PRODUCT COMPONENTS
1. Control box
2. Handpiece
3. ON/OFF foot controller
5. Power Cord
6. Dust Bag
7. Handpiece Holder
PART NAME OF CONTROL BOX
HANDLE
RPM DISPLAY
HANDPIECE HOLDER
VACUUM LEVEL DISPLAY
SPEED CONTROL DIAL
VACUUM LEVEL DIAL
BUTTONS ON THE OPERATION PANEL
■ Power button
: When you push the power button, the power turns on and figures
show on the rotation speed display and vacuum level display.
■ Rotation direction select button
: it is used to choose forward/reverse direction.
■ Hand/Foot select button
: When selecting "Hand", adjust rotation speed or vacuum level by rotating the dial.
: When selecting "Foot", you can operate by pressing foot pedal.
■ Auto/Manual select button
: When selecting "manual", rotation speed and vacuum level can be controlled separately.
: When selecting "auto", rotation speed and vacuum level can be controlled simultaneously.
■ Motor ON/OFF button
: You can operate production
or stop.
CONNECTION OF EACH PART
■ Control box connection
1. Dust bag connection
: Assemble dust bag into the handpiece connector.
2. Handpiece connection
: Connect pins of the hand-piece connector to thecontrol box
correctly. (Pins should be to the upperdirection.
FOOT PEDAL CONNECTOR
ON/OFF BUTTON
POWER SWITC
H
POWER CONNECTION
■ Vacuum level ON/OFF button
: You can operate vacuum function
or stop.
2
3. Foot pedal connection: Connect foot pedal.
4. Power cord connection
: Connect power cord to the back part of the
control box.
OPERATION PROCESS
① Turn power on at the back of the control box.
② Turn on at the front of the control box.
(Figures show on the rotation speed display and vacuum level display when turning power on.)
③ Select rotation direction.
When FWD lamp turns on, it rotates in forward direction and when REV lamp turns on, it rotates in
reverse direction.
④ Select vacuum mode.
· In the model of "AUTO", rotation speed and vacuum level can be controlled simultaneously. (Speed
FUSE BOX
changes simultaneously by turning rotation speed dial.)
· In the model of "MANUAL", rotation speed and vacuum level can be controlled separately.
(Adjustment by each dial)
⑤ Turn on vacuum function by turning on the vacuum ON/OFF button.
(Turn off this function if you don't use.)
⑥ Select hand/foot mode.
■ How to use when select the hand
- Motor rotates when pressing motor
on/off button.
- Motor stops when pressing motor
on/off button one more.
⑦ Turn off the power when finished work.
- Change rotation direction when motor stops compeletely.
- Replace the bur when motor stops completely.
- Never operate the motor wihtout dust bag.
MAINTENANCE OF THE PRODUCT
■ HOW TO CHANGE THE DUST BAG
① Take out the bur and disassemble handle case and motor cap.
② Separate motor by pushing handle case to the front direction.
③ Clean each part with dry cloth.
FOOT CONTROLLER
■ How to use when select the foot
- Press the foot pedal.
Motor rotates during pressing the pedal.
- Motor stops when foot pedal is released.