How To Use Ace-1000 - ACE INSTRUMENTS ACE-1000 Instruction Manual

Vw data recorder
Table of Contents

Advertisement

3-2 MENU Description (Applying display units by changes in Mode)
LCD light contrast
Change of Sweep mode
Applying display unit by
Changes in Mode

3-3 How to use ACE-1000

2.
5
In principle LCD adjust the contrast. When a product is shipped, LCD contrast is set at "10"
(25 as a standard). If the area is hot or it is summer time, set at "5∼7"(about 30∼35℃).
If it is cold or winter time, set at "12∼15"(about 0∼15℃). The contrast can be adjusted
from 1 to 20.
Adjust the frequency range of the unit to come into the Sweep Mode, and change the
sweep mode.
As the below table, apply the measuring unit, depending on the change in the Display
Mode. VW ACE-1000 readout unit shows frequency (Hz) always, and simultaneously the
chosen display unit is shown as below.
MODE
Measuring Unit
Mode 1
Period(μsec)
3
2
Mode 2
10
Hz
Strain(με0.391)
Mode 3
Strain(με0.7756)
Mode 4
Strain(με3.304)
Mode 5
Strain(με4.062)
Mode 6
Reference
As A1000Pro saves only frequency (Hz), the saved data should be converted by it if you
use other units.
Model ACE-1000 is effectively used as following.
 Connect a jumper cable to the connector.
 Press [ON] button.
(After popping up ACE's logo, it converts into the Main screen.)
 Check out the sweep and unit, and press [MENU], [ENTER],
[▲/LIGHT], [▼/SAVE] to set the basic menu.
(See [3-1] for detailed specifications and [3-2] for MENU)
 Connect with a signal cable by the clip of the jumper cable.
(See [3-1] for detailed specifications)
 Record or save the output frequency.
 Get rid of the clip and turn off power.
 Download the data by connecting RS-232 cable to the computer.
3. Description for Using
Main display
Applying sensors
μs
All VW Sensors
3
2
10
Hz
VW Load Cells
VW Spot-weldable Strain
0.391
Gage
0.7756
VW Shotcrete Strain Gage
VW Embedment Strain
3.304
Gage
4.062
VW Weldable Strain Gage
- 7

Advertisement