Error And Troubleshooting - Kanomax TABmaster 6715 Instruction Manual

Hide thumbs Also See for TABmaster 6715:
Table of Contents

Advertisement

7 Error and Troubleshooting

No display when power On
Display
Display
Press
Testing record can not be saved
New ID can not be added
Printing unreadable code
www.lazodecontrol.com
Symptom
incorrect specification of the AC
Adapter
Battery installation error
Low or dead battery power
Connection cable loss with the
base
Back-Pressure
incorrect
Low battery charge
but no response
Back Pressure Plate opened On
Over-limit Cycle quantity in
current ID
ID over-limit
Current ID is without Cycle
Unmatched printer
Wrong baud rate setting
Possible causes
mode
using
23
+51 960 810 089
Corrective action
Check and refer to Chapter4.1
in Manuel
Check and refer to Chapter4.1
in Manuel
Replace the batteries
Reseat the connect cable
Refer to Chapter 5.4 in Manuel
Replace new batteries
Shut off the plate
Add new ID or delete some
data
Delete some ID
Using the current ID
Using our special optional
printer supplied by us.
Reset referring to the Printer
Manuel
ventas@lazodecontrol.com

Advertisement

Table of Contents
loading

Table of Contents