7. Installation of pump casing (101)
− Insert an O-ring (412) into the groove of the shaft
sealing casing (441).
− Adjust the pump casing (101) (inlet/outlet nozzle
opposite the motor feet).
− Force the pump casing (101) into the shaft sealing
casing (441).
− SM55/75B only: Fasten the pump casing (101) to the
shaft sealing casing (441) using hexagon head screws
(901/550.1) and hexagon nuts (920.1/550.1).
− SM100/200/250B only: Fasten the pump casing (101)
to the shaft sealing casing (441) using hexagon head
screws (901/550.1).
25
255548 Rev.1