Villa Sistemi Medicali Visitor T15 User Manual

Villa Sistemi Medicali Visitor T15 User Manual

Mobile x-ray system
Hide thumbs Also See for Visitor T15:

Advertisement

Quick Links

Revision: 2
Version:
March 26, 2004
File:
[M_30010E00R02.doc]
WARNING: The information that is printed within this manual is vital for the correct use of the
equipment; please read it carefully before use.
U
s
e
r
'
U
s
e
r
s
M
a
'
s
M
a
n
u
a
l
n
u
a
l

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Visitor T15 and is the answer not in the manual?

Questions and answers

Summary of Contents for Villa Sistemi Medicali Visitor T15

  • Page 1 Revision: 2 Version: March 26, 2004 ’ ’ File: [M_30010E00R02.doc] WARNING: The information that is printed within this manual is vital for the correct use of the equipment; please read it carefully before use.
  • Page 2 User's Manual - Visitor T15 Villa Sistemi Medicali (This page is intentionally left blank) [File: M_30010E00.doc]...
  • Page 3: Table Of Contents

    Villa Sistemi Medicali Visitor T15 - User's Manual TABLE OF CONTENTS 1. 1. 1. 1. SAFETY AND COMPLIANC SAFETY AND COMPLIANCE E E E ........................................2 2 2 2 SAFETY AND COMPLIANC SAFETY AND COMPLIANC ........
  • Page 4: Safety And Compliance E E E

    Villa Sistemi Medicali Villa Sistemi Medicali Villa Sistemi Medicali cannot be held liable for any malfunction, damage, or danger resulting from improper cannot be held liable for any malfunction, damage, or danger resulting from improper cannot be held liable for any malfunction, damage, or danger resulting from improper...
  • Page 5: Electromagnetic Compatibility (Emc)

    General disposal General disposal Villa Sistemi Medicali produces radiological systems that are advanced in terms of safety and environmental protection. Assuming that the unit is properly used, there is no risk to people or the environment. In order to comply with applicable safety requirements, it is necessary to use materials that may be harmful to the environment (for example: monobloc oil, protective lead, boards and electronic components).
  • Page 6: Application & Final Destination

    (mod. DIAMENTOR PX). It complies with the safety requirements foreseen by the 93/42/EEC Directive. However, the liability of the interface, if it has not been evaluated and authorized by Villa Sistemi Medicali in writing, is of the operator and/or the person who has performed the interface.
  • Page 7: Compliance

    This x-ray unit is in compliance with the electromedical devices Directive 93/42 EEC and with the other national and international standards in force. The distributor (according to the European Directive 93/42/EEC) of the unit Visitor T15 is: Villa Sistemi Medicali...
  • Page 8: Component Identification Tion

    User's Manual - Visitor T15 Villa Sistemi Medicali COMPONENT IDENTIFICATION 2.1. 2.1. 2.1. 2.1. Overview Overview Overview Overview Monobloc 10. Cassette holder Goniometer 11. Magneto-thermic switch Monobloc handle 12. Supply cable Collimator 13. Handle for unit movement Supply cable-winding 14. Control panel Equipotential node 15.
  • Page 9: Control Panel

    Villa Sistemi Medicali Visitor T15 - User's Manual 2.3. 2.3. Control panel Control panel 2.3. 2.3. Control panel Control panel Figure 3 Here is a brief description of the keys function in standard use and the signals meaning. Unit OFF.
  • Page 10: Audible Signals

    User's Manual - Visitor T15 Villa Sistemi Medicali kV+ kV- To modify KV value mAs+ mAs- To modify mAs value RESET Return to the main menu. Function keys: F1, F2, F3, F4 They refer to the display line number (4 lines display).
  • Page 11: Signals And Error Messages

    Villa Sistemi Medicali Visitor T15 - User's Manual 2.5. 2.5. Signals and error messages Signals and error messages 2.5. 2.5. Signals and error messages Signals and error messages The unit foresees three types of alarm that can appear on the display: •...
  • Page 12: Messages On The Disp

