Contents I. Introduction: adjustment.......................1 II.The main function..........................1 III Technical parameters........................2 IV. Principle of operation:........................ 3 V. Operation Instruction........................4 VI. Replacement of spare part......................29 VII.Diagnoses for breakdown....................... 33 VIII.Instrument storage and maintenance:................. 34 IX.Responsibility and guarantee of the factory................34 X.Appendix............................34 XI.Packing list..........................35...
I. Introduction: adjustment HPC500/400automobile exhaust analyzer according to the national standard GB18285-2018" ignition engine exhaust emission limits and methods of measurement ( double idle speed and simple driving mode" ) and development of a test apparatus. We determine the technical indicators, also referring to the people's Republic of China communications industry standard JT/T386-2004"...
minimum value and average value. 9, can enter the number plate of the car, and can store500 cars measurement data, for ready retrieval printing. 10, with single, double idle idle measurement measurement, general measuring three kinds of measurement. 11, with a speed, oil temperature measuring function. III Technical parameters 1, measurement range: HC:0 to 10000 ×10-6( PPM ) Vol ( petrol, hexane equivalent )
13, weight: 9kg IV. Principle of operation: The test of HC, CO and CO via HPC500/400 Auto Exhaust Analyser, which adopts nondispersive infrared ray analytical method (NDIR), is based on the principle that different gases selectively absorb infrared rays. The Instrument, by adopting narrowband infrared interference filter, is capable of analysing the content of required gas from tested gas, and therefore has good selectivity.
expressions of λ and carbon monoxide (CO) emendation value (COcor) whose function is to show that auto exhaust pipe has leakage if difference between COcor and CO test value is significant. At the moment, COcor represents actual CO value, namely the value when exhaust pipe has not leakage.
Page 8
and count-down is also adopted for displaying of the 30 sec.). After warm-up and zeroing finish, the following main menu image is displayed on display screen: (S: select, E: confirm) hh:mm:ss 0000 0.00 0.35 0.00 λ x.xxx Flow 2105sccm COcor 0.00% O.T 114.1℃...
Page 9
test. 3.2 ZERO (used for inspecting whether gas pipeline is severely polluted) Choose ZERO via Key-S, Key-▲ or Key-▼ and then press Key-E so that the instrument opens air pump and reversing valve for 20-sec “Zeroing” (or reset). A prompt appears on the screen: Zeroing, wait…19s(S: exit).
Page 10
Setup menu hh : mm :ss Stroke: √4 stroke 2 stroke IGN. mode: double √once Fuel type: √gasoline Test mode : Idle Time setup resturn S:select, E:confirm,∧∨ modify The setup menu is used by users for SET Stroke, IGN. mode, Fuel type, test mode and time setup (refer to the 5th item below for time setup).
Page 11
menu is displayed on display screen: Adjust hh :mm :ss 1:HC/CO/CO 2:O /NO ADJ 3:O T. ADJ 4:Config reset 5:Return 6:Voltage display 7:Gauge gas check (S: select, E:confirm) The menu is used by users for, when necessary, ADJ (also called “scaling”) or inspection ADJ of each item has been done before the instrument leaves factory.
Page 12
Press Key-S for exiting from the image and returning to adjust menu. Choose Item-7 Gauge gas check via Key-S, and then press Key-E, then the following image is displayed: Gauge check 11:07:35 C.Cur Check out HC/C S : select, E: confirm TEST Note: whether HC/ C appears on the screen is determined by the choice of fuel type in...
Page 13
is displayed: Database NO. ×××× hh:mm:ss 1:Delete all data 2:Index on number plates 3:Index on serial number 4: Return The image shows how much data is saved. SCH can be done either by index on number-plates or index on serial number . (the one saved last shall be the first Item and from back to front in sequence are 2nd, 3rd, ……Item) Item 1, 2, 3, 4 may be chosen in sequence by pressing Key-S.
Page 14
displayed on the screen: IDLE hh:mm:ss 0000 0.00 0.35 0.00 RPM 0 λ 1.000 Flow 2105sccm COcor 0.00% O.T 114.1℃ P 101.6kpa t 27.6℃ TEST SAVE and SAVE may not appear and cursor is at ESC when the test menu is just displayed STR , PRT , SAVE and CUR may be chosen in sequence by pressing Key-S, as well...
