Page 1
Instruction Manual Combining with 取扱説明書 Water Bath Rotary Evaporator SB-1300 230V This instruction manual is designed to use the product safely while keeping its best performance. Be sure to read “Safety precautions” before use. IMPORTANT Please keep this manual in a place easily accessible to every user.
Page 2
Important Safety Notice 1. Warning Signal Word To ensure the safety, this manual defines the The unit is not explosion-proof. Take extreme information on such matters as requiring particular care no to spill any liquid when using a care in the safety for each danger and attaches the flammable specimen or organic solvent.
Page 3
The warning labels should be read before using the product. * If warning labels become illegible due to damage on them, replace them with new ones. Contact us for spare labels. Water Bath (Model SB-1300) CAUTION Hot surface Do not move the unit When liquid...
N-1300 series and N-1200 series when set-up, operations, troubleshooting, maintenance, using this product. check-up, and disposal of SB-1300, which is water bath for rotary evaporator. Be sure to carefully read this manual and understand its description before using this product.
For safe operation This bath is intended for water only. Do not put liquid other than water (i.e. oil, etc.) into this bath. Warning Use of liquid other than water (pure water) may cause fire or failure. Do not put a cover on the bath. Caution If you use the bath with the cover, heat remains in the bath and it gets hot.
Preset temperature can be changed and operation can be stopped if needed as long as the operation is underway. 2-3 Specifications Product Name Constant temperature water bath (Water bath) Model SB-1300 Temperature set range Room temp.+ 5℃~90℃ ±1℃( during stirring ) Temperature Adjustment Accuracy※1 Temperature Set Accuracy ※1 ±5℃(...
2-4 Names of parts Bath Handle Bath cover Bath case Control panel Power switch Power supply inlet with Fuse(8A×2) Overheat preventive device reset button Communication connector ※ Power cord ※A special communication cable is used for connection with the vacuum control unit model NVC-3000. - 3 -...
3 Names and functions of control assembly 3-1 Control panel ①Power switch ②Display ③RUN lamp ④HEATING lamp ⑤RUN/STOP ⑥UP key ⑦DOWN key ⑧SET key Name Function ① Power switch Turning this switch ON will turn power ON. The presented measured temperature of the bath, or the set temperature in ②...
3-2 Alarm functions measures referring to P.14“Possible causes of troubles The product has the following safety functions. and solutions”. If you encounter with an abnormality, take appropriate Safety functions Safety unit Operation Causes How to recover The fuse will blow to shut off The power circuit was short-circuited See P15 “7-2 How to replace Fuse...
4 Installation 4-1 Installation environment Warning Never install the product in a potentially hazardous atmosphere. The product is not explosion-proof. Use in a potentially hazardous atmosphere may cause a fire or other accidents. Select a place that meets the conditions below for installing this product: ●...
4-2 Connecting utilities Warning Warning Do not use the branching socket or Confirm the voltage, phase, capacity, and the type of receptacle of power table tap. supply. Over-current may cause cable burn, fire. Wrong connection of power supply may cause fire or electric shock Warning Connect the ground wire correctly.
5 Operation 5-1 Preparation for operation Warning Do not put the liquid other than water into the bath. This product is intended for water (fresh water) only. If liquid other than water is used, it may cause fire or failure. ※Also refer to instruction manual for rotary evaporator (N-1300,N-1200 series).
5-2 Operating procedures Caution Caution Stop using the product if you find Do not touch the hot part during any trouble. use and for a while after use. When you face the trouble, turn off the power switch Bath and flask for specimen are in hot temperature immediately and refer to “6.Troubleshooting”on during use and for a while after use.
Page 14
(6) When the liquid temperature in the bath has been Power key RUN/STOP key stable, press the power key of the evaporator to turn it to ON, lower the jack slowly to soak the specimen flask in the liquid. Rotation adjusting knob Take extreme care not to make the liquid spill over.
5-3 How to set the alarm function You can set the upper temperature for the upper temperature limit alarm [A-0] or the power outage recovery function in the setting of the alarm function. ※You cannot set the sensor alarm [F-1] or the outside the measurable temperature range alarm [---]. *Shifting to the alarm function setting mode (1) Press the SET key longer in the measurement (current ℃...
Page 16
2.Setting the power outage recovery function You can set whether temperature control will be resumed when power is turned on again after power outage during temperature control by setting ON/OFF of the power outage recovery function. ・ When the power outage recovery function is ON…Temperature control resumes after power recovery in the state immediately before the recovery.
5-4 How to recover over heat protector Caution Hot surface In case that over heat protector does not recover even after pressing reset switch, the temperature in the bath is still hot. For preventing physical injury and etc., do not put your hands into the bath. Press the switch after the temperature falls.
Contact your dealer or the nearest service 6 Causes of troubles and solutions center for troubles not listed here. Symptom Possible causes Solutions The power plug has come off the outlet. Turn the power switch OFF and securely insert the Or is not inserted securely.
7 Maintenance・check-up 7-1 Cleaning and caring the product Warning Caution Never attempt to disassemble the Do not clean and care the product product. while it is still hot. The unit contains parts with high voltage applied or Cleaning and caring the product while it is may become hot, and disassembly may cause an still hot may burn yourself.
Principal components parts and disposal method External dimensions (Including Components Model Weight How to discard handle) Approx.3.9 Request the disposal operator for SB-1300 285(W)×295(D)×244(H) Main body disposal. Components Major materials Heater integrated bath Stainless steel, aluminum, ceramics Bath cover PET(containing glass fibers)...
9 After-sale Services 1. In case the product does not function satisfactorily, 3. Repair during the guarantee period will be made check first referring page according to the guarantee stipulations. troubleshooting to see if this is actually a trouble. 4. After expiration of the guarantee period, the 2.
List of consumable & replacement parts/optional parts 10-1 Consumable & replacement parts 230V Power Cord C Type Time-delay Fuse 230V Power Cord B Type 230V Power Cord O Type Code No. Std. Code No. Std. Code No. Std. Code No. Std.
Need help?
Do you have a question about the SB-1300 and is the answer not in the manual?
Questions and answers