Step 1: Selection of measurement point & pipe preparation
Avoid installation of sensors in the vicinity of deformations and
pipe defects, near welding seams or where deposits could have
accumulated.
Select a measuring point with sufficient straight pipe to obtain
accurate measurements, generally a length of ten times the
pipe diameter on the inlet of the measuring point, and a five
times pipe diameter on the outlet.
For a horizontal pipe, mount the sensors on the side of the
pipe. For a vertical pipe, mount the sensors at a location where
the liquid flows upwards. (Pic 1)
Mount the sensors in the direction of the flow. (Pic 2)
Clean the pipe at the measuring point. Remove loose paint and
rust with a wire brush or file.
Apply coupling paste to the bottom of the clamp-on sensors
before installation. Leave straps slightly loose so sensors can
be moved along the pipe during the positioning process.
Pic 1: Mounting points
Step 2: Keyboard Familiarisation
Step 3: Quick Start Menu and Setup Wizard
After the provisional installation of the sensors, connect the trans-
mitter to a power source. You can then enter the parameters into
the Setup Wizard to prepare the flow meter for measurement.
On subsequent restoration of power the flow meter will be in
measurement mode. Press <ESC> to return to the main menu.
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2010. I Issue: KF150QuickStartV22E0710
KATflow 150 Quick Start Manual
Flow in pipe
Pic 2: Flow direction
Show NEXT (1) available item
Q
(2) = Start totaliser function
ON
Show next DISPlay (3)
Q
(8) = Stop totaliser function
OFF
DIRECT (9) access to trend plot
Move menu / selection item UP
Move menu / selection item DOWN
ESCape entry without saving
ENTER selection with saving
Upon first power on and the
boot-up sequence, the main
menu is displayed. Use <UP>
and <DOWN> cursor keys to
select Quick start. Confirm by
pressing <ENTER>.
Continue on page 2.
1 / 3
Use the cursor keys to select
Setup Wizard Single for single
channel measurement (one set
of clamp-on sensors).
Select Setup Wizard Dual, for
dual channel measurement (two
sets of clamp-on sensors).
Confirm selection by pressing
<ENTER>.
If the sensors are recognised,
the serial number will be shown.
If they are not recognised or not
connected, they may be selec-
ted from a list.
Select the units of measurement
using cursor keys and confirm
by pressing <ENTER>.
Choose the pipe material using
cursor
keys
and
confirm
selection by pressing <ENTER>.
Enter outside pipe diameter
using alphanumerical keys and
confirm by pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.
If "0" is entered, an additional
screen appears that allows entry
of the pipe circumference. Press
<ENTER> to confirm.
Enter pipe wall thickness using
alphanumerical keys and con-
firm by pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.
Select fluid using cursor keys.
Confirm by pressing <ENTER>.
Enter process temperature using
alphanumerical keys. Confirm by
pressing <ENTER>.
Use <UP> key as a backspace
to correct for data entry errors.

Advertisement

loading

Summary of Contents for Katronic KATflow 150

  • Page 1 KATflow 150 Quick Start Manual Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I info@katronic.co.uk I www.katronic.co.uk Step 1: Selection of measurement point & pipe preparation Use the cursor keys to select...
  • Page 2 KATflow 150 Quick Start Manual Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I info@katronic.co.uk I www.katronic.co.uk Step 3: Quick Start Function and Setup Wizard (Continued) Sensor Positioning Screen Select pipe lining material using Mount the transducers with the suggested spacing.
  • Page 3 KATflow 150 Quick Start Manual Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I info@katronic.co.uk I www.katronic.co.uk RS 485 Serial Interface Heat Quantity Measurement (HQM), optional The RS 485 interface is used for connecting up to 32 flow meters Where equipped, heat quantity (energy) and heat flow (energy to a computer system.