Page 15
cursor to the third character and press Key-▲ or Key-▼ for reset. Likewise, perform revision until the reset of the 8th character is done. Characters from the 2nd to 7th can choose from the Arabic numerals from 0 to 9 or the 26 capital English letters from A to Z. After reset of number plate is done, press Key-E for confirmation and cursor shall be at √BRP12345 (supposed to be new number plate)...
Page 16
First input number plate of vehicle to be tested as stated in Section 3.8 and 3.5, and then choose Stroke, IGN.M. mode and idle. 4.1.1 Choose STR via Key-S in test menu, and then press Key-E, then the following prompt is displayed on the screen: Set speed :5000 ∧∨...
Page 17
4.1.6 When RPM is adjusted to appropriate value, the following prompt image is displayed: Put the probe in the exhaust keep idle,S :exit 4.1.7 After the probe is inserted into the exhaust and exhaust gas enters into optical path of the instrument, the instrument enters into idle test. Testing,wait……44 keep idle,S:exit Test lasts altogether 45 seconds, and the former 15 seconds is at preparation stage.
Page 18
of the data saved will be displayed on the screen. Press key –s for return to test menu. 4.1.11 Choose AVE via Key-S and then press Key-▲ or Key-▼ to inspect the display of the minimum value, real-time value and maximum value, and then return to display of average value.
Page 19
Retard to idle(≤1100r/min) E: continue , S: exit After RPM is adjusted to idle, the instrument will prompt notify: Ready to test,wait…… keep idle,S :exit 2 seconds later, it shows: testing,wait……44 keep idle,S : exit Idle test is 45 seconds altogether, first 15 seconds are for preparation and the following 30 seconds are for formal test.
Page 20
4.3.2 After 20-second zeroing ends, the screen shows: Put the probe in the exhaust S : exit 4.3.3 With probe injected into exhaust, the exhaust gas gets into optical path of the instrument and when the instrument shifts to test state, the screen will show: Testing,wait……44 S: exit After the 45 seconds count-down completes, the air pump does not need to be closed and...
Page 21
4.4.3After zeroing completes, 30 seconds countdown for HC remained check test will automatically start and the screen will show: HC remained check,wait……29 S : exit 4.4.4 HC remained passes test and the screen will show: After preheat,restard to idle E : continue S : exit 4.4.5 The engine preheats for 30 seconds at 70% of rated speed and then adjust speed to idle.
Page 22
select double idle 4.5.1 Start up test like 4.2.1 4.5.2 Press S for 20-second automatic zeroing count down. The same as 4.2.2 (namely, the same as 4.1.2) 4.5.3 When zeroing completes, 30 seconds count-down for HC leftover test will automatically start. It is the same as 4.2.3 (e.g. 4.1.3): 4.5.4 HC leftover passes test and the screen will show: After preheat ,retard to fast idle E : continue,S : exit...
Page 23
4.5.11 After the probe is took out and the concentration of HC is reduced to 30% of average value, valve is automatically switched on and the optical path cleaning is accelerated while the screen shows: zeroing……59 S : exit) When 60 seconds countdown ends, close air pump, carry on 5-second cleaning (reverse blowing), shut off valve and automatically return to test menu.
Page 24
HC/CO/CO hh : mm : ss Gsuge Current New scale H3C8 1600 1.002 3.60 0.00 1.031 13.00 0.35 1.001 S :select ,E :confirm,∧∨ : modify: Choose in sequence ADJ ,HC CALC、 CO CALC、 CO CALC、 HC(at this time press▲ and then C appears)...
Page 25
adjusting,wait……119 S: cancel Open air pump and reversing valve, clean optical path to facilitate fresh air’s (20.90% O passing through oxygen sensor. When 2-minute countdown ends, save the current voltage value of O as the adjust value. Shut off air pump and valve, and O ADJ ends.
Page 26
Then, put OIL. T probe and standard thermometer into high temperature source and reset high standard temperature according to thermometer and after indicating value stabilizes, press S to select H.T. and E for adjusting high temperature point. Select ESC via S and return to adjust menu via E. 6.4 Restore original data When errors occur in adjust, adjust data set before coming off line can be taken into consideration.
