Functional Grounding And Product Safety - Gamry Instruments IMX8 Operator's Manual

Hide thumbs Also See for IMX8:
Table of Contents

Advertisement

This binding post is not intended for any use other than connecting the System Potentiostat to a functional
earth ground to improve shielding against noise. Connecting this binding post to a hazardous voltage can create
a significant safety hazard.
Warning:
earth ground. This is especially dangerous if the cell can source significant power (e.g., a battery, fuel cell,
or capacitor). The cell can discharge through the IMX8 and/or System Potentiostat, possibly causing severe
damage to the IMX8 or System Potentiostat.

Functional Grounding and Product Safety

As defined in IEC Publication 348, Safety Requirements for Electronic Measuring Apparatus, the IMX8 is a
Class I grounded device.
When you use the IMX8 with an approved line cord, the connection to the functional earth ground is
automatically made prior to making any power connections. The USB connection can also sometimes connect
the IMX8 to functional ground if the computer is earth grounded.
Warning:
use the IMX8 with a two-wire extension cord, with an adapter that does not provide for earth grounding, or
with an electrical outlet that is not properly wired with an earth ground. If the earth ground is not properly
connected, it creates a safety hazard, which could result in personal injury or death.
The IMX8 is supplied with a line cord suitable for use in the United States. In other countries, you may have to
replace the line cord with one suitable for your type of electrical outlet. You must always use a line cord with a
CEE 22 Standard V female connector on the instrument end of the cable. This is the same connector used on
the US standard line cord supplied with your IMX8.
Warning:
that supplied with the IMX8. An improper line cord can create a safety hazard, which could result in
personal injury or death.
The wiring polarity of a properly wired connector is shown in Table 1-2 for both US line cords and European
line cords that follow the "harmonized" wiring convention.
Safety Considerations
Do not earth ground the System Potentiostat if your electrochemical cell has
Do not negate the protection of the IMX8's earth ground by any means. Do not
If you replace the line cord, you must use a line cord with the same polarity as
9