Maintenance Instructions
11
It is highly recommended to test the system weekly by stimulating spasm movements on the bed for 15 seconds
(the default time unless changed by a user via programming) & ensure the alarm call is sent to the pager. To help
keep a record of testing, we have provided a test logging sheet on page 12, please write the serial number in the
space provided.
How to Clean the Devices
It is also recommended to regularly clean the units by wiping with cotton wool pads moistened (compressed until
dripping stops) with a mild detergent (0.5% washing up liquid) solution or by using an alcohol or baby wipe. Avoid
getting any liquid into the micro-USB connector.
Changing Battery's in the Companion
or Reset Button
mini
Remove the base of the monitor. Also please notice the polarity marked into the case (Fig. 3) when replacing the 2x
AAA batteries.
After changing the battery's, the monitor MUST be turned on again by pressing the recessed "TEST" button and
waiting for the 60 second test mode to finish.