Example of Incorrect Use
Waveform: 1022 points
Marker range: 110–1022
Marker polarity: Negative
This figure shows that a negative polarity marker goes low during
the marker on points; the marker signal goes high during the off
points. The ALC samples the waveform during the off marker
points.
Sampling both on and off time sets the modulator circuitry
incorrectly for higher signal levels. Note the increased amplitude
at the beginning of the pulse.

Accessing Marker Utilities

Use the following procedure to display the marker parameters. This procedure uses the Dual ARB
player, but you can access the marker utilities through the
1. Select the ARB waveform player:
press
>
Mode
Dual ARB
2. Press
>
ARB Setup
Marker Utilities
Chapter 3
.
Marker
Marker On
Sample range begins on first point of signal
Marker On
Negative range set between signal and
off time
softkey in all ARB formats.
ARB Setup
Basic Digital Operation
Using Waveform Markers
Marker On
Off
Marker
Marker
On
Off
101