HI 739
Fluoride High Range
w w w . h a n n a i n s t . c o m
sale@ponpe.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

Preliminary examination:

Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 739 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 1 bottle (30 mL) of HI 739AS Fluoride HR Reagent A
• 1 bottle (30 mL) of HI 739BS Fluoride HR Reagent B
• 2 x 1 mL syringe with tip
• 1 dropper
• 1 x 1.5V AAA Battery
• Instruction Manual
For more details about spare parts and accessories
see "Accessories".

Technical specifications:

Range
0.0 to 20.0 ppm
Resolution
0.1 ppm
Accuracy
±0.5 ppm ±5 % of reading @ 25 °C / 77 °F
Typical EMC Dev.
±0.1 ppm
Light Source
Light Emitting Diode @ 575 nm
Light Detector
Silicon Photocell
Adaptation of the Standard Methods for the Examination
Method
of Water and Wastewater, 18
Environment
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type
1 x 1.5V AAA
Auto-Shut off
After 10 minutes of non-use
Dimensions
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
Weight
64 g (2.25 oz.)

Functional description:

Errors and warnings:

th
edition , SPADNS method.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking "0.00" indicates that the sample absorbs
more light than the zero reference. Check the procedure and make
sure you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
www.ponpe.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 739 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 739

  • Page 1 2. Cuvette with cap. any damage occurred during shipment, please notify your Dealer. 3. Cuvette holder. Each HI 739 meter is supplied complete with: 4. Liquid Crystal Display. • Two Sample Cuvettes and Caps • 1 bottle (30 mL) of HI 739AS Fluoride HR Reagent A 5.
  • Page 2 • Turn the instrument upside down and remove the battery cover with a screwdriver. • Remove the cuvette from the meter and Hanna Instruments reserves the right to modify the design, construction or unscrew the cap. Using the unused syringe, appearance of its products without advance notice.

Table of Contents