Chapter 1 Specifications

Performance Specifications

Performance Specifications
Output Ratings (@ 0°C - 40°C)
+6V Output
0 to +6 V ; 0 to 5 A
+25V Output
0 to +25 V ; 0 to 1 A
-25V Output
0 to -25 V ; 0 to 1 A
[1]
Programming Accuracy
+6V Output
Voltage
0.1% + 5 mV
Current
0.2% + 10 mA
[1]
12 months (over GPIB and RS-232 or front panel with respect
Readback Accuracy
±
to actual output @ 25°C
5°C),
+6V Output
Voltage
0.1% + 5 mV
Current
0.2% + 10 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
+6V Output
Voltage
<0.35 mV rms
<2 mV p-p
Current
<2 mA rms
<1.5 μA rms
Common mode current
±
Load Regulation,
(% of output + offset)
Change in output voltage or current for any load change within ratings
Voltage
<0.01% + 2 mV
<0.01% + 250 μA
Current
±
Line Regulation,
(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage
<0.01% + 2 mV
Current
<0.01% + 250 μA
[1]
Accuracy specifications are after an 1-hour warm-up and calibration at 25°C
±
12 months (@ 25°C
5°C),
+25V Output
0.05% + 20 mV
0.15% + 4 mA
±
(% of output + offset)
+25V Output
0.05% + 10 mV
0.15% + 4 mA
+25V Output
<0.35 mV rms
<2 mV p-p
<500 μA rms
±
(% of output + offset)
-25V Output
0.05% + 20 mV
0.15% + 4 mA
-25V Output
0.05% + 10 mV
0.15% + 4 mA
-25V Output
<0.35 mV rms
<2 mV p-p
<500 μA rms
17