Description For Using; Detail Description For The Function - ACE INSTRUMENTS ACE-1000 Instruction Manual

Vw data recorder
Table of Contents

Advertisement

3-1 Detail description for the function

[MENU], [ENTER]
[▲/LIGHT]
■ Setting the Sweep and Measuring Unit
Like the below table it is recommendable to change them by using [MENU], [ENTER] and
[▲/LIGHT] buttons, depending on the situations on the site.
[MENU]
[ENTER]
-
-
-
-
1 Click
-
-
-
-
-
-
-
-
1 Click
-
-
-
1 Click
-
-
-
-
-
-
-
-
-
-
-
1 Click
-
2 Click
-
3 Click
Reference
(1) Temperature sensor
[RTD] indicates the resistance temperature device when the temperature sensor type is
specified. The resistance value changes according to the temperature change is positive
(+), [NTC] indicates the thermistor, and the resistance value change according to the
temperature change is negative (-).
(2) LCD Contrast
When the brightness of the LCD changes according to the weather, use the arrow keys to
adjust the brightness appropriately.
(3) POWER OFF TIMER
Use the arrow keys to turn off the auto power off and set it in every 10-minutes.

3. Description for Using

[▲/LIGHT]
Indication of LCD Display
Main screen Backlight ON
1 Click
Main screen Backlight OFF
2 Click
-
Sweep Mode A / 0.45 kHz∼1.2 kHz
Sweep Mode B / 0.8 kHz ∼2.0 kHz
1 Click
Sweep Mode C / 1.4 kHz ∼ 3.5 kHz
2 Click
Sweep Mode D / 2.4 kHz ∼ 6.0 kHz
3 Click
-
Temperature Sensor Type / RTD
1 Click
Temperature Sensor Type / NTC
Display Mode 1 / Period(μs)
-
1 Click
Display Mode 2 / 10
Display Mode 3 / με0.391
2 Click
Display Mode 4 / με0.7756
3 Click
Display Mode 5 / με3.304
4 Click
Display Mode 6 / με4.062
5 Click
LCD Contrast 10 (LCD Brightness
-
adjustment)
-
POWER OFF TIMER(10min interval)
Battery status
-
Display Version Information
3
2
Hz
- 5

Advertisement

Table of Contents
loading