Infinite Element Installation And Setup Manual page 16

For the prosthetist
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Element System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Element System should assure that it is used in such an
environment.
Emissions Test
Compliance
RF Emissions
Group 2
CISPR 11
RF Emissions
Class A
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
Not applicable
IEC 61000-3-3
Electromagnetic Environment - Guidance
The Element System must emit electromagnetic
energy in order to perform its intended function.
Nearby electronic equipment may be affected.
The Element System is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
30
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Element System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Element System should assure that it is used in such an
environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for input/
IEC 61000-4-4
output lines
± 1 kV line(s) to
Surge
line(s)
IEC 61000-4-5
± 2 kV line(s) to
earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
Voltage dips, short
T
(60 % dip in U
)
interruptions and
T
for 5 cycles
voltage variations
on power supply
input lines
70 % U
T
(30 % dip in U
)
IEC 61000-4-11
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
Power frequency
(50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8
NOTE: U
is the A.C. mains voltage prior to application of the test level.
T
Compliance
Electromagnetic Environment -
level
Guidance
±6 kV
Floors should be wood, concrete or
contact
ceramic tile. If floors are covered
with synthetic material, the relative
±8 kV air
humidity should be at least 30%.
Not
Not applicable
applicable
Not
Not applicable
applicable
Not
Not applicable
applicable
Power frequency magnetic fields
should be at levels characteristic
30 A/m
of a typical location in a
typical commercial or hospital
environment.
31