Page 9
881A/L-GS Scanning Sonar Acronyms and Abbreviations Acronym Definition ASCII American Standard Code for Information Interchange Decibel Echo Sounder Gigabyte Global Positioning System Gyro Stabilized Internet Protocol Kilohertz Meters per second Megahertz NMEA National Marine Electronics Association Personal Computer Random Access Memory Remotely Operated Vehicle Terabytes User Datagram Protocol...
Both units deliver a high level of performance with the 881L-GS ethernet unit able to transfer data at a higher data rate. A profiling version of the 881A-GS has been tailored for use in a borehole surveying role and contains features for auto scanning boreholes (see 3.6.9 Profile > Auto Profile > Horizontal Scan (Borehole)…).
881A/L-GS Scanning Sonar The general features are: 310 kHz, 675 kHz, or 1 MHz (standard default settings) -Other Frequency frequencies can be selected through programmable software configurations (Tunable from 280 kHz to 1.1 MHz in 5 kHz steps) Transducer Imaging type, fluid compensated Range Resolution 1 m –...
Ensure 24 VDC (1A minimum) power supply is off. Configure the network card’s IP address or COM port when using 881L-GS and 881A-GS respectively. (see Sections 3.4.2 System > IP Address and 3.4.1 System > Choose COM Ports) ...
If an 881A-GS is used – an available RS-485 or RS-232 (depending on the type of receiver) port is required. If no COM port is present on the PC, an RS-485 to USB adaptor or an RS-232 to USB adaptor is used.
881A/L-GS Scanning Sonar 2.1.2. Sonar Considerations Figure 1 External View FRAGILE. This area must not be obstructed. Do not attach any brackets or fastenings to this area. Ensure locking sleeve is used. Do not bend pins. Do not remove. Represents moving transducer zero position. Use this area to mount the sonar.
2.2. SOFTWARE SETUP The latest version of the Win881AL_GS program is available from the Imagenex Technology Corp. website (https://imagenex.com/interior-page/software-download). The installation of the software on the PC is straightforward. It can be installed in its own directory, and run by double clicking, or you can create a shortcut from the desktop for it.
881A/L-GS Scanning Sonar 2.3.1. Shortcut Buttons Connect to a physical sonar head. Open a File play-back session. Start / Stop recording an 81R data file. Scan in polar mode which means the sonar will do 360° rotations. Direction (C/W CC/W) is toggled by clicking Scan in a sector defined by the Sector setting.
881A/L-GS Scanning Sonar 2.3.2. File Playback Table 5 Playback Control Used during playback and allows the user to adjust the playback speed and file position. 2.3.3. Sonar Settings Used to control various sonar attributes. Only available in real-time mode: Figure 3 Sonar Settings ...
881A/L-GS Scanning Sonar Description Display current acoustic range and frequency of sonar in yellow. Displays grid scale and current gain in orange. Gyro enabled / disabled in green. Displays only when Gyro Stabilized Image is enabled. One of: North Up, Heading Up, or Target Steering. Displays “Not Recording”...
881A/L-GS Scanning Sonar Acquire Target was selected and the mouse cursor changed The desired target was selected. The screen automatically to the selection cursor. The target cursor is positioned over centred the target within a sector and rotated the sector to the target of interest and once clicked, >>...
881A/L-GS Scanning Sonar 2.3.7. Compass Tape Display Enable View > Compass Display. Displays the current heading as a moving tape display: Figure 7 Compass Tape Display 2.3.8. ROV Overlay Displays the ROV overlay if View > ROV Overlay is selected. Size can be adjusted by using the Z speed key, which toggles through various size options.
881A/L-GS Scanning Sonar 3. SOFTWARE MENUS 3.1. FILE 3.1.1. File > Connect to Sonar Head Select data source to be from the sonar head and viewed in real-time. 3.1.2. File > Open File to Playback Data can be replayed from a previously recorded file from the Open File dialog box. 3.1.3.
881A/L-GS Scanning Sonar 3.1.8. File > Convert 81R File to xyz… Converts an 81R data file to an ASCII text XYZ file. X and Y are the output from a polar to rectangular conversion, and z is calculated from the Auto Profile settings (see 3.6.9 Profile > Auto Profile). X,Y and Z values are corrected for pitch and roll.
881A/L-GS Scanning Sonar 3.2.5. View > Zoom In Shortcut button: Zooms in by the factor specified in Zoom Factor. 3.2.6. View > Zoom Out Shortcut button: Zooms out by the factor specified in Zoom Factor. 3.2.7. View > Grid Choose between None, Rectangular and Polar grids. 3.2.8.
881A/L-GS Scanning Sonar 3.2.14. View > Show Sonar Settings Realtime only. Toggles a display window on / off that shows various sonar settings and allows the setting of the head ID. The head IDs have been programmed into the sonar at default 16. Care should be taken when changing this value as it may lead to loss of communication.
Allows for the configuration of communication ports: If an 881A-GS is being used, the sonar communication port parameters are set in the Sonar Head section. The Sonar Head configuration is not relevant with 881L-GS as it is connected via ethernet.
881A/L-GS Scanning Sonar 3.4.5. System > Sonar Orientation Realtime only. Select Up if sonar orientated with red polyurethane housing (transducer) towards the top and Down if inverted. (See Figure 9 and Figure 10). Note that although the unit is rotated through 180° the pitch, roll and heading sensors will be corrected for sign conversion in both orientations.
