Herida Healthcare Grampian II User & Service Manual

Herida Healthcare Grampian II User & Service Manual

Pressure relieving / alternating air mattress & pump

Advertisement

Quick Links

Grampian II™ User/Service Manual
User & Full Service Manual
Pressure Relieving / Alternating Air Mattress & Pump
Pump Model: HH/P08 Mattress Model: HH/T05
NB: Please ensure this document is fully read & understood
by the Clinician / Carer / Service Engineer prior to use.
1 |
P a g e
A U G U S T 2 0 1 9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Grampian II and is the answer not in the manual?

Questions and answers

Roby
August 14, 2025

Showing error in air mattresses pressure pump model number APP-PO6

User image 689db1ba59e3d

Summary of Contents for Herida Healthcare Grampian II

  • Page 1 Grampian II™ User/Service Manual User & Full Service Manual Pressure Relieving / Alternating Air Mattress & Pump Pump Model: HH/P08 Mattress Model: HH/T05 NB: Please ensure this document is fully read & understood by the Clinician / Carer / Service Engineer prior to use.
  • Page 2 Grampian II™ User/Service Manual P a g e A U G U S T 2 0 1 9...
  • Page 3: Table Of Contents

    Grampian II™ User/Service Manual CONTENTS SAFETY………………………………………………………4 INTRODUCTION………………………………………….…5 Product Description Application Contra-Indications Working Principle OPERATION………………………..………………………..6 Static Mode Alternating Mode Profiled Bed – Clinical Considerations Installation Safety Instructions Pump Operation Mattress & Pump Characteristics MAINTENANCE……………………………………………..10 Cleaning Maintenance & Repair Storage Transportation Other Considerations Parts &...
  • Page 4: Safety

    & these should only be carried out by suitably trained & authorised personnel. For further advice please call Herida Healthcare on Freephone 0800 193 6030. P a g e...
  • Page 5: Introduction

    This product forms part of a holistic care package and should only be used by trained personnel. For training or further advice contact Herida Healthcare on Freephone +44 800 193 6030. Working Principle The dynamic mattress system consists of a manual air pump and an alternating air mattress. The air mattress has three distinct airway chambers: 1.
  • Page 6: Operation

    (*see below). Unless the clinical condition dictates, it is not advised to adjust the bed profile beyond 45°. The Grampian II™ will continue to operate beyond this but may cause increased pressure points on the patient. A fist or hand test to ensure patient clearance under the sacral area is a best practice guide (see image 1DA and diagram of profiled mattress 2DA or visit our website (www.heridahealthcare.co.uk) under “Education”...
  • Page 7: Installation

    Grampian II™ User/Service Manual OPERATION Image 1DA (NB: use hand to check ensure appropriate clearance when profiled as below. Add more pressure on the pump if required in order to achieve this. NB: Weight setting is a guide only and cell pressure must also be taken into consideration for individual patients).
  • Page 8: Safety Instructions

    Grampian II™ User/Service Manual OPERATION Safety Instructions 1. Ensure that there are no sharp objects, fire or abrasive materials in the vicinity. If the mattress is damaged. Remove the patient & transfer to an alternative mode of pressure relief. 2. If there is an issue with the pump, remove the patient & transfer to another mode of pressure relief.
  • Page 9: Mattress & Pump Characteristics

    Grampian II™ User/Service Manual OPERATION Pump Characteristics Power AC220V / AC110V 30 – 120 mmHg Pressure Range Internal Fuse F0.5A 250V Plug Fuse Frequency 50Hz / 60 Hz Power ≤45dB Noise Alternating Time 1 complete cycle is 11 ½ minutes which is 2 x 5 minute periods of alternative cell inflation with a change over time of 1 ½...
  • Page 10: Maintenance

    Maintenance & repair work should only be carried out by an appointed engineer who have been trained by Herida Healthcare Ltd. and only original manufacturers parts should be used. Failure to do so could result in a void warranty situation.
  • Page 11: Storage

    Grampian II™ User/Service Manual MAINTENANCE At point of service the following aspects must be maintained to ensure that manufacturer’s warranty remains intact. These can be done in a non-alphabetic way, if all areas are covered: • Ensure the pump unit and mattress / cells are fully decontaminated prior to service and are cleaned in accordance with the schedule provided.
  • Page 12: Transportation

    Grampian II™ User/Service Manual MAINTENANCE Transportation • Care should be taken when loading, unloading and during transportation & refrain from placing heavy objects on top of the mattresses. • Do not expos to direct sunshine for any period of time when loading and unloading &...
  • Page 13: Routine Checks

    Grampian II™ User/Service Manual ROUTINE CHECKS Mattress • Remove top cover and inspect for signs of wear or any tears. • Check zips are still operative & function as normal. • Check integrity of all connectors including cells, tubing and CPR toggle.
  • Page 14: Warnings & Symbols

    Grampian II™ User/Service Manual WARNINGS & SYMBOLS Warning – Follow guidance at all times as per the instructions within the user manual. BF application symbol – Provides a higher degree of protection against electric shock. Class II electrical appliance – Double insulated.
  • Page 15: Trouble Shooting

    Grampian II™ User/Service Manual TROUBLE SHOOTING Issue Possible Cause Possible Resolution • The plug is not • Make sure the power is properly connected. correctly connected and the • Power Cut. voltage is normal. • Await re-connection of • The fuse is broken.
  • Page 16: Company Information

    VAT Registration Number: 233587691 Disclaimer Please be aware that all products should be disposed of in an environmentally friendly & infection control manner in compliance with any local or national requirements. Contact Herida Healthcare Ltd. for responsible WEEE disposal methods. 16 |...