Before using centrifuge, please carefully read this user manual for its efficient operation and safety.
Before using centrifuge, please carefully read this user manual for its efficient operation and safety.
United Kingdom:
slsinfo@scientific-labs.com
International:
slsexport@scientific-labs.com
USER MANUAL
SLS Lab Basics 12 Place Microcentrifuge
High Speed Mini Centrifuge
Northern Ireland:
NI.sales@scientific-labs.com
Republic of Ireland:
ROI.sales@scientific-labs.com
Code: SLS4600
12301570
Ver.20211215

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SLS4600 and is the answer not in the manual?

Questions and answers

Summary of Contents for SLS SLS4600

  • Page 1 USER MANUAL SLS Lab Basics 12 Place Microcentrifuge High Speed Mini Centrifuge Code: SLS4600 12301570 Ver.20211215 Before using centrifuge, please carefully read this user manual for its efficient operation and safety. Before using centrifuge, please carefully read this user manual for its efficient operation and safety.
  • Page 2: Table Of Contents

    SLS Lab Basics 12 Place Microcentrifuge User Manual Contents Safety Reminder............................... 1 1. Specifications............................... 3 2. Declaration of Conformity..........................3 3. Required Operational Condition........................4 4. Installation..............................5 4.1 Location..............................5 4.2 Connection of the power cord and grounding..................5 5. Structure............................... 6 6.
  • Page 3: Safety Reminder

    SLS Lab Basics 12 Place Microcentrifuge User Manual Safety Reminder Common safety precautions Carefully read the following safety precautions for a thorough understanding. • Follow the instructions and procedures described in this manual to operate this centrifuge safely • Carefully read all safety messages in this manual and the safety instructions on the instrument •...
  • Page 4 • If you require services at site, please sterilize and decontaminate it in advance, and then notify the SLS Service Centre involved in the details of the particular materials •...
  • Page 5: Specifications

    SLS Lab Basics 12 Place Microcentrifuge User Manual 1. Specifications Maximum speed 15000rpm increment: 100rpm Maximum RCF 15100 × g, increment: 100 × g Maximum capacity 2ml × 12 Timer 30 seconds - 99 minutes HOLD continuous operation Driving Motor...
  • Page 6: Required Operational Condition

    SLS Lab Basics 12 Place Microcentrifuge User Manual Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. NOTE: This centrifuge has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
  • Page 7: Installation

    SLS Lab Basics 12 Place Microcentrifuge User Manual SLS Lab Basics 12 Place Microcentrifuge User Manual 4. Installation This section describes the instructions that you should abide when installing the centrifuge to ensure your safety and the optimum performance. Before moving the centrifuge, the rotor must be removed.
  • Page 8: Structure

    SLS Lab Basics 12 Place Microcentrifuge User Manual 5. Structure Door Lock Hook Rotor LCD Display Door release hole Operation Panel Foot Figure 5.1 Front view of centrifuge Door View Window Air Vents Power Inlet Power Switch Figure 5.2 Rear view of centrifuge...
  • Page 9: Operation Panel

    High Speed Mini Micro-Centrifuge SLS Lab Basics 12 Place Microcentrifuge User Manual 6. Operation panel 6. Operation panel Figure 6-1 Operation Panel Item Symbol Name Function High Speed Mini Micro-Centrifu Figure 6.1 Operation Panel Select button Press the button to choose the program which you want to 6.
  • Page 10: Rotor Preparation

    SLS Lab Basics 12 Place Microcentrifuge User Manual 7. Rotor Preparation 7.1 Prepare the samples 7.2 Inject the samples into tubes CAUTION: • Do not overload samples into the centrifuge which will cause leaking • Do not exceed the actual capacity allowed in the user manual 7.3 Keep the tubes balanced...
  • Page 11: Operation

    • Always keep the centrifugal chamber clean when the centrifuge is running. • If the centrifuge makes strange noise during operation, stop it immediately and contact the SLS Press the button to start running. The centrifuge will brake to Service Centre. Notify them of the warning code if displayed Start/ Stop button stop running if pressed during centrifugation.
  • Page 12 SLS Lab Basics 12 Place Microcentrifuge User Manual High Speed Mini Micro-Centrifuge Figure 8.2 The rotor installation Figure 8-2 the rotor installation • Load the rotor to the shaft to ensure rotor is in position until it is connected well with the shaft •...
  • Page 13 SLS Lab Basics 12 Place Microcentrifuge User Manual • There is a circulating function to increase decrease the speed values. Rotate the program button clockwise to change settings from small → large → maximum → minimum. Rotate the program button anti-clockwise to change settings from large → small → minimum → maximum (2) Set the time •...
  • Page 14: Rcf Operation

    SLS Lab Basics 12 Place Microcentrifuge User Manual 8.2 RCF Operation (1) Turn on the power switch. (2) Set an RCF (Relative Centrifugal Force) value. • Press the select button and choose speed unit × g, the speed symbol will flash into RCF value...
  • Page 15: Troubleshooting

    SLS Lab Basics 12 Place Microcentrifuge User Manual • Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution (2) Rotor chamber CAUTION: Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber.
  • Page 16: How To Open The Door

    SLS Lab Basics 12 Place Microcentrifuge User Manual Rotor lid loose Tighten the rotor lid firmly E–02 Door fault The door opened in running Close the door immediately The Run/Stop button is pressed Close the door and then start while the door opening to operate.
  • Page 17: Instructions Of Rotor And Tube

    11. Instructions of rotor and tube 11. Instructions of rotor and tube SLS Lab Basics 12 Place Microcentrifuge User Manual CAUTION: CAUTION: Read the instructions thoroughly, correct use rotor. ⚫ Read the instructions thoroughly, correct use rotor. ⚫ Do not exceed the allowable maximum speed of rotor、 tube and adapters etc., be care that the allowable ⚫...
  • Page 18: Tubes

    SLS Lab Basics 12 Place Microcentrifuge User Manual High Speed Mini Micro-Centrifuge tube Tube 0.5ml 0.5mL A02P2 15000 12780 A02P2 15000 12780 Microtube micro tube 11.1.3 Notice 11.1.3 Notice • The centrifuge rotor can separate samples with density lower than 2.0g/ml. If the separated samples The centrifuge rotor can separate sample which density lower than 2.0g/ml.
  • Page 19 SLS Lab Basics 12 Place Microcentrifuge User Manual Formaldehyde PA: Polyallomer PC: Polycarbonate PP: Polypropylene 11.2.2 Cleaning PC tubes PC materials are low in chemical resistance against alkaline solutions. Avoid using neutral detergents with pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according to the instruction in the makers catalog.
  • Page 20: Calculating Relative Centrifuge Force (Rcf)

    SLS Lab Basics 12 Place Microcentrifuge User Manual 12. Calculation Relative Centrifuge Force (RCF) Relative Centrifuge Force (RCF) can be determined with the following calculation formula. RCF= 1.118 × r × n × 10 R– rotating radius, unit: cm; n-rotating speed, unit: rpm 13.
  • Page 21 Fault Find & Fix On-Site Repairs With service prices starting at just £17.95, For an agreed fee, the SLS Service Centre the Fault Find & Fix Service will collect your will dispatch an engineer to your site to benchtop laboratory equipment, diagnose the...