Hanna Instruments HI 9026V Instruction Manual
Hanna Instruments HI 9026V Instruction Manual

Hanna Instruments HI 9026V Instruction Manual

Complete portable kit for ph, mv & temperature measurements in wine and must with calibration-check

Advertisement

Quick Links

Instruction Manual
HI 9026V
Complete Portable Kit
for pH, mV & Temperature
Measurements in
Wine and Must
with Calibration-Check
TM
w w w . h a n n a i n s t . c o m
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 9026V and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 9026V

  • Page 1 Instruction Manual HI 9026V Complete Portable Kit for pH, mV & Temperature Measurements in Wine and Must with Calibration-Check w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    • HI 7662 stainless steel temperature probe • pH 3.00 and pH 7.01 buffer solutions, 20 mL each All Hanna Instruments meters are warranted for two years against defects in workmanship • HI 700635 and HI 700636 specific cleaning solutions, 20 mL each and materials when used for their intended purpose and maintained according to instructions.
  • Page 3: Functional Description

    FUNCTIONAL DESCRIPTION SPECIFICATIONS Range -2.00 to 16.00 pH ±699.9 mV / ±1999 mV -20.0 to 120.0 °C / -4.0 to 248.0 °F Resolution 0.01 pH / 0.1 mV / 1 mV / 0.1 °C / 0.1 °F Accuracy ±0.01 pH / ±0.2 mV / ±1 mV ±0.4 °C / ±0.8 °F (excluding probe error) Typical EMC Deviation ±0.02 pH...
  • Page 4: Operational Guide

    The pH reading is directly affected by temperature. In order for the meter to OPERATIONAL GUIDE measure the pH accurately, temperature must be taken in consideration. If the sample temperature is quite different from the temperature at which the pH electrode was kept, allow a few minutes for a perfect thermal equilibrium INITIAL PREPARATION between them.
  • Page 5: Setup Menu

    Notes: SETUP MENU • When reading is out of range, the display will flash the closest full-scale value. • If using pH electrode while in mV mode, the meter will measure the mV The instrument allows the user to program several parameters though a Setup generated by the pH electrode.
  • Page 6: Calibration

    If you’re going to calibrate with a different buffer, select the desired value CALIBRATION by pressing the arrow keys. • Submerge the electrode approximately 4 cm (1½") into the second buffer solution, place the temperature probe as close as possible to the electrode HI 9026 is factory calibrated for ORP and temperature ranges.
  • Page 7 After calibrating the first point (see above), press the CAL key to end the CONDITION calibration procedure. The display is provided with a 5-dot bargraph (unless the feature is disabled) Note: With one-point calibration there is not “Condition” and only the frame is which gives indications about the electrode status after calibration as follows: shown.
  • Page 8: Ph Buffer Temperature Dependence

    pH BUFFER TEMPERATURE DEPENDENCE LCD MESSAGE GUIDE pH readings are strongly affected by temperature. The calibration buffer TAGS & SYMBOLS solutions are affected by temperature changes to a lesser degree than normal solutions. For manual temperature calibration please refer to the following chart. TEMP pH BUFFER VALUES °C °F...
  • Page 9: Troubleshooting Guide

    TROUBLESHOOTING GUIDE pH ELECTRODE MAINTENANCE • Blinking full scale value: reading is out of range. Note: To prevent damage to the electrode, remove the pH electrode from the sample before turning the meter off. With the meter OFF, detach the •...
  • Page 10: Accessories

    Cleaning procedure ACCESSORIES After use, clean the electrode by immersing it in the proper cleaning solution (HI 700635 or HI 700636) for 5 minutes. PROBES If case of slow and unreliable readings, it is recommended to perform the HI 1048B Glass body, double junction, refillable pH electrode with following procedure for cleaning the junction: BNC connector and 1 m (3.3') cable...
  • Page 11 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.