Summary of Contents for Hanna Instruments HI9064

  • Page 1 Instruction Manual HI 9064 • HI 9065 Professional Water-Resistant Hygrometers w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    Dear Customer, PRELIMINARY EXAMINATION Thank you for choosing a HANNA instruments ® product. Please read this instruction manual carefully before using the instru- Remove the instrument from the packing material and examine it to ment. make sure that no damage has occurred during shipping. If there is any damage, immediately notify your dealer.
  • Page 3: Functional Description

    FUNCTIONAL DESCRIPTION SPECIFICATIONS HI 9064 HI 9065 Range 5.0 to 95.0% RH H °C 0.0 to 60.0°C / 32 to 140°F Resolution 0.1% RH 0.1°C / 1°F Accuracy ±2% RH ±0.4°C / ±1 °F H °C (for 1 year, excluding probe error) HI 9065 HI 9064 Thermo...
  • Page 4: Operational Guide

    TEMPERATURE MEASUREMENTS (HI 9065 ONLY) OPERATIONAL GUIDE Connect the temperature probe (HI 765 series, see “Accessories”) to the socket on the top of the meter and the unit will automatically INITIAL PREPARATION display the temperature reading. Each meter is supplied complete with four 1.5V Note: HI 9065 can be used with just one probe (RH or temperature) AA batteries.
  • Page 5: Calibration

    CALIBRATION • Pour approximately 12 cc of distilled water into the other chamber marked "RH 75.4%". All HANNA hygrometers have been precalibrated at the factory. HANNA instruments ® uses state-of-the-art thermal humidity cham- Na Cl Li Cl 75.4 % 12 cc 11.1 %...
  • Page 6: Battery Replacement

    • Remove the probe and tightly seal the chamber containing the BATTERY REPLACEMENT LiCl solution. HI 9064 and HI 9065 are provided with the advanced BEPS • Remove the cap from the "RH75.4%" cham- (Battery Error Prevention System) technology that detects a low power ber containing the NaCl solution and insert condition and automatically turns the instruments off to avoid Na Cl...
  • Page 7: Accessories

    HI 765W Wire temperature probe, whitout handle, for hard-to- reach places Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
  • Page 8: Ce Declaration Of Conformity

    CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the envi- ronment in which they are used. Operation of these instruments in residential area could cause unacceptable inter- ferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences.
  • Page 9 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.

This manual is also suitable for: