5. Cleaning and Disinfection and Sterilization

The probe and accessory must be reprocessed after each use.
Refer to the reprocessing instruction in this chapter.
WARNINGS
Limitations on
reprocessing
Transportation
before using
The probe is delivered unsterile. Prior to the
first use, reprocess the probe.
Temperature should not exceed 60°C during
reprocessing.
Probe connector is not water resistant.
The probe is not completely submersible. The
immersible part is shown in Fig.1. The
un-immersible part should be disinfected by wipe
disinfection.
The probe should be packed in a sterile pouch
or container to transport from Central Sterile
Supply Department (CSSD) to an operating room.
Be careful not to damage the sterile pouch or
container during transportation.
-12-