Maintenance And Cleaning - INSIGHT Vital Signs Plus ECG Operator's Manual

Table of Contents

Advertisement

Prompt and Alarm
Notification
Overrange

8.3 Maintenance and Cleaning

Warning
Turn off the power and disconnect the AC power before cleaning the monitor or the
sensor connected to it.
Reusable temperature probe:
1) The heating of the temperature probe should not exceed 100°C.
2) Do not steam sterilise the probe.
3) Only use alcohol detergent to disinfect.
4) When using the straight probe, it should be possible to cover it with protective
tape.
5) When cleaning the probe, hold the top of the probe in one hand and scrub the
probe downward toward the coupling with a wet, lint free cloth in the other hand.
Warning
If you are using a disposable temperature probe, do not disinfect again or re-use.
Warning
In order to protect the environment, the disposable temperature probe should be
disposed of according to local regulations.
Reason
Out of range
41

Advertisement

Table of Contents
loading

Related Products for INSIGHT Vital Signs Plus ECG

Table of Contents