TM
LFA 447
Reassembly to the 300°C-furnace
The reassembly to the 300°C furnace with
four samples must be carried out in reverse
order.
01/09
J:\LFA447\CHAPTER5.DOC
Appendix
ATTENTION!
• sensor:
guide pins
sensor
For the 300°C furnace the sensor must be
placed in the lower position so that the guide
pins go through the holes in the sensor plate
as shown in the figure.
• Software settings:
software settings for the 300°C furnace:
- The standard sample holder with Four
Samples must be selected in the instrument
configuration:
Menu: File
Menu item: Instrument Configuration
- The furnace type 300 must be
selected in the instrument configuration:
Menu: File
Menu item: Instrument Configuration
→
Sample Tray
Four
Samples
→
Furnace Type
300