Page 29
881A/L-GS Scanning Sonar Orientation Axis Convention (Compass North = Diagram Forward) Pitch - Rotating + = Pitch Up around axis A (into the page) Transducer Up Roll – Rotating around Axis B + = Starboard Down (into the page) Doc No: 430-044-00 Page 29 of 46...
Not applicable Table 8 Conventions 3.4.6. System > Sonar Type Realtime only. Various models of sonar use the same interface software. Select 881A-GS or 881L-GS depending on model. 3.5. OPTIONS 3.5.1. Options > Units Meters or feet can be selected as the units used.
881A/L-GS Scanning Sonar 3.5.3. Options > RealTime Play Speed Playback only. Limits the playback speed to the same speed as what was used when data was acquired. 3.5.4. Options > Enable Gyro Enables the internal gyro to be used in North Up, Heading Up and Target Steering and Acquire Target mode. 3.5.5.
881A/L-GS Scanning Sonar Suspend sonar operations, and mount the sonar on a level stable surface, or settle the ROV on a flat bottom. Select Options > Calibrate Gyro. The message Calibrating Gyro will appear. Calibration takes approximately 30 seconds. ...
881A/L-GS Scanning Sonar 3.5.7. Options > Set Heading Reference Realtime only. Manual and Automatic are generally used when the sonar is used in borehole survey applications, and not when it is used, for example, as an imager mounted on an ROV. If the sonar is used as an imager, Disable should be selected.
881A/L-GS Scanning Sonar 3.5.9. Options > Calibrate Motor Realtime only. Calibrates the motor and realigns the sonar head position relative to the transducer position. This can be done periodically to remove unwanted drift. 3.5.10. Options > Local Lat / Compass Declination Realtime only.
Profile > Auto Profile > Horizontal Scan (Borehole)… Realtime only. 881A-GS Profiler Only. Used to generate the Z (depth) component of the XYZ files via an auto increment function. This allows the user to set the expected depth increment for successive scans and if the sonar is paid out to these increments, a realistic profile will be generated of the internal surface of the borehole.
Page 36
881A/L-GS Scanning Sonar This will set a reference angle which is used for subsequent scans to minimize heading changes during the generation of the XYZ output files. The sonar should be running in North Up mode. The procedure detailed below should be followed prior to commencing auto profile scans.
Page 37
881A/L-GS Scanning Sonar A typical deployment procedure is detailed below: An 81R file logging session is started. The sonar is lowered on a marked cable to some predetermined depth. Auto Profile Enabled is checked (see Figure 12). ...
881A/L-GS Scanning Sonar Do not re-calibrate compass, gyro or motor between scans. To terminate the process, uncheck Auto Profile Enable and click OK, or click Cancel. Figure 12 Auto Profile 3.7. 3.7.1. GPS > Enabled Realtime only. If a GPS receiver is connected via a comm port, it may be enabled and disabled here. 3.7.2.
881A/L-GS Scanning Sonar 3.7.4. GPS > Zoom In Realtime only. Zooms in on the track plot. Meters / feet per division is shown in the track-plot window. 3.7.5. GPS > Zoom Out Realtime only. Zooms out on the track plot. Meters / feet per division is shown in the track-plot window. 3.8.
881A/L-GS Scanning Sonar 4. ADDITIONAL DOCUMENTATION Document Number Description 410-013 USB Converters and the Windows OS 430-054 881L Ethernet Troubleshooting Guide Table 9 Additional Document Doc No: 430-044-00 Page 40 of 46...
881A/L-GS Scanning Sonar APPENDIX C SPEC SHEETS Doc No: 430-044-00 Page 46 of 46...
Page 47
200 m radius, 360° surrounding area. The Model 881A-GS still has low power, simple set- up, and small size that make it an ideal tool for large work ROV’s and small inspection vehicles. On it’s own it is now an...
Page 48
1000 m unit: 1.6 kg (3.5 lbs) 3000 m unit: TBA In Water 1000 m unit: TBA 3000 m unit: TBA MATERIALS 1000 m unit: 6061-T6 Aluminum & Polyurethane 3000 m unit: Titanium, Polyurethane & 300 series stainless steel FINISH Hard Anodize 881A-GS Imaging 445-093 www.imagenex.com...
Page 49
3000 m UNIT Standard 881-000-401 RS-232 Option -006 End mount connector Option -009 Right angle connector Option -010 Gyro Stabilization Option -048 Product and company names listed are trademarks or trade names of their respective companies. 881A-GS Imaging 445-093 www.imagenex.com...
Page 50
Drop sonar • • Orientation module Scientific Research Gyro-stabilization of the Imagenex Model 881L-GS makes the high resolution 881L sonar into a system capable of crystal clear visualization of the ocean environment from moving platforms, no longer compromised by the blurring effects of host vehicle rotation.
Page 51
1000 m unit: 1.8 kg (4 lbs) 3000 m unit: TBA In Water 1000 m unit: TBA 3000 m unit: TBA MATERIALS 1000 m unit: 6061-T6 Aluminum & Polyurethane 3000 m unit: Titanium, Polyurethane & 300 series stainless steel FINISH Hard Anodize 881L-GS Imaging 445-092 www.imagenex.com...
Page 52
IP Address* Option -020 Gyro Stabilization Option -048 *Note: Standard IP Address is 192.168.0.5 A different IP Address may be specified upon ordering. Product and company names listed are trademarks or trade names of their respective companies. 881L-GS Imaging 445-092 www.imagenex.com...