Recommended separation distances between portable and mobile RF communications equipment and the AC-
300MD
The AC-300MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the AC-300MD can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the AC-300MD as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Caution
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital regarding
environmental pollution.
Warning on power connection for medical use
Please use the following power supply cord. With
connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord that
is approved for use in your area.
United States and Canada
Plug Type
HOSPITAL GRADE*
Cord Type
Min. Type SJT
Min. 18 AWG
Minimum Rating
10A/125V
for Plug and
Appliance
Couplers
Safety Approval
UL Listed and CSA
*Note: Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked 'Hospital Only' or
'Hospital Grade'.

Warning on power connection

Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts that
conforms to the safety regulations of each country if
applicable.
GB
6
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2
d = 1.2
0.12
0.38
1.2
3.8
12
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power
Cord / Appliance Connector / Plug, please consult a
qualified service personnel.
Caution
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as vases,
shall be placed on the apparatus.
IMPORTANT
The nameplate is located on the bottom.
For the customers in Europe
This product has been manufactured by or on behalf of
Sony Corporation, 1-7-1 Konan Minato-ku Tokyo, 108-
0075 Japan.
Inquiries related to product compliance based on European
Union legislation shall be addressed to the authorized
representative, Sony Deutschland GmbH, Hedelfinger
Strasse 61, 70327 Stuttgart, Germany.
For any service or guarantee matters, please refer to the
addresses provided in the separate service or guarantee
documents.
m
800 MHz to 2.5 GHz
d = 2.3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23