM 3 MENU RELAY REGULATION MENU...
ABOUT Bamophox ABOUT BAMOPHOX ENTER VERSION 2.04 ENTER SERIAL N° ENTER 20451 05 ENTER CONSULTATION / MODIFICATION CONSULTATION ▲ MODIFICATION ENTER CODE ? 0000 ▲ ENTER CODE ? 5105 The last 4 digits (of serial number) are the key code to access to the MODIFICATION menu.
P.I.D. regulation with the 4-20 mA signal There is a specificity of alarm 3 in Bamophox 194. It can be setup for an external sensor. Signal is coming from an inductive flow rate sensor type NPN. This sensor is mounted in the flow cell.
Page 11
ADJUST ALARM 1 Next menu: MENU ADJUST ALARM 2 ENTER ALARM 1 ON/OFF ▲ ENTER ALARM 1 MEASURE/TEMP MEASURE= Threshold against measured value ▲ TEMP.= Threshold against temperature measured value ENTER ALARM 1 LOW/HIgH HIgH= Contact closes when value goes over the limit ▲...
Page 12
ADJUST ALARM 3 Next menu: RELAY REGULATION MENU ENTER ▲ ALARM 3 ON/OFF ENTER EXTERN NO/YES NO= set up (ALARM 3 MEASURE/TEMP) ▲ YES= flow rate sensor wired SAVINg ? VALID ENTER ▲ ALARM 3 MEASURE/TEMP MEASURE= Threshold against measured value TEMP.= Threshold against temperature measured value ENTER ▲...
RELAY REgULATION Next menu: REGUL PID MENU ENTER ENTER ON/OFF MODE REgUL. PID REGUL ON/OFF ▲ ENTER ENTER ENTER SET VAL. 0,000 ppm ▲ Set point ENTER ▲ T. CYCLE 0000 SEC Cycle time ENTER Highest limit for ▲ HIGH PB 00.00 ppm proportional bandwith ENTER Lowest value for...
Page 14
Example With process configuration: - Set point: 1.5 ppm - High dead zone: 0.4 ppm between 1.5 and 1.9 ppm - Low dead zone : 0.1 ppm between 1.4 and 1.5 ppm - High BANDWIDTH: 3.5 ppm (that is a limit of 5 ppm for the maximum) - Low BANDWIDTH: 0.5 ppm (that is a limit of 1 ppm for the minimum) - Over the highest limit (>5 ppm), S2 is “ON”...
Next menu: Output mA MENU REgUL PID MEASUREMENT This operating mode allows a PID regulation with an analogic output 0/20 or 4/20 mA (terminals 3 and 4). This output is no more available for temperature output. ▲ REGUL ON/OFF To switch ON or OFF the regulation mode ENTER ▲...
Output mA MEASUREMENT Next menu: MENU Output mA TEMP. Measurement signal copy on the analog output ENTER ▲ HIGHER 00.00 ppm Value corresponding to 20,00 mA ENTER ▲ LOWER 00.00 ppm Value corresponding to 0.00 or 04,00 mA ENTER ▲ OUTPUT 4/20 mA / 0/20 mA Output type ENTER...
ELECTRODE CALIBRATION Next menu: FORCED RELAY MENU CAUTION - Temperature compensation: set up the instrument on MANUAL mode at 20°C During calibration in mode “MODIFICATION”, measurement value is on stand-by on the last reading value, the regulation is on stand-by. A delay after ending the calibration allows a restart of the complete system before to automatically restart the regulation mode.
FORCED RELAY Next menu: ADJUST ALARM MENU ENTER ▲ ALARM 1 ON / OFF ENTER ▲ ALARM 2 ON / OFF Diagnostic mode to test manually the threshold configurations ENTER ▲ ALARM 3 ON / OFF ENTER ▲ ALARM 4 ON / OFF ENTER ADJUST ALARM Next menu:...