Bamo BAMOPHOX 194 Instruction Manual

Amperometric sensors signal monitor

Advertisement

Quick Links

BAMOPHOX 194
Amperometric sensors signal monitor
INSTRUCTIONS MANUAL
MES
Amperometric signal monitor
BAMOPHOX 194
194-01
/1
13-12-2013
194 M1 01 F

Advertisement

Table of Contents
loading

Summary of Contents for Bamo BAMOPHOX 194

  • Page 1: Adjusting Threshold

    BAMOPHOX 194 Amperometric sensors signal monitor INSTRUCTIONS MANUAL Amperometric signal monitor BAMOPHOX 194 194-01 13-12-2013 194 M1 01 F...
  • Page 2: Table Of Contents

    Amperometric signal monitor BAMOPHOX 194 (Technical information and Manual for LOGGER /RS422 version are on a specific document) Table of contents Page 1. TECHNICAL SPECIFICATIONS 2. DIMENSIONS 3. WIRING 4. FRONT PANEL SCROLLING MENU ABOUT BAMOPHOX CONSULTING / MODIFYING TESTING RELAY OUTPUTS...
  • Page 3: Technical Specifications

    1. TECHNICAL FEATURES Displayed parameters: Measured values - Configuration Menu - Temperature Display: Back lighted - 2 lines of 16 alphanumerical characters ; 9 mm high Indication: LED alarms status Configuration: 8 push buttons keyboard on front board - Keyword protected Sensor input: For amperometric sensors 4-20 mA (2 wire technique) or 0/-2 V (4 wire technique) Scales:...
  • Page 4: Wiring

    3. WIRINg PANEL MOUNTINg INSTRUMENT OPTION: LOGGER & RS 422 Signal input WALL MOUNTINg INSTRUMENT OPTION: LOGGER (access by upper cover) & RS 422 Signal input Page 4 194 M1 01 F 12-13-2013...
  • Page 5 Screwing terminals + mA ouput / measurement - mA ouput / measurement - mA ouput / temperature or Regulation PID + mA ouput / temperature or Regulation PID + 24 V - 24 V Flow controller connection - 24 V + 24 V Pt 100 Ω...
  • Page 6 TERMINAL FOR SENSOR CONNECTION TERMINAL FOR SENSOR CONNECTION Load resistor supplied with the BAMOPHOX 194 4-20 mA sensor 0 to -2000 mV sensor Alim = 24 V Shunt Brown = + Blue = – Black = output – Flow rate sensor (194 832)
  • Page 7: Front Panel

    4. FRONT PANEL S1, S2, S3, and S4 2 lines /16 alphanumeric characters indicate relays status: 9.22 mm high - Back lighted LED lighting = contact ON LED OFF = contact OFF LED flashing = Timer in use Key “A” To display the parameters of upper line.
  • Page 8: Scrolling Menu

    PRESENTATION ET DEFILEMENT DU MENU Press ENTER to change the display ENTER MESU. 00,00 ppm +20,0°C from MEASUREMENT or TEMPERATURE MENU ABOUT BAMOPHOX MENU CONSULTATION / MODIFICATION MENU SENSOR CONFIGURATION MENU ADJUST ALARM 1 MENU ADJUST ALARM 2 MENU ADJUST ALARM 3 MENU RELAY REGULATION MENU...
  • Page 9: About Bamophox

    ABOUT Bamophox ABOUT BAMOPHOX ENTER VERSION 2.04 ENTER SERIAL N° ENTER 20451 05 ENTER CONSULTATION / MODIFICATION CONSULTATION ▲ MODIFICATION ENTER CODE ? 0000 ▲ ENTER CODE ? 5105 The last 4 digits (of serial number) are the key code to access to the MODIFICATION menu.
  • Page 10: Sensor Configuration

    P.I.D. regulation with the 4-20 mA signal There is a specificity of alarm 3 in Bamophox 194. It can be setup for an external sensor. Signal is coming from an inductive flow rate sensor type NPN. This sensor is mounted in the flow cell.
  • Page 11 ADJUST ALARM 1 Next menu: MENU ADJUST ALARM 2 ENTER ALARM 1 ON/OFF ▲ ENTER ALARM 1 MEASURE/TEMP MEASURE= Threshold against measured value ▲ TEMP.= Threshold against temperature measured value ENTER ALARM 1 LOW/HIgH HIgH= Contact closes when value goes over the limit ▲...
  • Page 12 ADJUST ALARM 3 Next menu: RELAY REGULATION MENU ENTER ▲ ALARM 3 ON/OFF ENTER EXTERN NO/YES NO= set up (ALARM 3 MEASURE/TEMP) ▲ YES= flow rate sensor wired SAVINg ? VALID ENTER ▲ ALARM 3 MEASURE/TEMP MEASURE= Threshold against measured value TEMP.= Threshold against temperature measured value ENTER ▲...
  • Page 13: On/Off Regulation

    RELAY REgULATION Next menu: REGUL PID MENU ENTER ENTER ON/OFF MODE REgUL. PID REGUL ON/OFF ▲ ENTER ENTER ENTER SET VAL. 0,000 ppm ▲ Set point ENTER ▲ T. CYCLE 0000 SEC Cycle time ENTER Highest limit for ▲ HIGH PB 00.00 ppm proportional bandwith ENTER Lowest value for...
  • Page 14 Example With process configuration: - Set point: 1.5 ppm - High dead zone: 0.4 ppm between 1.5 and 1.9 ppm - Low dead zone : 0.1 ppm between 1.4 and 1.5 ppm - High BANDWIDTH: 3.5 ppm (that is a limit of 5 ppm for the maximum) - Low BANDWIDTH: 0.5 ppm (that is a limit of 1 ppm for the minimum) - Over the highest limit (>5 ppm), S2 is “ON”...
  • Page 15: Pid Regulation

    Next menu: Output mA MENU REgUL PID MEASUREMENT This operating mode allows a PID regulation with an analogic output 0/20 or 4/20 mA (terminals 3 and 4). This output is no more available for temperature output. ▲ REGUL ON/OFF To switch ON or OFF the regulation mode ENTER ▲...
  • Page 16 ppm measurement Page 16 194 M1 01 F 12-13-2013...
  • Page 17: Analog Output 4-20 Ma For Measurement Signal

    Output mA MEASUREMENT Next menu: MENU Output mA TEMP. Measurement signal copy on the analog output ENTER ▲ HIGHER 00.00 ppm Value corresponding to 20,00 mA ENTER ▲ LOWER 00.00 ppm Value corresponding to 0.00 or 04,00 mA ENTER ▲ OUTPUT 4/20 mA / 0/20 mA Output type ENTER...
  • Page 18: Sensor Calibration

    ELECTRODE CALIBRATION Next menu: FORCED RELAY MENU CAUTION - Temperature compensation: set up the instrument on MANUAL mode at 20°C During calibration in mode “MODIFICATION”, measurement value is on stand-by on the last reading value, the regulation is on stand-by. A delay after ending the calibration allows a restart of the complete system before to automatically restart the regulation mode.
  • Page 19: Testing Regulation Mode

    FORCED RELAY Next menu: ADJUST ALARM MENU ENTER ▲ ALARM 1 ON / OFF ENTER ▲ ALARM 2 ON / OFF Diagnostic mode to test manually the threshold configurations ENTER ▲ ALARM 3 ON / OFF ENTER ▲ ALARM 4 ON / OFF ENTER ADJUST ALARM Next menu:...