9.4 NIBP Alarm Information

The following situations may cause a longer measurement time or
unreliable values:
• Animal Movement
• Severe Shock
• Low Heart Rate
• Arrhythmia
• Rapid Pressure Changes
• Extremely Large Animals
Error
SysErr
SysErr2
CuffLoose
CuffErr
Leakage
PressErr
Weak
OveRange
Movement
Protect0
Saturate
TimeOut
Reset
Cause
Self-test fail
NIBP module system error
Cuff is too loose or cuff not connected
Using small cuff in big cuff mode
Valve or gas circuit leak
NIBP Valve is not working appropriately
Animal's pulse is too weak or cuff is loose
Animal's blood pressure exceeds the measurement range
During measurement, interruption in signal or too much
interference
Cuff pressure exceeds the range 300mmHg
Large signal amplitude caused by interference
Big Cuff: cuff pressure over 2kPa(15mmHg) lasting for more
than 3 minutes.
Small Cuff: cuff pressure over 0.67kPa (5mmHg) lasting for
more than 90s
NIBP module reset
45