Monitor Operation
Programming Delay, Failsafe, and Latch functions directly on the
E
Point Network transmitter is NOT recommended, because local
3
settings will override network settings. These functions should
typically be programmed directly from the Network Control device.

Network Model Buzzer Menu

The Buzzer menu allows users to select whether or not the built-in audible
alarm can be silenced by pressing a keypad button and which events will
activate the buzzer.
Press enter in the main Buzzer menu screen. The Buzzer screen
displays either Activate or Silence.
Use the arrow keys to scroll to the desired option and press enter
to select.
If Silence is selected, the buzzer can be deactivated (without affecting the
event) by pressing and holding any key on the monitor for 5 seconds.
Select Activate to display available menu options:
Network:
If the monitor is configured with the Modbus or BACnet
communication protocol, these options are replaced with
Network. No action can be taken from this option: it simply
informs you that the buzzer is controlled by the network
master device.
Disable:
Only available if the monitor is configured to emulate a 201T
(previous generation of Honeywell Analytics gas monitors for
toxic and combustible gases).
When the E
controller device controls the relay and buzzer together; they cannot be
controlled separately.
28
* Menu *
Buzzer
Buzzer
Activate
Point Network Model is configured in 201T EMU mode, the
3
E
Point Gas Monitor Technical Manual