Open the device web interface
On the device website, you can upgrade the device embedded software, get log files, roll back software
images, reboot the device and manage license files.
To open the device web interface:
1. In the home view, select the device from the list.
2. Click More beneath the Connect button, and then click Open device web interface.
A website with the IP address of the device opens in the default web browser.
4.5
Connect CIBANO 500 to the test object
WARNING
Death or severe injury caused by high voltage or current possible
► Do not connect CIBANO 500 to a test object before isolating the test object according
to the five safety rules.
► Always obey the five safety rules (see 1.2.2 "Safety rules" on page 3) and all additional
relevant laws and internal safety standards when connecting CIBANO 500 to a test
object.
You can connect CIBANO 500 to the test object without disconnecting other parts of the station or with
complete disconnection from the station as shown in the following figure.
+
For example,
relay
Trip coil
Figure 4-3: Principal connection of CIBANO 500 to the test object
To connect CIBANO 500 to the test object:
1. Connect the equipotential ground terminal of CIBANO 500 (see 3.2.2 "Side panel" on page 11) to
ground as close as possible to the operator.
OMICRON
For example,
control room
CIBANO 500
Close coil
Installation
17

Advertisement

Table of Contents
loading