Hansen Technologies HSC4H-N10K Specifications, Applications, Service Instructions & Parts page 4

Gas detection sensors, monitors & alert systems
Table of Contents

Advertisement

OPERATING TEMPERATURE
RANGE
OPERATING CURRENT @ 24V DC
OPERATING CURRENT @ 24V AC
VOLTAGE REQUIREMENTS
HUMIDIT Y RANGE
(NON-CONDENSING)
MA X LOOP RESISTANCE
OUTPUT
4-20mA SIGNAL OUTPUT
ALARM RELAY
TRUE ALTITUDE
POLLUTION DEGREE
EQUIPMENT CLASS
INSTALLATION CATEGORY
DUT Y CYCLE
*Approved Class2/SELV; -5%/+10% voltage tolerance. AC or DC voltage user selectable. See pages 6 and 7 for
configuration.
DO NOT SUBSTITUTE OR MODIFY
Because of the danger of introducing additional hazards,
do not install substitute parts or perform unauthorized
modifications to the equipment listed herein.
A100g
JULY 2017
GAS SENSOR SPECIFICATIONS
BASIC
STANDARD
0ºF TO 122ºF
0ºF TO 122ºF
(-17ºC TO 50ºC)
(-17ºC TO 50ºC)
230mA /6W
50mA /1W
460mA /11VA
100mA /2VA
2
3
TABLE 1
HARSH
VENT LINE
-40ºF TO 105ºF
-40ºF TO 122ºF
(-40ºC TO 40ºC)
(-40ºC TO 50ºC)
280mA /7W
280mA /7W
560mA /13VA
560mA /13VA
12-24V AC/DC* (-15%/+10%)
0-95%
1000 OHMS
4-20mA, 0-5V, 1-5V, 0-10V, 2-10V DC
Isolated, 2 wire
SPDT, 1 A, 120V
2000M
4
3
1
100%
CLEANING/DECONTAMINATION
EXTERNAL SURFACE
The external surface, excluding the sensing element,
can be cleaned using light detergent and water. Do not
allow any of the solution to come in contact with the
sensing element sintered surface.
4
EXPLOSION
EXTREME
PROOF
-40ºF TO 122ºF
-60ºF TO 105ºF
(-40ºC TO 50ºC)
(-51ºC TO 40ºC)
280mA /7W
1900mA /46W
560mA /13VA
2100mA /50VA
24V AC/DC*
0-100%

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hec4h-n250Hec4-n1k

Table of Contents