Attaching the Sensor Mat
The Solo-Call Transmitter will accept any Lifemax Sensor Mat. To attach the Sensor
Mat, plug the RJ9 connector into the input socket.
Setting the Detection Mode and Delay Setting
The Solo-Call system can be used with a chair or bed sensor mat. Commonly, this will
alarm when pressure is relieved from the mat as the patient leaves the chair/bed.
In some environments and to prolong the life of your sensor mat, it may be more
appropriate to use your mat as a floor sensor. Floor mats are commonly used to alarm
when pressure is applied to them, detecting when a patient moves through a certain
area.
In some instances, it is preferable to allow a small delay before the Solo-Call transmitter
is triggered. This can reduce false alarms for some applications, such as when a patient
moves in their bed or chair. The Solo-Call system offers the option of an immediate
trigger or a 2 second delay.
The Solo-Call can be switched between detection modes and delay settings as follows:
Remove the battery cover on the Pager unit by depressing the lock pin and
1.
sliding the cover downwards;
2.
Turn the transmitter unit on, by sliding the power switch downwards;
3.
Turn the pager unit on, by sliding the multi-position switch downwards to
position 1;
With the battery compartment facing upwards, switch the mode button to
4.
either the chair/bed or floor setting;
5.
Switch the delay button to either 0s (immediate) or 2s (2 second delay);
6.
Replace the battery cover on each unit and slide upwards until the lock pin
locates.
Mounting the Solo-Call Units
The Solo-Call Transmitter can be mounted using the screw slot in the rear of the unit
or by fitting the mounting strap, as described below. To maximise the wireless range,
it is recommended that the transmitter is mounted as high as possible and clear of
objects that may restrict a signal.
The Solo-Call Pager also features a screw slot or can be fitted with a belt clip/wall
mounting bracket, as described below.
To fit either the mounting strap, belt clip or wall mounting bracket:
1.
Place the transmitter or pager unit with the battery compartment facing
upwards;
2.
Insert the peg of the mounting strap or belt clip into the large hole of the
screw slot;
3.
Gently slide upwards and into place.
1510 Wireless Solo Call Instructions_160927.indd 4
Peg
27/09/2016 15:00:59