Hanna Instruments HI 739 Quick Reference Manual
Hanna Instruments HI 739 Quick Reference Manual

Hanna Instruments HI 739 Quick Reference Manual

Fluoride high range

Advertisement

HI 739 Fluoride High Range
Quick Reference Guide

TECHNICAL SPECIFICATIONS

Range:
0.0 to 20.0 ppm
Resolution:
0.1 ppm
±0.5 ppm ±5 % of reading @ 25 °C / 77 °F
Accuracy:
Typical EMC Dev.:
±0.1 ppm
Light Source:
Light Emitting Diode @ 575 nm
Light Detector:
Silicon Photocell
Method:
Adaptation of the Standard Methods for the Examination of
Water and Wastewater, 18
Environmental:
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type:
1 x 1.5V AAA
Auto-Shut off:
After 10 minutes of non-use
Dimensions:
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5 ")
64 g (2.25 oz.)
Weight:

TIPS FOR AN ACCURATE MEASUREMENT

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and
completely free of fingerprints, oil and dirt. Wipe it thoroughly with HI 731318 or a lint-
free cloth prior to insertion.
• Shaking the cuvette can generate bubbles, causing higher readings. To obtain accurate
measurements, remove bubbles by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent is added, as accuracy will
be affected.
• After the reading it is important to immediately discard the sample, otherwise the glass
might become permanently stained.
w w w . h a n n a i n s t . c o m
th
edition , SPADNS method.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 739 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 739

  • Page 1 HI 739 Fluoride High Range Quick Reference Guide TECHNICAL SPECIFICATIONS • Range: 0.0 to 20.0 ppm • Resolution: 0.1 ppm • ±0.5 ppm ±5 % of reading @ 25 °C / 77 °F Accuracy: • Typical EMC Dev.: ±0.1 ppm •...
  • Page 2 TAKING MEASUREMENTS • Turn the meter on by pressing the button, all segments will be displayed. When the display shows “Add”, “C.1” with “Press” blinking, the meter is ready. • Using supplied syringe add 2 mL (1 mL twice) of HI 739AS reagent to the cuvette and use the pipette to fill the cuvette up to the mark with HI 739BS reagent.

Table of Contents