AKSO CLORO Tester Instruction Manual

Advertisement

INSTRUCTION MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CLORO Tester and is the answer not in the manual?

Questions and answers

Summary of Contents for AKSO CLORO Tester

  • Page 1 INSTRUCTION MANUAL...
  • Page 2 1. SPECIFICATIONS - Measuring range: 0.00 a 3.50 ppm - Resolution: 0.01ppm - Accuracy: ± (0.05ppm + 3% reading) - Sample temperature: 15 a 35 °C - sample volume: 10mL - Light source: Light emitting diodo (LED) - Wavelength: 525nm - Measuring cell: Glass cuvette - Auto power off: After 2 minutes - Languages: English, Spanish and Portuguese...
  • Page 3 - 2 glass cuvette with lid (Ø18.8 mm x 68mm) - 1 cleaning wipe - 1 plastic case - 1 instruction manual Before use, examine the instrument carefully. If you detect any abnormalities, contact AKSO. 3. PRESENTATION Measuring cap Automatic shutdown Battery level...
  • Page 4 4. OPERATING INSTRUCTIONS MEASUREMENT Preparation 1) Make sure that the temperature of the water sample to be analyzed is between 15 and 35°C to ensure the best accuracy in the chlorine measurement; 2) Wash the cuvette 3 to 4 times using the water sample that will be analyzed to set the container, avoiding reading deviations;...
  • Page 5 Reading – Free chlorine While Insert reagent is on the display: 11) Open the measuring cover and remove the cuvette from the compartment; 12) Remove the cuvette cap and add 3 drops of DPD1 reagent and 3 drops of DPD2 to the sample; 13) Close the cuvette with its lid;...
  • Page 6 21) After counting, total chlorine will be read in the water sample. Wait; 22) View the total chlorine measurement value in ppm on the Oled display. NOTES:  IMMEDIATELY, after each measurement, discard the sample and wash the cuvette.  Be very careful when handling the cuvettes! Scratches, stains and marks can compromise the results.
  • Page 7 6. MAINTENANCE BATTERY INSTALATION/REPLACEMENT When the display displays Low Battery, replace the battery as described below: 1) Turn off the instrument by holding down the Power- Measurement button until the display is switched off; 2) Remove the base clamping screw at the bottom of the instrument, using a “Philips”...
  • Page 8 Failure to communicate with Contact Akso Technical Damaged sensor the sensor. Support. Go to the product page on the Akso website and make sure your manual version is up to date. AKSO PRODUTOS ELETRÔNICOS LTDA www.akso.com.br • sales@akso.com.br +55 51 3406 1717...