Connecting Cables
The Actiheart is supplied with a Micro USB cable of length 1.8m. Cables
exceeding 1.8m are not recommended.
The Actiheart patient lead is tested as part of the system and should not be
modified or extended.
WARNING: Use of cables and/or accessories other than those specified by
CamNtech may result in increased electromagnetic emissions or decreased
electromagnetic immunity and result in improper operation.
Use adjacent to Portable RF Communications equipment
WARNING: Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the Actiheart, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Electromagnetic Immunity
The Actiheart is intended for use in the electromagnetic environment specified below. The customer
or the user of the Actiheart should assure that it is used in such an environment.
Immunity
Test
Radiated Field
Immunity
Conducted RF
Immunity
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
The Actiheart is not mains powered hence no ac power line tests are applicable.
Guidance and Manufacturers Declaration – Electromagnetic Immunity
IEC 60601 Test
Level
IEC61000-4-3
IEC61000-4-3
IEC61000-4-6
IEC61000-4-6
IEC 61000-4-2
IEC 61000-4-2
Actiheart User Manual 5.1.31 (Jun 2022)
Compliance Level
10 V/m 80-2700 MHz
1 kHz 80% am
9-28 V/m 385-6000 MH
pulse mode
3 Vrms 0.15-80 MHz
1 kHz 80% am
6 Vrms 0.15-80 MHz
1 kHz 80% am
ISM/amateur radio band
+/-8kV Contact
+/-15kV Air
Electromagnetic
environment - guidance
74