8� Command Modes; Using Magnetic Pen And Pin Cal - gefran HMF Series Instruction Manual

Smart hart melt pressure transmitters
Table of Contents

Advertisement

The command modes must be used by the user only to make periodical recalibrations of the system in the
process, and are made through:
• Magnetic sensor/Pin CAL
• HART Communication
The event combinations give the following functions:
1.
Autozero
2.
Calibration (CAL)
3.
Reset autozero
1) A
utozero
Application mode
The Autozero function is activated
by :
1) positioning the magnet near the
Autozero label on the shell of the
sensor.
2) Short-circuiting the pin E-F
(external Autozero version).
The magnet has to be maintained
on the Autozero position for a time
within 2 to 5 seconds.
NOTE: For all transmitters with current outputs, during the Autozero phase, output can rise up to 7 mA.
That's a short variation only visible during the Autozero phase; it won't have any effects on the final signal.
Magnetic
2...5 seconds
pen
Warning: the periodical zero ricalibration through the Autozero function must be done at stable tempe-
rature and no pressure in the extrusion chamber.
80636B_MAN_HMF - HWF_05-2022_ENG
8� Command modes
8�1� Using Magnetic pen and Pin CAL
Limits
FS (bar)
% adjustment
≤ 35
> 35, < 100
> 100, < 200
≥ 200
Remove
100
The Autozero effect will be visible after
waiting 2 seconds starting from the fun-
40
ction removal.
20
The precision of the zero value will be
10
defined by the accuracy class of the
sensor.
1) Machine pressure = 0 bar and transducer
powered.
2) Put magnetic pen on marked spot
(2...5 seconds).
3) Remove magnetic pen.
4) Ready !