3.2.3

Warning lights

CIBANO 500 provides the following warning lights to indicate safe operation and possible hazards.
Table 3-1: Warning lights
Warning
Description
light
Green light on the front
panel is on.
Blue ring on the
Measurement
Start/Stop button is on.
Blue ring on the
Measurement
Start/Stop button is
flashing.
Red light on the front
panel is flashing.
Warning symbol on the
side panel is flashing.
3.2.4

Emergency Stop button

Pressing the Emergency Stop button immediately shuts off all CIBANO 500 outputs and stops the
running measurement. After pressing the Emergency Stop button, Primary Test Manager does not
allow starting a measurement.
To restart the measurement after the reason for the emergency stop has been resolved, release the
Emergency Stop button by carefully turning it, click the Start button in Primary Test Manager, and then
press the Measurement Start/Stop button.
For information about the CIBANO 500 accessories, see "Accessories" in the CIBANO 500 PTM User
Manual.
OMICRON
CIBANO 500 state
CIBANO 500 is up and
running in the stand-by
mode.
A test is prepared and
ready to start.
A test is starting.
Possibly there are
hazardous voltage
and/or current levels at
the CIBANO 500
outputs.
A test is running.
Possibly there are
hazardous voltage
and/or current levels at
the CIBANO 500
outputs.
There are hazardous
voltage levels (> 42 V)
at the CIBANO 500
inputs/outputs
independent of the
measurement state.
Operating condition
Safe operating condition as long as
no voltage is applied from outside
(As long as the warning symbol on
the side panel is off.)
Dangerous operating
condition
Dangerous operating
condition
Dangerous operating
condition