Removal Of Old Panel; Remove Ohmeda Front Panel; Installing New Panel; Installing Retrofit Kit - BeaconMedaes 4107 9027 47.00 Installation, Operation And Maintenance Instructions

Table of Contents

Advertisement

Wall Mount Gas Control Panel
2.

Removal of Old Panel

CAUTION:
Prior to removal and installation of control
panels, verify zone valve is shut off and line to
control panel is not pressurized.
2.1

Remove Ohmeda Front Panel

1. Prior to removal and installation of control
panels, verify zone valve is shut off and line
to control panel is not pressurized.
2. Loosen set screws of on-off valve handle and
remove handle.
3. Remove regulator adjusting screw and knob
by unscrewing from regulator.
4. Unscrew lens covers from gauges and
remove.
5. Remove screws from cover plate. Remove
cover plate. Discard
6. Disconnect both tubing connections at top of
box.
7. Remove screws in mounting brackets and
remove control panel manifold from box.
Discard.
Mounting
Bracket
Tubing
Connections
On-Off Valve Handle
4107 9027 47.00
Regulator
Cover Plate
Figure 1
3.

Installing New Panel

3.1
Installing Retrofi t Kit
1. Remove fl are nut assembly (supplied with
retrofi t kit) from polyethylene bag and install
to fi ttings at top of rough-in box.
Regulator
Adjusting Screw
and Knob
2. Remove protective dust caps from hose
ends. Insert hose assembly end into left inlet
connector inside mounting box. Push hose
adapter through collet into connector until it
bottoms out. Pull back on the hose assembly
to verify proper attachment.
Gauge Lens
Covers
2
Figure 2
Figure 3
Figure 4