Page 27
Press S to select 5:Return; return to adjust menu via E. Press E again, return to adjust menu. 7. Output Interface The instrument could be equipped with mini PRTer to PRT out test results. It could also be equipped with D/A analog output interface or RS-232 digital output interface.
Page 28
Jack port of an external computer 7.2.1 Parameter of RS232 serial communication Baud rate: 9600 BPS Length of characters: 1 byte is start, 8 bytes is data and 1 byte for stop Test method: no test7.2.2 O3H order given by external computer (host machine), referring to real-time data (without stops, e.g.
Page 29
External computer gives out: 07H Correspondence of the instrument: ACK(06H) 8.1.8 SET IGN.M. mode as ONCE External computer gives out: OAH Correspondence of the instrument: ACK(06H) 8.1.9 SET IGN.M. mode as DOUBLE External computer gives out: OBH Correspondence of the instrument: ACK(06H) 8.1.10 HC remained check External computer gives out: 08H Correspondence of instrument: testing HC remained check: 00H...
Page 30
out 10 groups of data at most. When data staved exceeds 10 groups, external computer could send out OFH orders many times and continually refer to the 10 groups of data until all data is obtained. 8.2Test process controlled by host computer (serial communication of idle double idle and general) After preheat, the instrument is in the state of test (cursor is on Test ) 8.2.0 External computer gives out orders, and selects Strokes, IGN.M.
Page 31
Correspondence of instrument: 00H (indicating that the instrument has opened air pump, zeroing valve and is carrying on 30-second zeroing) External computer sends 09H orders again and again during the period until the instrument corresponds ACK (06H). 8.2.5 External computer sends again: 09H (zeroing once again) Correspondence of instrument: 00H, sending back 06H (ACK) after 30 seconds 8.2.6 External computer transmission:02H(indicating switching off air pump, exit)...
8.4.5 O :The O data represents the same as CO。 8.4.6 NO :The NO data, stipulation is similar to HC。 8.4.7 n: RPM is the integerr with symbol (2 bytes). High byte precede, and its unit is r/min. 8.4.8 O. T data is the number with marks (2 bytes). The top digit in first, and the unit is ℃. 8.4.9 λ: The current data for excess air coefficient is the integer with symbol (2 bytes).Highby6te precede , and the data expanded 100 times, for example, the test result ofλis 1.05, then theλdata is 0069H.
Page 33
1. Screw lids 2. Lid for locating 3. Big seal packing collars 4. Small seal packing collar 5.Filter paper 6. Dust filter seat 7. Analyzing box Fig.4 Assembly drawing for dust filter 2. Replacement for paper The instrument has been installed scroll while it leaves the factory. If the paper has been used up and needs replacing, please do as following: (1)Open the door of paper box.
Page 34
Fig.5 Plastic support for installing scroll (3)Try your best to take down the scroll, replace the blank scroll with the new one. After the new scroll set in the axis, install the scroll on the plastic support in the guide channel (Please note you must make sure the scroll has been installed reliable and will not fall), cut the scroll like the style in Picture 6, in order to insert the PRTer well.
Page 35
Box gate Color ribbon PRTer nose Fig.7 PRTer nose inside box gate Take down the former color ribbon from PRTer nose (see Picture 8 schematic drawing for taking down color ribbon). (2) Attention: Please first lift the right margin of the color ribbon box, and then lift left side again to take down the color bandbox.
VII.Diagnoses for breakdown 1The automatic diagnoses for breakdown (1) Choose L.P. via S in the main menu, press E, then operate as showed. If it discovers gas pipeline leakage, then promptly, “leakage” will be shown in the display monitor. (2) when starting HC remained check, if discovers gas pipeline obstruction to cause the flow excessively to be low, then “gas pipeline obstruction”...
replaced; if the air pump work is normal, and the test data is too small, you should inspect gas tightness of instrument (monitoring instrument and front filter, pipeline attachment place whether leak). In the main menu press S to select ADJ , press E to enter adjust menu, again choose 6: VOLTAGE DISPLAY in adjust menu via S;...
(1) Exhaust gas system must have no leakage (2) The depth of pipe inserting the exhaust must not be less than 300mm (3) Engine must be up to the stipulated hot condition Four Strokes: Water cooling: Water temperature 60℃ above; Forced-air cooling: Oil temperature 40℃ above (4) Engine regulating to stipulated idle 3.Test procedure:...