SPECIFICATIONS
ENVIRONMENTAL CONDITIONS
Conditions
Normal Use
Storage & Transport
ELECTRICAL SPECIFICATIONS
Designed for:
Input Service
Current Draw
Voltage Output to
Actuators
Electric Shock
Protection
Tabletop Applied Part
Ingress Protection
Rating
Made of Operation
Temperature
50° (10°C) to 104° (40°C)
-20° (-29°C) to 135° (57°C)
North America
120 VAC/15 amp/60 Hz
5.0 amps
24 VDC
Class 1 Equipment
Type B Applied Part
IPX0
Intermittent Operation MAX
2 minutes ON
18 minutes off
Humidity
Atmospheric Pressure
20% to 60% RH
20% to 95% RH
Europe
220 VAC/10 amp/50/60 Hz
2.5 amps
24 VDC
Class 1 Equipment
Type B Applied Part
IPX0
Intermittent Operation MAX
2 minutes ON
18 minutes off
98 to 105 kPa
98 to 105 kPa
Japan
100 VAC/15 amp/50/60 Hz
5.0 amps
24 VDC
Class 1 Equipment
Type B Applied Part
IPX0
Intermittent Operation MAX
2 minutes ON
18 minutes off
26