en - Areas of application
Installing the ad-
apters
32

Installation

Adapters from the APWD 325 set are required to reprocess the hoses
used in dental nitrous oxide sedation systems. The adapters are suit-
able for all commercially available double hoses.
Screw the adapters into the connections on the injector manifold.
Secure the adapters using the two  knurled nuts and tighten them
at the same time.
Additional adapters must be installed if the APWD 328 extension set
is used. The adapters are available as retrofit set
APWD 326.
When the adapters are used, the wash water must be filtered. The
injector manifold must be fitted with an A 800 tubular filter. The
wash water consists of the inflowing water, the process chemicals
and the soiling.
The tubular filter is supplied with its own operating instructions.
Follow in particular the instructions on how to perform regular
cleaning of the tubular filter.