5.1 Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after
each use. The procedure is as follows:
A) EUP-O732T probe
1) Remove the protective cover.
2) Clean the probe of all patient's blood or fluid with running
tap water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B) Attachment
1) Remove the Attachment and the magnetic sensor from the probe.
2) Immerse the Attachment in sufficient amount of high quality
tap water. Scrub it using soft cloth to remove all visible soil
and from its surface.

5.2 Containment and transportation

Putting
the
exclusive shock and damage proof container for
transportation is recommended. It is recommended that instruments
are reprocessed as soon as possible and not later than 4 hours
after usage.

5.3 Manual Cleaning and disinfection

Prepare following items before manual
cleaning and disinfection:
a) Detergent:
Johnson, #2258) or another cleaning
agent
with
compatibility
device.
b) Disinfectant: Cidex® OPA (Johnson &
Johnson,
disinfectant with approved material
compatibility
device.
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water
(sufficient size for immersion of the submergible part of the
probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal
protective
protective skirt, face protection mask or protective glasses,
see also instructions of the manufacturer for the detergent and
the disinfectant)
contaminated
Cidezyme®
approved
for
this
#
20391)
or
for
this
equipment
equipment
(Johnson
&
material
medical
another
medical
(gloves,
-15-
Point of use
(Pre-cleaning)
Containment and
into
transportation
Manual cleaning
and disinfection
Manual Cleaning
Rinsing after manual
cleaning
Manual Disinfection
Rinsing after manual
disinfection
water
repellent
Q1E-EP1234