(4) Sample Area Blood should be applied to the tip of the
One Drop Test Strip here.
(5) Visual Fill Window This window will turn red when
enough blood has been applied.
Important Test Strip Information:
• Store the One Drop Test Strip vial in a cool, dry place at
46°F to 86°F (8°C to 30°C) and at a relative humidity of 10%
to 90%.
• Keep away from direct sunlight and heat.
• Store your One Drop Test Strips in their original vial only;
never store them in another vial, any other container, or
outside the vial.
• With clean, dry hands you may gently handle the One Drop
Test Strip when removing it from the vial or inserting it
into the One Drop Meter.
• After removing a One Drop Test Strip from the vial, imme-
diately close the vial cap tightly.
• Do not use One Drop Test Strips beyond the expiration
date or 180 days after first opening the vial. This may
cause inaccurate results. Write the discard date (180 days
from the first opening) on the One Drop Test Strip vial.
• Do not bend, cut, or alter One Drop Test Strips.