Page 1
VECTECH 496 ELECTROSTSTIC FIELDMETER OPERATION MANUAL Thank you for purchasing our electrostatic field meter. Please carefully read this manual before operating the system. Store this manual in a safe, easily accessible place for future reference.
Page 2
Note: * Please accord to the “note” for preventing the meter from damage. * Please not hit the sensor part. * Please not wet the meter and not use the meter with wet hands. * Please use the original parts when replacing the fittings. * Please turn off the power supply when finishing using and take out the battery when not using in a period of time.
2 The meter cannot run when the voltage is lower. And in the 2.Product Instruction and Parts testing process, the meter sounds firstly and turns off Product instruction automatically. At the moment, it must replace a new battery. 3 Replace a new battery: move the lid and take out the valid battery.
6 Press “HOLD/MAX” key into hold process or max process Display on the LCD when it is necessary to keep the testing value or maximal value. 7 Press “POWER” key to turn off the meter when finishing the test. 0 - 2 2 0 V 0 - 2 2 0 V Note: 0 - 1 .
2.3 Parts meter includes following parts as the table. Please check all the parts are well and if the parts are damaged, please contact with the agents or our company. Parts Quantity Electrostatic field meter Testing plate of the ion balance Grounding line 9VDC battery Operation Manual...
Page 6
±10% 5 the Static Eliminator device such as ionic air blower and keep Testing precise: some distance to test the ion balance. Waiting current: <3uA When it is necessary to keep the testing value or maximal Working current: <15mA value. Surroundings: 0℃~50℃, humidity<80% 6 Press “power”...
Page 7
Mode1 Mode 2 Mode 3 Mode 4 th e te s te d 2 5 m m Pressing o b je c t “power” 1s-2.5s 2.5s-3.5s 3.5s-4.5s 4.5s-5s " + " g r a p h e m ic time display display display...
Page 8
MAX: keep the current maximal value until testing more bigger 4.2 Zero value. In max process, lock the functions of “STATIC / IB” and “ZERO” until exiting the MAX process. Note: ENTER: press “HOLD/MAX” key until display “MAX” 1) Ground of the meter: default 0 of the meter *The grounding line is “Ground of the meter”...
Page 9
zeroing successfully. power supply. * Grounding first and then press “ZERO” key about two 2 press “STATIC/IB” key to the ion balance testing (display “IB”). seconds and avoid objects with electricity to zero. After 3 press “STATIC/IB” key again to return to the static testing. un-pressing the key, the meter sounds and displays 0 if zeroing successfully.