Summary of Contents for PCB Piezotronics 1230-11ADB
Page 1
Model 1230-11ADB LOW PROFILE LOAD CELL, DUAL BRIDGE Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
E-mail: info@pcb.com representative. Warranty – All equipment and repair services provided by PCB Piezotronics, Inc. are covered by a limited warranty against defective material workmanship for a period of one year from date of original purchase. Contact...
Page 5
Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
Page 6
Model Number Revision: NR LOW PROFILE LOAD CELL, DUAL BRIDGE 1230-11ADB ECN #: 43066 Performance ENGLISH OPTIONAL VERSIONS [5][6] Measurement Range(100 x 10^6 cycles) 100 klb 445 kN Optional versions have identical specifications and accessories as listed for the standard model [7][1] except where noted below.
Page 7
PCB Load & Torque Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others without DESCRIPTION the written consent of PCB Load and Torque Inc. RELEASED FOR PRODUCTION - 06.18.14, PTE 43066 4.50 [114.3] CONNECTOR "B"...
DUAL BRIDGE LOAD CELL OPERATION MANUAL 3.1 Dimensions 1.0 INTRODUCTION The following figure and table give the general outline Dual bridge fatigue-rated load cells manufactured by PCB dimensions of the dual bridge fatigue-rated load cells. Load and Torque, Inc. are guaranteed to last 100 million fully reversed cycles (full tension through full compression).
DUAL BRIDGE LOAD CELL OPERATION MANUAL 3.2 Standard Components 4.0 MECHANICAL INSTALLATION The following figure describes the standard components of the Standard components of the dual bridge fatigue-rated load dual bridge fatigue-rated load cells. cells have all been installed to factory specified values for optimum performance.
DUAL BRIDGE LOAD CELL OPERATION MANUAL 5.2 Cable & Grounding Considerations Proper grounding and shielding is required to prevent electrical noise in strain gage load cell measuring systems. The cable must be shielded twisted pairs with a drain wire. Cable shields must be grounded only at one end, for example, on the instrument or control system ground.
DUAL BRIDGE LOAD CELL OPERATION MANUAL 6.1.1 Measured Output 7.0 MOMENT COMPENSATION The applied load starting at zero is measured in five When an eccentric axial load is Primary increments to full scale. Output (mV/V) is measured at each applied to a load cell that is not Axis increment.
DUAL BRIDGE LOAD CELL OPERATION MANUAL Dual bridge fatigue rated load cells, when mounted on the 11.2.1 Estimating Bridge Balance using an Ohm Meter factory supplied base, are barometric compensated through a small port located in one of the spanner wrench holes on the A load cell that has been severely overloaded will exhibit a base.
DUAL BRIDGE LOAD CELL OPERATION MANUAL Certificate information includes tabulated measurement variable data zero balance, bridge input/output resistance, computer nonlinearity and hysteresis, static error band (SEB) calculations and entries abilities and traceability statements. If an initial evaluation shows that a transducer requires repair, PCB will provide the customer with an estimate prior to taking any corrective action.