Download Print this page

Performance Specifications - Agilent Technologies E3632A User Manual

Dc power supply
Hide thumbs Also See for E3632A:

Advertisement

Chapter 8 Specifications

Performance Specifications

Performance Specifications
Output Ratings(@0°C - 40°C)
Low range
0 to +15 V/0 to 7 A
High range
0 to +30 V/0 to 4 A
[1]
Programming Accuracy
Voltage
0.05% + 10 mV
Current
0.2% + 10 mA
[1]
Readback Accuracy
12 months(over GPIB and RS-232 or front panel with respect
to actual output @25°C ± 5°C), ±(% of output + offset)
Voltage
0.05% + 5 mV
Current
0.15% + 5 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
Normal mode voltage
Normal mode current
Common mode current
Load Regulation, ±(% of output + offset)
Change in output voltage or current for any load change within ratings with remote
sensing connected
Voltage
<0.01% + 2 mV
<0.01% + 250 μA
Current
Line Regulation,
±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage
<0.01% + 2 mV
Current
<0.01% + 250 μA
[1]
Accuracy specifications are after an 1-hour warm-up with no load and calibration at
25 °C.
12 months(@25°C ± 5°C), ±(% of output + offset)
<0.35 mV rms and 2 mV p-p
<2 mA rms
<1.5 μA rms
8
157

Advertisement

loading