    User's Manual - Visitor T15 Villa Sistemi Medicali MESSAGES ON THE DISPLAY F = fatal error W = alarm S = unit status Text Meaning Action READY The unit is ready to perform an exposure BUSY Preparation phase Wait for “READY” message...
  • Page 13 Villa Sistemi Medicali Visitor T15 - User's Manual The table below shows all the messages and signals in the five languages that can be set. English (GB) English (GB) Italian (I) Italian (I) French (F) French (F) German (D) German (D)
  • Page 14: Functioning

    User's Manual - Visitor T15 Villa Sistemi Medicali FUNCTIONING 4.1. 4.1. Transport Transport 4.1. 4.1. Transport Transport For the transport of the unit, consider the following instructions: • The unit must be OFF, the supply plug must be removed from the socket outlet and the cable wound.
  • Page 15: Positioning

    Villa Sistemi Medicali Visitor T15 - User's Manual 4.2. 4.2. 4.2. 4.2. Positioning Positioning Positioning Positioning For positioning the unit the following instructions should be considered: Don't move the unit when the brakes are activated. For the movements use the proper handles.
  • Page 16: Collimator Adjustment

    User's Manual - Visitor T15 Villa Sistemi Medicali 4.3. 4.3. Collimator adjustment Collimator adjustment 4.3. 4.3. Collimator adjustment Collimator adjustment 1. 1. 1. 1. On the frontal panel of the collimator, there are two knobs for the beam adjustment (width and length) as well as the push-button to turn ON the collimator lamp (see Figure 12).
  • Page 17: Start Up And Checks At The Ignition

    Villa Sistemi Medicali Visitor T15 - User's Manual 4.4. 4.4. 4.4. 4.4. Start up and checks at the ignition Start up and checks at the ignition Start up and checks at the ignition Start up and checks at the ignition Connect the unit to the mains.
  • Page 18: Exposures

    User's Manual - Visitor T15 Villa Sistemi Medicali 4.5. 4.5. Exposures Exposures 4.5. 4.5. Exposures Exposures 4.5.1. 4.5.1. 4.5.1. 4.5.1. Operative procedure Operative procedure Operative procedure Operative procedure Set the exposure data by following the operative procedure here below. IGNITION & START UP...
  • Page 19: Programs For The Anatomical Programming Mode (Apr Mode)

    Villa Sistemi Medicali Visitor T15 - User's Manual 4.5.2. Programs for the Anatomical programming mode (APR MODE) Here is a table with the APR default values. Consider that they are valid by placing the x-ray tube with a SID (Source-Image receptor Distance) of 100 cm without grid.
  • Page 20: Perform An Exposure

    User's Manual - Visitor T15 Villa Sistemi Medicali 4.5.3. 4.5.3. Perform an Perform an exposure exposure 4.5.3. 4.5.3. Perform an Perform an exposure exposure Before performing an exposure, make sure that all the necessary precautions against radiation Before performing an exposure, make sure that all the necessary precautions against radiation...
  • Page 21: Optional: Dosimeter (Dapmeter)

    Villa Sistemi Medicali Visitor T15 - User's Manual 4.5.5. Optional: dosimeter (DAPmeter) The unit can be supplied, on request, with a ionizing chamber dosimeter (dose-area product meter, DAP meter) installed. The DAPmeter function is to measure the dose-area product [cGgcm ] in output towards the patient.
  • Page 22: Optional: Radiography With Examination Table And Potter Bucky Grid

    User's Manual - Visitor T15 Villa Sistemi Medicali 4.5.6. 4.5.6. Optional: Radiography with examination table and Potter Bucky grid Optional: Radiography with examination table and Potter Bucky grid 4.5.6. 4.5.6. Optional: Radiography with examination table and Potter Bucky grid Optional: Radiography with examination table and Potter Bucky grid...
  • Page 23: Maintenance

    Villa Sistemi Medicali Visitor T15 - User's Manual MAINTENANCE Villa Sistemi Medicali can supply, on request, a programmed maintenance plan to be performed on the unit. 5.1. 5.1. 5.1. 5.1. General Warnings General Warnings General Warnings General Warnings As with any medical device, this system requires: •...
  • Page 24: Cleaning

    User's Manual - Visitor T15 Villa Sistemi Medicali 5.3. 5.3. Cleaning Cleaning 5.3. 5.3. Cleaning Cleaning Please take the following information into consideration before choosing a detergent: • To clean plastic surfaces, simply use water and soap, and nothing else. If other detergents are used (e.g.
  • Page 25: Technical Data

