  • Page 1 HZSF-641 SF6 Gas Analyzer Huazheng Electric Manufacturing (Baoding) Co., Ltd...
  • Page 3 Advice Respected user: If you are using this product for the first time, please note the following: At the beginning of the measurement, first open the flow valve on the panel, then use the needle valve on the instrument to measure the flow. This device has a built-in large-capacity lithium battery that can work continuously for more than 10 hours when fully charged.
  • Page 4: Table Of Contents

    Contents I.Technical Characteristics And Parameters................1 II.Panel Instruction........................3 III.Measurement..........................4 IV.Menu Operation........................5 V.Remarks............................. 8 VI.Packing List..........................9...
  • Page 6: I.technical Characteristics And Parameters

    I.Technical Characteristics And Parameters 1. Technical Characteristics The SF6 comprehensive tester integrates the SF6 micro-water meter, the SF6 purity meter and the SF6 decomposition product analyzer into one, and the functions that were originally realized with three instruments are concentrated in one instrument. With one on-site measurement, three indicators can be tested, which greatly saves gas in the equipment.
  • Page 7 2.Technical Parameter Measuing Range 0%~100% measurement ±0.5% accuracy Purity measure time <2mins Measuring range -80 ℃~+20 ℃ measurement ±1℃ accuracy Humidity 63% need 5S, 90% need 45s(-60℃~+20℃) Response time (+ 20℃) 63% need 10S, 90% need 240s(-80℃~-60℃) Measuing Range 0~100ppm Minimum ≤0.1ppm detection...
  • Page 8: Ii.panel Instruction

    II.Panel Instruction 1.Front Panel Bracket adjust button Flow Gas inlet meter Key button Note: The angle of the bracket can be adjusted by simultaneously pressing the bracket adjustment buttons on both sides. 2.Rear Panel RS232 outlet Charging Power indicator AC220V switch...
  • Page 9: Iii.measurement

    III.Measurement 1. Connect SF6 Equipment Connect the threaded end of the measuring pipe to the switch joint, tighten with a wrench, and close the needle valve at the other end of the measuring pipe; Then insert one end of the quick connector on the test pipe into the sampling port on the comprehensive tester;...
  • Page 10: Iv.menu Operation

    pipe and the regulating valve on the comprehensive tester. Remove the adapter from the SF6 electrical equipment. If you need to continue measuring other equipment, please do not turn off the power ,follow the steps above to make the next measurement. 7.
  • Page 11 menu. The data will not be saved at this time. 2. View The Record In the measurement state, you can enter the function menu by pressing the “OK” button, press the “Up” and “Down” buttons to select the “View Record” menu, and press the “OK”...
  • Page 12 successful, perform the above steps as many times as possible (this function should ensure that there is no corresponding gas in the pipeline); 6. Calibration (Cancel + Right Key, This Function Is Limited To Special Personnel Operations) 7. Dew Point Value Calibration (Cancel + Up Key Combination Use, This Function Is Limited To Dedicated Operation) Press the “OK”...
  • Page 13: V.remarks

    V.Remarks The instrument should be placed in a safe position to prevent damage and avoid severe vibration. 2.Before using the instrument, it should be charged in time. When charging, just plug the power cord into the 220V socket. The instrument will automatically charge without turning on the power switch.
  • Page 14: Vi.packing List

    VI.Packing List Item Main engine Gas Inlet pipe Gas outlet pipe Power line...