    Villa Sistemi Medicali Visitor T15 - User's Manual TECHNICAL DATA 6.1. 6.1. 6.1. 6.1. Electrical data Electrical data Electrical data Electrical data Description Description Data Data Description Description Data Data Voltage 115/230Vac ±10% standard monophase with automatic unit prearrangement in function of the mains (plug &...
  • Page 26: Radiological Data

    User's Manual - Visitor T15 Villa Sistemi Medicali 6.3. 6.3. Radiological data Radiological data 6.3. 6.3. Radiological data Radiological data Characteristic Characteristic Characteristic Characteristic Performances Performances Performances Performances Working technique 2 points technique with kV and mAs setting APR anatomic program 24 exams memorization (4 programs each of 6 exams) available in the 5 different selectable languages.
  • Page 27: Environmental Data

    Villa Sistemi Medicali Visitor T15 - User's Manual 6.4. 6.4. Environmental data Environmental data 6.4. 6.4. Environmental data Environmental data Description Description Description Description Normal use Normal use Normal use Normal use Transport and storage Transport and storage Transport and storage...
  • Page 28: Unit Sizes

    User's Manual - Visitor T15 Villa Sistemi Medicali 6.5.1. 6.5.1. Unit sizes Unit sizes 6.5.1. 6.5.1. Unit sizes Unit sizes Figure 15 Pag. 26/32- Rev.1 [File: M_30010E00.doc]...
  • Page 29: Components Specifications

    Villa Sistemi Medicali Visitor T15 - User's Manual 6.6. 6.6. Components specifications Components specifications 6.6. 6.6. Components specifications Components specifications 6.6.1. 6.6.1. 6.6.1. 6.6.1. Generator Generator Generator Generator Description Description Description Description Data Data Data Data Inverter IHF 2015 Working frequency 20kHz max Max.
  • Page 30: Tube-Housing Assembly

    User's Manual - Visitor T15 Villa Sistemi Medicali 6.6.2. 6.6.2. Tube Tube- - - - Housing Assembly Housing Assembly 6.6.2. 6.6.2. Tube Tube Housing Assembly Housing Assembly X X X X - - - - Ray Tube Ray Tube Ray Tube...
  • Page 31 Villa Sistemi Medicali Visitor T15 - User's Manual Monobloc Monobloc Monobloc Monobloc Description Description Data Data Description Description Data Data Monobloc MHF 2015 Weight 19 Kg (41,89Lb) Dimensions 320x140x255mm (12,60x5,51x10,04in.) X-ray tube X22 0.8/1.3 (Large focus not used) Anode Rotating (2850rpm at 50Hz)(3400rpm at 60Hz) External thermostat 57°...
  • Page 32: Collimator

    User's Manual - Visitor T15 Villa Sistemi Medicali 6.6.3. 6.6.3. Collimator Collimator 6.6.3. 6.6.3. Collimator Collimator Description Description Description Description Data Data Data Data Type, brand and model Manual with internal light source (Ralco R221) Collimator Square field, multilayers Light source Halogen Lamp 12V 100W with timed switching-on at approx.
  • Page 33: Labels And Symbols

    Villa Sistemi Medicali Visitor T15 - User's Manual 6.9. 6.9. Labels and symbols Labels and symbols 6.9. 6.9. Labels and symbols Labels and symbols 6.9.1. 6.9.1. 6.9.1. 6.9.1. Labels of the unit Labels of the unit Labels of the unit...
  • Page 34: Internal Symbols

    User's Manual - Visitor T15 Villa Sistemi Medicali 6.9.2. 6.9.2. Internal symbols Internal symbols 6.9.2. 6.9.2. Internal symbols Internal symbols Protective ground clamp Dangerous voltage Precautionary warning 6.9.3. 6.9.3. 6.9.3. 6.9.3. Various symbols Various symbols Various symbols Various symbols Unit OFF...
  • Page 35: Document Status

    Document status Visitor T15 – User's Manual DOCUMENT STATUS Rev. Rev. Date Date Pages Pages Modification description Modification description Rev. Rev. Date Date Pages Pages Modification description Modification description 26.08.03 Document approval 14.01.04 Documents general revision. 26.03.04 Performed modification of the APR values in the unit software.

Table of Contents